DOI QR코드

DOI QR Code

Effects of Screw Configuration on Biomechanical Stability during Extra-articular Complex Fracture Fixation of the Distal Femur Treated with Locking Compression Plate

잠김 금속판(LCP-DF)을 이용한 대퇴골 원위부의 관절외 복합골절 치료시 나사못 배열에 따른 생체역학적 안정성 분석

  • Kwon, Gyeong-Je (Department of Biomedical Engineering, Inje University) ;
  • Jo, Myoung-Lae (Department of Biomedical Engineering, Inje University) ;
  • Oh, Jong-Keon (Department of Orthopedic Surgery, Korea University) ;
  • Lee, Sung-Jae (Department of Biomedical Engineering, Inje University)
  • Received : 2009.10.12
  • Accepted : 2010.04.28
  • Published : 2010.06.30

Abstract

The locking compression plates-distal femur(LCP-DF) are being widely used for surgical management of the extra-articular complex fractures of the distal femur. They feature locking mechanism between the screws and the screw holes of the plate to provide stronger fixation force with less number of screws than conventional compression bone plate. However, their biomechanical efficacies are not fully understood, especially regarding the number of the screws inserted and their optimal configurations. In this study, we investigated effects of various screw configurations in the shaft and the condylar regions of the femur in relation to structural stability of LCP-DF system. For this purpose, a baseline 3-D finite element (FE) model of the femur was constructed from CT-scan images of a normal healthy male and was validated. The extra-articular complex fracture of the distal femur was made with a 4-cm defect. Surgical reduction with LCP-DF and bone screws were added laterally. To simulate various cases of post-op screw configurations, screws were inserted in the shaft (3~5 screws) and the condylar (4~6 screws) regions. Particular attention was paid at the shaft region where screws were inserted either in clustered or evenly-spaced fashion. Tied-contact conditions were assigned at the bone screws-plate whereas general contact condition was assumed at the interfaces between LCP-DF and bone screws. Axial compressive load of 1,610N(2.3 BW) was applied on the femoral head to reflect joint reaction force. An average of 5% increase in stiffness was found with increase in screw numbers (from 4 to 6) in the condylar region, as compared to negligible increase (less than 1%) at the shaft regardless of the number of screws inserted or its distribution, whether clustered or evenly-spaced. At the condylar region, screw insertion at the holes near the fracture interface and posterior locations contributed greater increase in stiffness (9~13%) than any other locations. Our results suggested that the screw insertion at the condylar region can be more effective than at the shaft during surgical treatment of fracture of the distal femur with LCP-DF. In addition, screw insertion at the holes close to the fracture interface should be accompanied to ensure better fracture healing.

