Comparison of Blood Glucose Measurements Using Samples Obtained from the Forearm, Finger Skin Puncture, and Venous Serum

  • Park, Kyung-Soon (Personalized Tumor Engineering Research Center, College of Medicine, Chungbuk National University) ;
  • Park, Mi-Sook (Personalized Tumor Engineering Research Center, College of Medicine, Chungbuk National University) ;
  • Cha, Young-Joo (Department of Clinical Pathology, College of Medicine, Chung-Ang University) ;
  • Kim, Wun-Jae (Personalized Tumor Engineering Research Center, College of Medicine, Chungbuk National University) ;
  • Choi, Seong-Su (Personalized Tumor Engineering Research Center, College of Medicine, Chungbuk National University) ;
  • Kim, Kyoung-Ok (Department of Nursing, Seoul Asan Medical Center) ;
  • Cha, Eun-Jong (Personalized Tumor Engineering Research Center, College of Medicine, Chungbuk National University) ;
  • Kim, Kyung-Ah (Personalized Tumor Engineering Research Center, College of Medicine, Chungbuk National University)
  • Received : 2009.11.28
  • Accepted : 2010.04.19
  • Published : 2010.06.30

Abstract

Background : Blood glucose testing (BGT) at the forearm minimizes the pain experienced during sampling of capillary blood. We compared the BGT results for forearm sampling with those for standard finger skin puncture and venous serum to evaluate the clinical validity of forearm BGT. Methods : BGT was performed on the finger ($G_F$) and forearm ($G_A$) with a portable glucometer in 555 subjects, including 61 diabetic patients, under fasting conditions. BGT with venous serum ($G_V$) was followed within an hour in 514 subjects. Simple linear regression, intraclass correlation, and Passing-Bablok regression analyses were performed using the $G_A-G_F$ and $G_A-G_V$ data. Results : $G_A$ showed an excellent linear relationship with both $G_F$ and $G_V$ with a Pearson correlation coefficient (r) of 0.97 (P<0.0001) in the patient group, which was similar to the findings in the normal group except for the lower r values. The mean bias between $G_A$ and $G_F$ and between $G_A$ and $G_V$ were within ${\pm}10$ mg/dL in both groups. The intraclass correlation coefficients were slightly smaller than the corresponding r values, but they showed the same tendency in both groups. In the Passing-Bablok analyses, the 95% confidence intervals of the slope and intercept parameters were < ${\pm}20%$ of unity and < ${\pm}20$ mg/dL, respectively, which were within the acceptable ranges. All 3 statistical analyses supported the satisfactory agreement of $G_A$ with $G_F$ or $G_V$. Conclusions : BGT at the forearm was highly consistent with the standard BGT, thereby confirming its applicability in clinical practice for self-testing under steady fasting conditions.

