Comparison of $HbA_{1c}$ Analyzers: D-10, Variant II Turbo, Cobas Integra 800, and Afinion AS100

  • Lee, Jin-Young (Department of Laboratory Medicine, School of Medicine, Ewha Womans University) ;
  • Hong, Ki-Sook (Department of Laboratory Medicine, School of Medicine, Ewha Womans University) ;
  • Cho, Sung-Eun (Eone Reference Laboratory)
  • Received : 2009.10.16
  • Accepted : 2010.07.15
  • Published : 2010.08.31

Abstract

Background : The purpose of this study was to evaluate the performance and agreement among $HbA_{1c}$ values measured using selected analyzers certified by the National Glycohemoglobin Standardization Program (NGSP) and standardized by the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC). Methods : $HbA_{1c}$ determined using D-10 (Bio-Rad, USA), Variant II Turbo (Turbo; Bio-Rad, USA), Cobas Integra 800 (Integra; Roche, Switzerland) and Afinion AS100 (Afinion; Axis-Shield, Norway) were compared with each other. Precision and method comparisons with Deming regression were evaluated according to CLSI recommendations. We also compared the $HbA_{1c}$ values obtained with each analyzer using either IFCC or NGSP methods by correlation analysis and kappa statistics. Results : The repeatability and method/device precisions of D-10 and Afinion were acceptable. The correlation coefficients of $HbA_{1c}$ were 0.986 for D-10 vs. Afinion, 0.997 for D-10 vs. Turbo, 0.988 for D-10 vs. Integra, and 0.991 for Integra vs. Afinion. The average biases of $HbA_{1c}$ Afinion (IFCC) and $HbA_{1c}$ Integra (IFCC) against $HbA_{1c}$ D-10 (NGSP) were -1.90% and -1.79%, respectively. Kappa agreement statistics for the three diabetic control group $HbA_{1c}$ values of "less than 6.5%", "6.5%-7.5%", and "greater than 7.5%" for D-10 vs. Turbo, D-10 vs. Integra, and D-10 vs. Afinion were 0.872, 0.836, and 0.833, respectively. Conclusions : The strong correlations and good clinical agreements of $HbA_{1c}$ between each analyzer expressed in terms of either NGSP or IFCC-derived NGSP indicate that these analyzers can be used interchangeably.

배경 : National Glycohemoglobin Standardization Program (NGSP)법 인증을 받거나 International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) 표준화를 따르는 $HbA_{1c}$ 측정기기들의 검사 수행능력을 비교 평가하였다. 방법 : D-10 (Bio-Rad, USA), Variant II Turbo (Turbo; Bio-Rad, USA), Cobas Integra 800 (Integra; Roche, Switzerland)과 Afinion AS100 (Afinion; Axis-Shield, Norway)을 서로 비교하였다. CLSI 권고에 따라 정밀도 검증 및 검사 방법 비교를 하였고, 카파 통계를 이용하여 일치도 분석을 하였다. 결과 : D-10과 Afinion의 재현성과 방법/장비 정밀도는 적합하였고, D-10과 Afinion, D-10과 Turbo 및 D-10과 Integra의 상관계수는 각각 0.986, 0.997 및 0.988이었다. Integra와 Afinion의 상관계수는 0.991이었다. D-10 (NGSP)에 비해 Afinion (IFCC) 및 Integra (IFCC) $HbA_{1c}$값의 평균 오차는 각각 -1.90%, -1.79%이었다. $HbA_{1c}$ 측정값을 당뇨병 진단 기준에 의한 분류에 의하여 '6.5% 미만', '6.5-7.5%' 및 '7.5 이상' 세 군으로 나누었을 때, D-10과 Turbo, D-10과 Integra 및 D-10과 Afinion의 카파 통계량은 각각 0.872, 0.836 및 0.833이었다. 결론 : 각 기기들의 NGSP $HbA_{1c}$값 사이에 높은 상관성과 일치도를 보였고, NGSP $HbA_{1c}$값과 IFCC-변환 NGSP $HbA_{1c}$값 사이의 호환성을 확인할 수 있었다.