Keywords

References

  1. R.P. Mashru, E.A. Perez, "Fractures of the distal femur: current trends in evaluation and management," Curr. Opin. Orhop., vol. 18, pp.41-48. 2007. https://doi.org/10.1097/BCO.0b013e328011fbc0
  2. B.R. Bolhfner, B. Carmen. Clifford, "The results of open reduction and internal fixation of dstal femur fractures using a biologic(indirect) reduction technique," J. Orthop. Trauma, vol. 10, no. 6, pp.372-377, 1996. https://doi.org/10.1097/00005131-199608000-00002
  3. R.J. Placide, J.H. Lonner, "Fracture of the distal femur," Curr. Opin. Orthop., vol. 10, no. 1, pp.2-9, 1999. https://doi.org/10.1097/00001433-199902000-00002
  4. D.J. Shewring, B.F. Meggitt, "Fracture of the distal femur treated with the AO dynamic condylar screw," J. Bone Joint Surg. [Br], vol. 74, no. B, pp.122-125, 1992.
  5. R. Sanders, P. Regazzoni, T.P. Ruedi, "Treatment of supracondylar-intracondylar fractures of the femur using the dynamic condylar screw," J. Orthop. Trauma, no. 3. pp.214-222, 1989.
  6. M.J. Albert, "Supracondylar fractures of the femur," J. A.M. Acad. Orthop. Surg., vol. 5, pp.163-171, 1997. https://doi.org/10.5435/00124635-199705000-00005
  7. A. Marti, C. Fankhauser, A. Frenk et al., "Biomechanical evaluation of the less invasive stabiliation system for the in ternal fixation of distal femur fractures," J. Orthop. Trauma, vol. 15, no. 7, pp.482-487, 2001. https://doi.org/10.1097/00005131-200109000-00004
  8. K.J. Koval, J.J. Hoelhl, F.J. Kummer et al., "Distal femoral fixation : A biomechanical comparison of the standard condylar buttress plate, a locked buttress plate, and the 95-degree blade plate," J. Orthop. Trauma, vol. 11, no. 7, pp.521-524, 1997. https://doi.org/10.1097/00005131-199710000-00010
  9. R. Frigg, "Development of the Locking Compression Plate," injury, Int. J. Care Injured, vol. 32:S-B6-S-B10, 2003.
  10. M.J. Kaab, A. Frenk, A. Schmeling et al., "Locked internal fixator - Sensitivity of screw/plate stability to the correct insertion angle of the screw," J. Orthop. Trauma, vol. 18, pp.483-487, 2004. https://doi.org/10.1097/00005131-200409000-00002
  11. T. Goesling, A. Frenk, A. Appenzeller et al., "LISSPLT : Design, mechanical and biomechanical characteristics," Injury, Int. J. Care Injured, vol. 34:S-A11-S-A15, 2003.
  12. R. Frigg, "Locking compression Plate(LCP). An osteosynthesis plate based on the dynamic compression plate and the point contact fixator(PC-Fix)," Injury, Int. J. Care Injured, vol. 32:S-B-63-66, 2001.
  13. C.F. Luo, "Locking compression plating : a new solution for fractures in fheumatoid patients," Mod Rheumatol, vol. 15, pp.169-172, 2005. https://doi.org/10.1007/s10165-005-0392-z
  14. R. Frigg, "The development of the distal femur Less Invasive Stabilization System(LISS)," Injury, vol.32, no. 3, pp.48-54, 2001. https://doi.org/10.1016/S0020-1383(01)00183-8
  15. H.T. huang, P.J. Huang, J.Y. Su et al., "Indirect reduction and bridge plating of supracondylar fractures of the femur," Injury, vol. 32 no. 2, pp.135-140, 2003.
  16. A.D. Heiner, T.D. Brown, "Structural properties of a new design of composite replicate femurs and tibias," J. Biomechanics, vol. 32, pp.773-781, 2001.
  17. L. Cristofolini, M. Viceconti, "In vitro stress shielding measurements can be affected by large errors," The Journal of Strain Analysis, vol. 32, no. 3, pp. 193-200, 1997. https://doi.org/10.1243/0309324971513337
  18. R.D. McLeish, J. Charnley, "Abduction forces in the one legged stance," J. Biomechanics, vol. 3, pp.191-209, 1970. https://doi.org/10.1016/0021-9290(70)90006-0
  19. M.E. Taylor, K.E. Tanner KE, M.A.R. Freeman, et al., "Stress and strain distribution within the intact femur: compression or bending?," Med. Eng. Phys., vol. 18, no. 2, pp.122-131, 1996. https://doi.org/10.1016/1350-4533(95)00031-3
  20. "Standard specification and test method for metallic bone plates," ASTM F382-99.
  21. www.sawbone.com
  22. "Standard specification for wrought titanium-6 aluminium-7 niobium alloy tor surgical implant applications(UNS R56700)," ASTM F1295-05.
  23. G. Bergmann, F. Graichen, A. Rohlmann, "Hip joint loading during walking and running, measured in two patients," J. Biomechanics, vol. 26, no. 8. pp.969-990, 1993. https://doi.org/10.1016/0021-9290(93)90058-M
  24. N.W. Rydell, "Forces acting on the femoral head prosthesis." Acta. Orthop. Scand. vol. 37, no. 88, pp.1-132, 1966.
  25. J. Cordey, M. Borgeaud. S.M. Perren, "Force transfer between the plate and the bone: relative importance of the bending stiffness of the screw and the friction between plate and bone," Injury, Int. J. Care injured, vol. 31:S-C21-28, 2000.
  26. H. Tomkvist, T.C. Heam, J. Schatzker, "The strength of plate fixation in relation to the number and spacing of bone screws," J Orthop Trauma, vol. 10, pp.204-208. 1996. https://doi.org/10.1097/00005131-199604000-00009