The validation of Periotest values for the evaluation of orthodontic mini-implants' stability

즉시 부하 교정용 미니임플랜트의 안정성 평가를 위한 Periotest$^{(R)}$의 유효성

  • Cha, Jung-Yul (Department of Orthodontics, College of Dentistry, Yonsei University) ;
  • Yu, Hyung-Seog (Department of Orthodontics, College of Dentistry, Yonsei University) ;
  • Hwang, Chung-Ju (Department of Orthodontics, College of Dentistry, Yonsei University)
  • 차정열 (연세대학교 치과대학 교정학교실, 두개안면기형 연구소, 구강과학 연구소) ;
  • 유형석 (연세대학교 치과대학 교정학교실, 두개안면기형 연구소, 구강과학 연구소) ;
  • 황충주 (연세대학교 치과대학 교정학교실, 두개안면기형 연구소, 구강과학 연구소)
  • Received : 2008.12.12
  • Accepted : 2009.10.12
  • Published : 2010.06.30

Abstract

Objective: The aim of this study was to validate the Periotest values for the prediction of orthodontic mini-implants' stability. Methods: Sixty orthodontic mini-implants (7.0 mm $\times$ $\emptyset1.45$ mm; ACR, Biomaterials Korea, Seoul, Korea) were inserted into the buccal alveolar bone of 5 twelve month-old beagle dogs. Insertion torque (IT) and Periotest values (PTV) were measured at the installation procedure, and removal torque (RT) and PTV were recorded after 12 weeks of orthodontic loading. To correlate PTV with variables, the cortical bone thickness (mm) and bone mineral density (BMD) within the cortical bone and total bone area were calculated with the help of CT scanning. Results: The BMD and cortical bone thickness in mandibular alveolus were significantly higher than those of the maxilla (p < 0.05). The PTV values ranged from -3.2 to 4.8 for 12 weeks of loading showing clinically stable mini-implants. PTV at insertion was significantly correlated with IT (-0.51), bone density (-0.48), cortical bone thickness (-0.42) (p < 0.05) in the mandible, but showed no correlation in the maxilla. PTV before removal was significantly correlated with RT (-0.66) (p < 0.01) in the mandible. Conclusions: These results show that the periotest is a useful method for the evaluation of mini-implant stability, but it can only be applied to limited areas with thick cortical and high density bone such as the mandible.

Periotest$^{(R)}$는 보철용 임플랜트 식립 3 - 4개월 후 골유착 정도를 평가하기 위해 사용되는 동요도 측정 기기로 교정용 미니임플랜트 안정성 평가에 적용가능성이 높다. 본 연구는 교정용 미니임플랜트의 안정성 평가를 위해 Periotest$^{(R)}$의 유효성을 검증하고자 하였다. 성견 다섯 마리의 상악, 하악 협측골에 교정용 미니임플랜트를 식립하고, 식립 시 식립토오크와 동요도를 측정하고 12주 동안 교정력을 부여한 후, 제거 시 제거 토오크와 동요도를 측정하였다. 동요도 (Periotest$^{(R)}$ value, PTV) 측정은 재현성을 위해 동일 부위에 2회 측정하였다. PTV와 다른 변수와의 상관성 분석을 위해 모든 실험견은 CT 촬영 후 식립 부위의 골밀도와 피질골 두께를 계측하였다. 동요도 측정에 대한 재현성은 0.96의 상관계수를 나타내 매우 높았으며 (p < 0.001) 골밀도와 피질골 두께에 있어 하악골 협측치조골이 상악골에 비해 유의하게 컸다 (p < 0.05). 12주 동안 PTV 값은 -3.2에서 4.8의 범위를 나타내며 모든 임플랜트는 임상적으로 안정적이었다. 식립 시 PTV 값은 하악에서 식립 토오크(-0.51), 골밀도(-0.48), 피질골 두께(-0.42)와 상관성이 있었으나 (p < 0.05), 상악에서는 상관성이 없었다. 제거 시 PTV 값은 하악에서 제거 토오크(-0.66)와 상관성이 있었으며(p < 0.05), 상악에서는 상관성이 없었다. 이상의 결과는 Periotest$^{(R)}$를 이용하여 교정용 미니임플랜트의 안정성을 간접적으로 평가할 수 있으나 골밀도가 높고 피질골이 두꺼운 부위에서 제한적으로 응용할 수 있음을 시사한다.

