High-Resolution Diffusion Tensor MR Imaging for Evaluating Myocardial Anisotropy and Fiber Tracking at 3T: the Effect of the Number of Diffusion-Sensitizing Gradient Directions

  • Choi, Sang-Il (Department of Radiology, Asan Medical Center, University of Ulsan College of Medicine) ;
  • Kang, Joon-Won (Department of Radiology, Asan Medical Center, University of Ulsan College of Medicine) ;
  • Chun, Eun-Ju (Division of Cardiovascular Imaging, Department of Radiology, Seoul National University Bundang Hospital, Seoul National University College of Medicine) ;
  • Choi, Seong-Hoon (Department of Radiology, Ulsan University Hospital, University of Ulsan College of Medicine) ;
  • Lim, Tae-Hwan (Department of Radiology, Asan Medical Center, University of Ulsan College of Medicine)
  • Received : 2009.04.26
  • Accepted : 2009.08.11
  • Published : 2010.02.01

Abstract

Objective: We wanted to evaluate the effect of the number of diffusion-sensitizing gradient directions on the image quality for evaluating myocardial anisotropy and fiber tracking by using in vitro diffusion tensor MR imaging (DT-MRI). Materials and Methods: The DT-MR images, using a SENSE-based echoplanar imaging technique, were acquired from ten excised porcine hearts by using a 3T MR scanner. With a b-value of 800 s/$mm^2$, the diffusion tensor images were obtained for 6, 15 and 32 diffusion-sensitizing gradient directions at the midventricular level. The number of tracked fibers, the fractional anisotropy (FA), and the length of the tracked fibers were measured for the quantitative analysis. Two radiologists assessed the image quality of the fiber tractography for the qualitative analysis. Results: By increasing the number of diffusion-sensitizing gradient directions from 6 to 15, and then to 32, the FA and standard deviation were significantly reduced (p < 0.01), and the number of tracked fibers and the length of the tracked fibers were significantly increased (p < 0.01). The image quality of the fiber tractography was significantly increased with the increased number of diffusionsensitizing gradient directions (p < 0.01). Conclusion: The image quality of in vitro DT-MRI is significantly improved as the number of diffusion-sensitizing gradient directions is increased.

