Abstract
In this article, the performance of the THC analyzer was inspected using two different span gases of methane ($CH_4$) and propane ($C_3H_8$). To explore the effect of standard gas selection, MicroFID system was tested by the following procedures. Initially, the system is spanned by propane gas of 60 ppm (or 180 ppmC). The system is then run against methane standards prepared at 5 different concentrations of 200, 250, 300, 400, and 500 ppm. According to the suggestion of the KMOE's test procedure to use multiplying a factor of 3 (for propane), the resulting THC values derived by methane standards were systematically biased with ~500% error relative to true value. This paper discusses the interpretation procedures to obtain the least biased THC values for a given span set-up.