Abstract

For treatment of Total Skin Electron beam Therapy (TSET), measurement of dose at various conditions is need on the contrary to usual radiotherapy. When treating TSET with modified Stanford technique based on linear accelerator, the energy of treatment electron beam, the spatial dose distribution and the actual doses deposited on the surface of the patient were measured by using EBT2. The measured energy of the electron beam was agreed with the value that measured by ionization chamber, and the spatial dose distribution at the patient position and the doses at several point on the patient's skin could be easily measured by EBT2 film. The dose on the patient that was measured by EBT2 film showed good agreement with the data measured simultaneously by TLD. With the results of this study, it was proven that the EBT2 film can be one of the useful dosimeter for TSET.

전신 피부 전자선 치료(total skin electron beam therapy, TSET)를 위해서는 흔히 행해지는 일반적인 방사선 치료와 달리 치료 전 다양한 조건에서의 선량 측정과 더불어, 치료 중 지속적인 환자 신체부위별 선량 측정이 요구된다. 본 연구에서는 선형 가속기에 기반한 modified Stanford Technique으로 전신 피부 전자선 치료를 할 때 치료에 필요한 치료 전자선의 에너지와 선량의 공간적 분포 및 치료 중 환자의 각 부위별 조사선량을 EBT2 필름을 이용하여 측정 하였다. 전자선의 에너지는 이온전리함으로 측정한 값과 비교하였을 때 잘 일치하였고, 선량의 공간 분포 및 환자 각 신체부위에서의 선량 분포는 EBT2 필름을 이용하여 편리하게 측정할 수 있었다. 또한 TSET 치료중 EBT2 필름을 사용하여 in-vivo로 측정된 환자 신체 부위별 선량분포의 변화는 열형광선량계(thermoluminescent dosimeter, TLD)로 동시에 측정한 값과 비교하였을 때 잘 일치함을 알 수 있었다. 이로써 EBT2 필름의 전신 피부 전자선 치료를 위한 선량계로서의 가능성을 확인하였다.