Keywords

References

  1. Rademakers FE, Rogers WJ, Guier WH, Hutchins GM, Siu CO, Weisfeldt ML, et al. Relation of regional cross-fiber shortening to wall thickening in the intact heart. Three-dimensional strain analysis by NMR tagging. Circulation 1994;89:1174-1182 https://doi.org/10.1161/01.CIR.89.3.1174
  2. Taber LA, Yang M, Podszus WW. Mechanics of ventricular torsion. J Biomech 1996;29:745-752 https://doi.org/10.1016/0021-9290(95)00129-8
  3. Spotnitz HM. Macrodesign, structure, and mechanics of the left ventricle. J Thorac Cardiovasc Surg 2000;119:1053-1057 https://doi.org/10.1016/S0022-5223(00)70106-1
  4. Streeter DD Jr, Spotnitz HM, Patel DP, Ross J Jr, Sonnenblick EH. Fiber orientation in the canine left ventricle during diastole and systole. Circ Res 1969;24:339-347 https://doi.org/10.1161/01.RES.24.3.339
  5. Wu EX, Wu Y, Nicholls JM, Wang J, Liao S, Zhu S, et al. MR diffusion tensor imaging study of postinfarct myocardium structural remodeling in a porcine model. Magn Reson Med 2007;58:687-695 https://doi.org/10.1002/mrm.21350
  6. Edelman RR, Gaa J, Wedeen VJ, Loh E, Hare JM, Prasad P, et al. In vivo measurement of water diffusion in the human heart. Magn Reson Med 1994;32:423-428 https://doi.org/10.1002/mrm.1910320320
  7. Reese TG, Weisskoff RM, Smith RN, Rosen BR, Dinsmore RE, Wedeen VJ. Imaging myocardial fiber architecture in vivo with magnetic resonance. Magn Reson Med 1995;34:786-791 https://doi.org/10.1002/mrm.1910340603
  8. Tseng WY, Reese TG, Weisskoff RM, Wedeen VJ. Cardiac diffusion tensor MRI in vivo without strain correction. Magn Reson Med 1999;42:393-403 https://doi.org/10.1002/(SICI)1522-2594(199908)42:2<393::AID-MRM22>3.0.CO;2-F
  9. Holmes AA, Scollan DF, Winslow RL. Direct histological validation of diffusion tensor MRI in formaldehyde-fixed myocardium. Magn Reson Med 2000;44:157-161 https://doi.org/10.1002/1522-2594(200007)44:1<157::AID-MRM22>3.0.CO;2-F
  10. Scollan DF, Holmes A, Winslow R, Forder J. Histological validation of myocardial microstructure obtained from diffusion tensor magnetic resonance imaging. Am J Physiol 1998;275:H2308-H2318
  11. Tseng WY, Wedeen VJ, Reese TG, Smith RN, Halpern EF. Diffusion tensor MRI of myocardial fibers and sheets: correspondence with visible cut-face texture. J Magn Reson Imaging 2003;17:31-42 https://doi.org/10.1002/jmri.10223
  12. Wu MT, Tseng WY, Su MY, Liu CP, Chiou KR, Wedeen VJ, et al. Diffusion tensor magnetic resonance imaging mapping the fiber architecture remodeling in human myocardium after infarction: correlation with viability and wall motion. Circulation 2006;114:1036-1045 https://doi.org/10.1161/CIRCULATIONAHA.105.545863
  13. Lee JW, Kim JH, Kang HS, Lee JS, Choi JY, Yeom JS, et al. Optimization of acquisition parameters of diffusion-tensor magnetic resonance imaging in the spinal cord. Invest Radiol 2006;41:553-559 https://doi.org/10.1097/01.rli.0000221325.03899.48
  14. Wu EX, Wu Y, Tang H, Wang J, Yang J, Ng MC, et al. Study of myocardial fiber pathway using magnetic resonance diffusion tensor imaging. Magn Reson Imaging 2007;25:1048-1057 https://doi.org/10.1016/j.mri.2006.12.008
  15. Tanenbaum LN. 3-T MR imaging: ready for clinical practice. AJNR Am J Neuroradiol 2004;25:1626-1627
  16. Zhai G, Lin W, Wilber KP, Gerig G, Gilmore JH. Comparisons of regional white matter diffusion in healthy neonates and adults performed with a 3.0-T head-only MR imaging unit. Radiology 2003;229:673-681 https://doi.org/10.1148/radiol.2293021462
  17. Choi SI, Choi SH, Kim ST, Lim KH, Lim CH, Gong GY, et al. Irreversibly damaged myocardium at MR imaging with a necrotic tissue-specific contrast agent in a cat model. Radiology 2000;215:863-868 https://doi.org/10.1148/radiology.215.3.r00jn01863
  18. Jeong AK, Choi SI, Kim DH, Park SB, Lee SS, Choi SH, et al. Evaluation by contrast-enhanced MR imaging of the lateral border zone in reperfused myocardial infarction in a cat model. Korean J Radiol 2001;2:21-27 https://doi.org/10.3348/kjr.2001.2.1.21
  19. Jones DK. The effect of gradient sampling schemes on measures derived from diffusion tensor MRI: a Monte Carlo study. Magn Reson Med 2004;51:807-815 https://doi.org/10.1002/mrm.20033
  20. Okada T, Miki Y, Fushimi Y, Hanakawa T, Kanagaki M, Yamamoto A, et al. Diffusion-tensor fiber tractography: intraindividual comparison of 3.0-T and 1.5-T MR imaging. Radiology 2006;238:668-678 https://doi.org/10.1148/radiol.2382042192
  21. Jaermann T, Crelier G, Pruessmann KP, Golay X, Netsch T, van Muiswinkel AM, et al. SENSE-DTI at 3 T. Magn Reson Med 2004;51:230-236 https://doi.org/10.1002/mrm.10707
  22. Nagae-Poetscher LM, Jiang H, Wakana S, Golay X, van Zijl PC, Mori S. High-resolution diffusion tensor imaging of the brain stem at 3 T. AJNR Am J Neuroradiol 2004;25:1325-1330
  23. Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: sensitivity encoding for fast MRI. Magn Reson Med 1999;42:952-962 https://doi.org/10.1002/(SICI)1522-2594(199911)42:5<952::AID-MRM16>3.0.CO;2-S
  24. Cercignani M, Horsfield MA, Agosta F, Filippi M. Sensitivity-encoded diffusion tensor MR imaging of the cervical cord. AJNR Am J Neuroradiol 2003;24:1254-1256
  25. Naganawa S, Koshikawa T, Kawai H, Fukatsu H, Ishigaki T, Maruyama K, et al. Optimization of diffusion-tensor MR imaging data acquisition parameters for brain fiber tracking using parallel imaging at 3 T. Eur Radiol 2004;14:234-238 https://doi.org/10.1007/s00330-003-2163-6
  26. Pierpaoli C, Basser PJ. Toward a quantitative assessment of diffusion anisotropy. Magn Reson Med 1996;36:893-906 https://doi.org/10.1002/mrm.1910360612
  27. Jiang Y, Guccione JM, Ratcliffe MB, Hsu EW. Transmural heterogeneity of diffusion anisotropy in the sheep myocardium characterized by MR diffusion tensor imaging. Am J Physiol Heart Circ Physiol 2007;293:H2377-H2384 https://doi.org/10.1152/ajpheart.00337.2007
  28. Chang Y, Lee YJ, Kim YS, Kang DS, Sohn CH, Woo SK, et al. Histogram analysis of noise performance on fractional anisotropy brain MR image with different diffusion gradient numbers. J Korean Radiol Soc 2005;52:87-92 [Korean] https://doi.org/10.3348/jkrs.2005.52.2.87

Cited by

  1. Abnormal Motion of the Interventricular Septum after Coronary Artery Bypass Graft Surgery: Comprehensive Evaluation with MR Imaging vol.11, pp.6, 2010, https://doi.org/10.3348/kjr.2010.11.6.627