Keywords

References

  1. Nathan DM, Singer DE, Hurxthal K, Goodson JD. The clinical information value of the glycosylated hemoglobin assay. N Engl J Med 1984;310:341-6. https://doi.org/10.1056/NEJM198402093100602
  2. Benjamin RJ and Sacks DB. Glycated protein update: implications of recent studies, including the diabetes control and complications trial. Clin Chem 1994;40:683-7.
  3. Steffes M, Cleary P, Goldstein D, Little R, Wiedmeyer HM, Rohlfing C, et al. Hemoglobin A1c measurements over nearly two decades: sustaining comparable values throughout the Diabetes Control and Complications Trial and the Epidemiology of Diabetes Interventions and Complications study. Clin Chem 2005;51:753-8. https://doi.org/10.1373/clinchem.2004.042143
  4. Little RR, Rohlfing CL, Wiedmeyer HM, Myers GL, Sacks DB, Goldstein DE; NGSP Steering Committee. The national glycohemoglobin standardization program: a five-year progress report. Clin Chem 2001;47:1985-92.
  5. Jeppsson JO, Kobold U, Barr J, Finke A, Hoelzel W, Hoshino T, et al. Approved IFCC reference method for the measurement of HbA1c in human blood. Clin Chem Lab Med 2002;40:78-89. https://doi.org/10.1515/CCLM.2002.016
  6. Kobold U, Jeppsson JO, Dülffer T, Finke A, Hoelzel W, Miedema K. Candidate reference methods for hemoglobin A1c based on peptide mapping. Clin Chem 1997;43:1944-51.
  7. Hoelzel W, Weykamp C, Jeppsson JO, Miedema K, Barr JR, Goodall I, et al. IFCC reference system for measurement of hemoglobin A1c in human blood and the national standardization schemes in the United States, Japan, and Sweden: a method-comparison study. Clin Chem 2004;50:166-74. https://doi.org/10.1373/clinchem.2003.024802
  8. Sacks DB. Global harmonization of hemoglobin A1c. Clin Chem 2005; 51:681-3. https://doi.org/10.1373/clinchem.2004.047431
  9. Penttila IM, Halonen T, Punnonen K, Tiikkainen U. Best use of the recommended IFCC reference method, material and values in HbA1C analyses. Scand J Clin Lab Invest 2005;65:453-62. https://doi.org/10.1080/00365510510025809
  10. Mosca A, Goodall I, Hoshino T, Jeppsson JO, John WG, Little RR, et al. Global standardization of glycated hemoglobin measurement: the position of the IFCC Working Group. Clin Chem Lab Med 2007; 45:1077-80. https://doi.org/10.1515/CCLM.2007.246
  11. Consensus Committee. Consensus statement on the worldwide standardization of the hemoglobin A1c measurement: the American Diabetes Association, European Association for the Study of Diabetes International Federation of Clinical Chemistry and Laboratory Medicine, and the International Diabetes Federation. Diabetes Care 2007;30:2399-400. https://doi.org/10.2337/dc07-9925
  12. Marzullo C, Minery M. Evaluation of D10 hemoglobin testing system for hemoglobin A1c assay. Ann Biol Clin (Paris) 2008;66:95-9.
  13. Leon-Justel A, Santotoribio JD, Dominguez-Pascual I, Delgado AL, Macias C, Herrera MT, et al. Influence of reduction in the elution times on HPLC glycohaemoglobin results. Clin Biochem 2009;42:1582-4. https://doi.org/10.1016/j.clinbiochem.2009.07.003
  14. Fleming JK. Evaluation of HbA1c on the Roche COBAS Integra 800 closed tube system. Clin Biochem 2007;40:822-7. https://doi.org/10.1016/j.clinbiochem.2007.03.017
  15. Tholen DW, Kallner A, et al. eds. Evaluation precision performance of quantitative measurement method: approved guideline. CLSI document EP5-A2. 2nd ed. Wayne, PA: NCCLS, 2004.
  16. Krouwer JS, Tholen DW, et al. eds. Method comparison and bias estimation using patient samples: approved guideline. CLSI document EP9-A2. 2nd ed. Wayne, PA: NCCLS, 2002.
  17. Song JH, Kwon KC, Kim JH, Kim JW, Min WK, Lee SY, et al. Annual report on external quality assessment in metabolic disorders in Korea (2008). J Lab Med Qual Assur 2009;31:143-59. (송정한, 권계철, 김정호, 김종원, 민원기, 이수연 등. 대사질환검사 신빙도조사 결과 보고(2008). 임상검사와정도관리 2009;31:143-59.)
  18. Goldstein DE, Chenault VM, et al. eds. Harmonization of gylcohemoglobin measurement. CLSI document C44-A. Wayne, PA: NCCLS, 2002.
  19. Miedema K. Standardization of HbA1c and optimal range of monitoring. Scand J Clin Lab Invest Suppl 2005;240:61-72.
  20. Dhatt GS, Agarwal MM, Bishawi B. HbA1c: a comparison of NGSP with IFCC transformed values. Clin Chim Acta 2005;358:81-6. https://doi.org/10.1016/j.cccn.2005.02.009
  21. Paisooksantivatana K, Kongsomgan A, Khupulsup K. NGSP and IFCC-derived NGSP HbA1c can be used interchangeably. Diabetes Res Clin Pract 2009;85:e22-5. https://doi.org/10.1016/j.diabres.2009.06.005
  22. Nordin G and Dybkaer R. Recommendation for term and measurement unit for "HbA1c". Clin Chem Lab Med 2007;45:1081-2. https://doi.org/10.1515/CCLM.2007.245

Cited by

  1. Analytical Performance of Bio-Rad D-100 on a Hemoglobin A1c Assay vol.7, pp.2, 2017, https://doi.org/10.3343/lmo.2017.7.2.59
  2. Assessment of Two Diabetes Point-of-care Analyzers Measuring Hemoglobin A1c in the Peruvian Amazon vol.84, pp.4, 2018, https://doi.org/10.29024/aogh.2368