Keywords

References

  1. Meyler TS, Blumberg AL, Purser P: Total skin electron beam therapy in mycosis fungoides. Cancer 42:1171-1176 (1978) https://doi.org/10.1002/1097-0142(197809)42:3<1171::AID-CNCR2820420321>3.0.CO;2-8
  2. Micaily B, Vonderheid EC, Brady LW, Andrews C: Total skin electron beam and total nodal irradiation for treatment of patients with cutaneous T-cell lymphoma. Int J Radiat Oncol Biol Phys 11:1111-1115 (1985) https://doi.org/10.1016/0360-3016(85)90057-4
  3. Maingon P, TrucG, Dalac S, et al: Radiotherapy of advanced mycosis fungoides: Indications and results of total skin electron beam and photon beam irradiation. Radiot Oncol 54:73-78 (2000) https://doi.org/10.1016/S0167-8140(99)00162-0
  4. Chen Z, Agostinelli AG, Wilson LD, Nath R: Matching the dosimetry characteristics of a dual-field Stanford technique to a customized single-field Stanford technique for total skin electron therapy. Int J Radiation Oncology Biol Phys 59:872-885 (2004) https://doi.org/10.1016/j.ijrobp.2004.02.046
  5. Tetennes PJ, Goodwin PN: Comparative study of superficial whole-body radiotherapeutic techniques using a 4 MeV non angled electyron beam. Radiology 122:219-226 (1977)
  6. Welson LW, Jones GW, Kim D, et al: Experience with total skin electron beam therapy in combination with extracorporeal photopheresis in the management of patients with erythrodermic (T4) mycosis fungoides. J Am Acad Dermatol 43:54-60 (2000) https://doi.org/10.1067/mjd.2000.105510
  7. Karzmark CJ, Loevinger R, Steele RE, Weissbluth M: A technique for large field superficial electron therapy. Radiology 74:633-644 (1960)
  8. Page V, Gardener A, Karzmark CJ: Patient dosimetry in the electron treatment of large superficial lesions. Radiology 94:635-641 (1970)
  9. Piermattei A, Rossi G, Azario L, Fidanzio A, Balducci M, Valentini V: Electron beam dosimetry of total skin electron therapy. Radiol Med 108:549-559 (2004)
  10. Gamble LM, Farrell TJ, Jones GW, Hayward JE: Twodimensional mapping of underdosed areas using radiochromic film for patients undergoing total skin electron beam radiotherapy. Int J Radiation Oncology Biol Phys 62:920-924 (2005) https://doi.org/10.1016/j.ijrobp.2005.02.057
  11. Weaver RD, Gerbi BJ, Dusenbery KE: Evaluation of dose variation during total skin electron irradiation using thermoluminescent dosimeters. Int J Radiat Oncol Biol Phys 33:475-478 (1995) https://doi.org/10.1016/0360-3016(95)00161-Q
  12. Antolak JA, Cundiff JH, Ha CS: Utilization of thermoluminescent dosimetry in total skin electron beam radiotherapy of mycosis fungoides. Int J Radiat Oncol Biol Phys 40:101-108 (1998) https://doi.org/10.1016/S0360-3016(97)00585-3
  13. Fraass BA, Roberson PL, Glatstein E: Whole-skin electron treatment: patient skin dose distribution. Radiology 146:811-814 (1983)
  14. Bufacchi A, Carosi A, Adorante N, et al: In vivo EBT radiochromic film dosimetry of electron beam for total skin electron therapy (TSET). Physica Medica 23:67-72 (2007) https://doi.org/10.1016/j.ejmp.2007.03.003
  15. Reynard E, Evans M, Devic S, et al: Rotational total skin electron irradiation with a linear accelerator. J Appl Clin Med Phys 9:123-134 (2008)
  16. Jones GW, Kacinski M, Wilson LD, et al: Total skin electron radiation in the management of mycosis fungoides: Consensus of the European Organization for Research and Treatment of Cancer cutaneous lymphoma project group. J Am Acad Dermatol 47:364-370 (2002) https://doi.org/10.1067/mjd.2002.123482
  17. Trump JG, Wright KA, Evans WW, et al: High-energy electrons for the treatment of extensive superficial malignant lesions. Am J Ront Rad Ther Nuc Med 69:623-629 (1953)
  18. Johnston DO, Smedal MI, Wright KA, Trump JG: Electron beam therapy of widespread superficial malignant lesions. Surg Clin North Am 39:1-6 (1959)
  19. Wu JM, Leung SW, Wang CJ, Chui CS: Lying-on position of total skin electron therapy. Int J Radiation Oncology Biol Phys 39:521-528 (1997) https://doi.org/10.1016/S0360-3016(97)00141-7
  20. Cheung T, Butson MJ, Yu PKN: Energy dependence corrections to MOSFET dosimetric sensitivity. Australasian Physical&Engineering Science in Medicine 32:16-20 (2009) https://doi.org/10.1007/BF03178623
  21. Cho SK, Choi SH, Na SH, et al: Determination of dose correction factor for energy and directional dependence of the MOSFET dosimeter in an anthropomorphic phantom. J Korea Asso Radiat Prot 31:97-104 (2006)
  22. Fuss M, Sturtewagen E, De Wagter C, Georg D: Dosimetric characterization of GafChromic EBT film and its implication on film dosimetry quality assurance. Phys Med Biol 52:4211-4225 (2007) https://doi.org/10.1088/0031-9155/52/14/013
  23. Todorovic M, Fischer M, Cremers F, Thom E, Schmidt R: Evaluation of GafChromic EBT prototype B for external beam dose verification. Med Phys 33:1321-1328 (2006) https://doi.org/10.1118/1.2188077
  24. Cheung T, Butson MJ, Yu PK: Independence of calibration curves for EBT gafchromic films of the size of high-energy X-ray. Applied Radiation and Isotopes 64:1027-1030 (2006) https://doi.org/10.1016/j.apradiso.2006.04.006
  25. http://online1.ispsorp.com/_layouts/Gafchromic/content/products/ebt2/pdfs/GAFCHROMICEBT2TechnicalBrief-Rev1.pdf
  26. Su FC, Liu Y, Stathakis S, Shi C, Esquivel C, Papanikolaou N: Dosimetry characteristics of GAFCHROMIC EBT film responding to therapeutic electron beams. Appl Radiat Isot 65:1187-1192 (2007) https://doi.org/10.1016/j.apradiso.2007.05.005
  27. Butson MJ, Cheung T, Yu PK: Weak energy dependence of EBT gafchromic film dose response in the 50 kVp∼10 MVp X-ray range. Appl Radiat Isot 64:60-62 (2006) https://doi.org/10.1016/j.apradiso.2005.07.002
  28. Khan FM: The Physics of Radiation Therapy. 3rd ed. Williams & Wilkins, Philadelphia, PA (2003), pp. 301
  29. AAPM Report 23: Total Skin Electron Therapy: Technique and Dosimetry. New York, NY, Amerian Institue of Physics, Inc.,335 East 45 Street (1987)