Keywords

Acknowledgement

Supported by : NRF

References

  1. Zimmet P, Alberti KG, Shaw J. Global and societal implications of the diabetes epidemic. Nature 2001;414:782-7. https://doi.org/10.1038/414782a
  2. Kim MH, Kim MK, Choi BY, Shin YJ. Educational disparities in the metabolic syndrome in a rapidly changing society: the case of South Korea. Int J Epidemiol 2005;34:1266-73. https://doi.org/10.1093/ije/dyi175
  3. Cho NH. Diabetes epidemiology in Korean. J Korean Diabetes Assoc 2001;25:1-10.
  4. Korea National Statistical Office. Annual report on the cause of death statistics (based on vital registration) 2006. Statstice Korea, 2007.
  5. Kim EJ, Min HK, et al. eds. Diabetics. 3rd ed. Seoul: Korea Medical Book Company, 2005:1.
  6. Parchman ML, Pugh JA, Noel PH, Larme AC. Continuity of care, self-management behaviors, and glucose control in patients with type 2 diabetes. Med Care 2002;40:137-44. https://doi.org/10.1097/00005650-200202000-00008
  7. Kim EJ, Min HK, et al. eds. Diabetics. 2nd ed. Seoul: Korea Medical Book Company, 1998:303-8.
  8. Skyler JS. Self-monitoring of blood glucose. Med Clin North Am 1982;66:1227-50. https://doi.org/10.1016/S0025-7125(16)31360-8
  9. Choi JY. A comparison on the degree of pain according to methods of blood sugar test between DM patients and healthy group. J Korean Acad Nurs 2003;33:928-35. https://doi.org/10.4040/jkan.2003.33.7.928
  10. Park SW, Kim DJ, Kim JY, Kim HY, Min KW, Baek SH, et al. Current status of diabetes management in Korea using national insurance database. Korean Diabetes Association 19th Spring Meeting 2006:227-8.
  11. Cowett RM and D'Amico LB. Capillary (heelstick) versus venous blood sampling for the determination of glucose concentration in the neonate. Biol Neonate 1992;62:32-6. https://doi.org/10.1159/000243850
  12. Barker DP and Rutter N. Exposure to invasive procedures in neonatal intensive care unit admissions. Arch Dis Child Fetal Neonatal Ed 1995;72:F47-8. https://doi.org/10.1136/fn.72.1.F47
  13. McIntosh N, van Veen L, Brameyer H. Alleviation of the pain of heel prick in preterm infants. Arch Dis Child Fetal Neonatal Ed 1994;70:F177-81. https://doi.org/10.1136/fn.70.3.F177
  14. Cunningham DD, Henning TP, Shain EB, Young DF, Elstrom TA, Taylor EJ, et al. Vacuum-assisted lancing of the forearm: an effective and less painful approach to blood glucose monitoring. Diabetes Technol Ther 2000;2:541-8. https://doi.org/10.1089/15209150050501952
  15. Park MS, Park KS, Kim KA, Jun MH, Kim TI, Lee TS, et al. Vacuum assisted auto-lancing technique for capillary blood sampling on the forearm with minimized pain. J Biomed Eng Res 2004;25:557-63.
  16. Lock JP, Szuts EZ, Malomo KJ, Anagnostopoulos A. Whole-blood glucose testing at alternate sites: glucose values and hematocrit of capillary blood drawn from fingertip and forearm. Diabetes Care 2002;25:337-41. https://doi.org/10.2337/diacare.25.2.337
  17. Cohen M, Boyle E, Delaney C, Shaw J. A comparison of blood glucose meters in Australia. Diabetes Res Clin Pract 2006;71:113-8. https://doi.org/10.1016/j.diabres.2005.05.013
  18. Reinauer H, Home RD, et al. eds. Laboratory diagnosis and monitoring of diabetes mellitus. Geneva, Switzerland: World Health Organization, 2002:11-2.
  19. U. S. Food and Drug Administration. Review criteria for assessment of portable blood glucose monitoring in-vitro diagnostic devices using glucose oxidase, dehydrogenase, or hexokinase methodology. Guidance document (medical devices). U. S. Department of Health and Human Service, 1997:5.
  20. Koch GG. Intraclass correlation coefficient. In: Kotz S and Johnson NL, eds. Encyclopedia of statistical sciences. New York: John Wiley & Sons, 1982: 213-7.
  21. Passing H and Bablok W. A new biometrical procedure for testing the equality of measurements from two different analytical methods. Application of linear regression procedures for method comparison studies in clinical chemistry, Part I. J Clin Chem Clin Biochem 1983;21:709-20.
  22. National Committee for Clinical Laboratory Standards. Point-of-care blood glucose testing in acute and chronic care facilities: approved guideline. Document C30-A2. 2nd ed. Wayne, PA: NCCLS, 2002;22.
  23. International Organization for Standardization. ISO 15197:2003. In vitro diagnostic test systems requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. Geneva: International Organization for Standardization, 2003.
  24. Clarke WL, Cox D, Gonder-Frederick LA, Carter W, Pohl SL. Evaluating clinical accuracy of systems for self-monitoring of blood glucose. Diabetes Care 1987;10:622-8. https://doi.org/10.2337/diacare.10.5.622
  25. Kim DK, Won JH, Potapov SN, Meriakri VV, Chigryai EE. Basic investigation for the non-invasive measurement of blood glucose concentrations by millimeter waves. IEEK Journal (SC) 1998;42:3-6.
  26. Waynant RW and Chenault VM. Overview of non-invasive fluid glucose measurement using optical techniques to maintain glucose control in diabetes mellitus. LEOS newsletter 1998;12:3-6.
  27. Paul EK, Ruth SW, John BN. Carbohydrates. In: Henry JB, ed. Clinical diagnosis and management by laboratory methods. Philadelphia, PA: WB Sanders, 2001:214.
  28. Kwon MJ and Lee SY. Evaluation of GLUCOCARD X-METER glucose monitoring system. Korean J Lab Med 2008;28:8-15. https://doi.org/10.3343/kjlm.2008.28.1.8
  29. Jungheim K and Koschinsky T. Glucose monitoring at the arm: risky delays of hypoglycemia and hyperglycemia detection. Diabetes Care 2002;25:956-60. https://doi.org/10.2337/diacare.25.6.956
  30. Jungheim K and Koschinsky T. Risky delay of hypoglycemia detection by glucose monitoring at the arm. Diabetes Care 2001;24:1303-6. https://doi.org/10.2337/diacare.24.7.1303
  31. Ellison JM, Stegmann JM, Colner SL, Michael RH, Sharma MK, Ervin KR, et al. Rapid changes in postprandial blood glucose produce concentration differences at finger, forearm, and thigh sampling sites. Diabetes Care 2002;25:961-4. https://doi.org/10.2337/diacare.25.6.961
  32. Fagrell B. Microcirculation of the skin. In the physiology and pharmacology of the microcirculation. Orlando: Academic Press, 1984;133-80.
  33. Rendell M, Bergman T, O'Donnell G, Drobny E, Borgos J, Bonner RF. Microvascular blood flow, volume, and velocity measured by laser Doppler techniques in IDDM. Diabetes 1989;38:819-24. https://doi.org/10.2337/diabetes.38.7.819
  34. Stansberry KB, Peppard HR, Babyak LM, Popp G, McNitt PM, Vinik AI. Primary nociceptive afferents mediate the blood flow dysfunction in non-glabrous (hairy) skin of type 2 diabetes: a new model for the pathogenesis of microvascular dysfunction. Diabetes Care 1999;22:1549-54. https://doi.org/10.2337/diacare.22.9.1549
  35. McGarraugh G. Response to Jungheim and Koschinsky: glucose monitoring at the arm. Diabetes Care 2001;24:1304-6. https://doi.org/10.2337/diacare.24.7.1304
  36. Pohl SL, Gonder-Frederick L, Cox DJ, Evans WS. Self-measurement of blood glucose concentration: clinical significance of patient-generated measurements. Diabetes Care 1985;8:617-9. https://doi.org/10.2337/diacare.8.6.617
  37. Tieszen KL and New JP. Alternate site blood glucose testing: do patients prefer it? Diabet Med 2003;20:325-8. https://doi.org/10.1046/j.1464-5491.2003.00927.x
  38. Dai KS, Tai DY, Ho P, Chen CC, Peng WC, Chen ST, et al. Accuracy of the EasyTouch blood glucose self-monitoring system: a study of 516 cases. Clin Chim Acta 2004;349:135-41. https://doi.org/10.1016/j.cccn.2004.06.010

Cited by

  1. Influence of sample collection and preanalytical sample processing on the analyses of biological markers in the European multicentre study IDEFICS vol.35, pp.suppl1, 2010, https://doi.org/10.1038/ijo.2011.41
  2. Comparison of Hematologic and Biochemical Test Results in Blood Samples Obtained by Jugular Venipuncture Versus Nail Clip in Moluccan Cockatoos (Cacatua moluccensis) vol.29, pp.4, 2010, https://doi.org/10.1647/2009-046r5
  3. Evaluation of four portable blood glucose meters in diabetic and non-diabetic dogs and cats. vol.36, pp.1, 2010, https://doi.org/10.1080/01652176.2015.1092617