Keywords

References

  1. Huja SS, Litsky AS, Beck FM, Johnson KA, Larsen PE. Pull-out strength of monocortical screws placed in the maxillae and mandibles of dogs. Am J Orthod Dentofacial Orthop 2005;127:307-13. https://doi.org/10.1016/j.ajodo.2003.12.023
  2. Huja SS, Katona TR, Burr DB, Garetto LP, Roberts WE. Microdamage adjacent to endosseous implants. Bone 1999;25:217-22. https://doi.org/10.1016/S8756-3282(99)00151-9
  3. Roberts WE, Simmons KE, Garetto LP, DeCastro RA. Bone physiology and metabolism in dental implantology: risk factors for osteoporosis and other metabolic bone diseases. Implant Dent 1992;1:11-21. https://doi.org/10.1097/00008505-199200110-00002
  4. Chen Y, Shin HI, Kyung HM. Biomechanical and histological comparison of self-drilling and self-tapping orthodontic microimplants in dogs. Am J Orthod Dentofacial Orthop 2008;133:44-50. https://doi.org/10.1016/j.ajodo.2007.01.023
  5. Brown GA, McCarthy T, Bourgeault CA, Callahan DJ. Mechanical performance of standard and cannulated 4.0-mm cancellous bone screws. J Orthop Res 2000;18:307-12. https://doi.org/10.1002/jor.1100180220
  6. Heidemann W, Gerlach KL, Grobel KH, Kollner HG. Influence of different pilot hole sizes on torque measurements and pullout analysis of osteosynthesis screws. J Craniomaxillofac Surg 1998;26:50-5. https://doi.org/10.1016/S1010-5182(98)80035-8
  7. Miyawaki S, Koyama I, Inoue M, Mishima K, Sugahara T, Takano-Yamamoto T. Factors associated with the stability of titanium screws placed in the posterior region for orthodontic anchorage. Am J Orthod Dentofacial Orthop 2003;124:373-8. https://doi.org/10.1016/S0889-5406(03)00565-1
  8. Schulte W, d'Hoedt B, Lukas D, Muhlbradt L, Scholz F, Bretschi J, et al. Periotest--a new measurement process for periodontal function. Zahnarztl Mitt 1983;73:1229-40.
  9. van Steenberghe D, Quirynen M. Reproducibility and detection threshold of peri-implant diagnostics. Adv Dent Res 1993;7:191-5.
  10. Friberg B, Sennerby L, Meredith N, Lekholm U. A comparison between cutting torque and resonance frequency measurements of maxillary implants. A 20-month clinical study. Int J Oral Maxillofac Surg 1999;28:297-303. https://doi.org/10.1016/S0901-5027(99)80163-5
  11. Alsaadi G, Quirynen M, Michiels K, Jacobs R, van Steenberghe D. A biomechanical assessment of the relation between the oral implant stability at insertion and subjective bone quality assessment. J Clin Periodontol 2007;34:359-66. https://doi.org/10.1111/j.1600-051X.2007.01047.x
  12. Schulte W, Lukas D. Periotest to monitor osseointegration and to check the occlusion in oral implantology. J Oral Implantol 1993;19:23-32.
  13. Schulte W, Lukas D. The Periotest method. Int Dent J 1992;42:433-40.
  14. Kim JW, Chang YI. Effects of drilling process in stability of micro-implants used for the orthodontic anchorage. Korean J Orthod 2002;32:107-15.
  15. Park SH, Kim SH, Ryu JH, Kang YG, Chung KR, Kook YA. Bone-implant contact and mobility of surface-treated orthodontic micro-implants in dogs. Korean J Orthod 2008;38:416-26. https://doi.org/10.4041/kjod.2008.38.6.416
  16. Meredith N, Alleyne D, Cawley P. Quantitative determination of the stability of the implant-tissue interface using resonance frequency analysis. Clin Oral Implants Res 1996;7:261-7. https://doi.org/10.1034/j.1600-0501.1996.070308.x
  17. Tricio J, Laohapand P, van Steenberghe D, Quirynen M, Naert I. Mechanical state assessment of the implant-bone continuum: a better understanding of the Periotest method. Int J Oral Maxillofac Implants 1995;10:43-9.
  18. Truhlar RS, Morris HF, Ochi S. Stability of the bone-implant complex. Results of longitudinal testing to 60 months with the periotest device on endosseous dental implants. Ann Periodontol 2000;5:42-55. https://doi.org/10.1902/annals.2000.5.1.42
  19. Truhlar RS, Lauciello F, Morris HF, Ochi S. The influence of bone quality on periotest values of endosseous dental implants at stage II surgery. J Oral Maxillofac Surg 1997;55(12 Suppl 5):55-61.
  20. Kim JW, Ahn SJ, Chang YI. Histomorphometric and mechanical analyses of the drill-free screw as orthodontic anchorage. Am J Orthod Dentofacial Orthop 2005;128:190-4. https://doi.org/10.1016/j.ajodo.2004.01.030
  21. Chavez H, Ortman LF, DeFranco RL, Medige J. Assessment of oral implant mobility. J Prosthet Dent 1993;70:421-6. https://doi.org/10.1016/0022-3913(93)90078-3
  22. Olive J, Aparicio C. Periotest method as a measure of osseointegrated oral implant stability. Int J Oral Maxillofac Implants 1990;5:390-400.
  23. Salonen MA, Raustia AM, Kainulainen V, Oikarinen KS. Factors related to periotest values in endosseal implants: a 9-year follow-up. J Clin Periodontol 1997;24:272-7. https://doi.org/10.1111/j.1600-051X.1997.tb01842.x
  24. Tricio J, van Steenberghe D, Rosenberg D, Duchateau L. Implant stability related to insertion torque force and bone density: an in vitro study. J Prosthet Dent 1995;74:608-12. https://doi.org/10.1016/S0022-3913(05)80313-0
  25. van Steenberghe D, Tricio J, Naert I, Nys M. Damping characteristics of bone-to-implant interfaces. A clinical study with the periotest device. Clin Oral Implants Res 1995;6:31-9. https://doi.org/10.1034/j.1600-0501.1995.060104.x
  26. Branemark R, Ohrnell LO, Nilsson P, Thomsen P. Biomechanical characterization of osseointegration during healing: an experimental in vivo study in the rat. Biomaterials 1997;18:969-78. https://doi.org/10.1016/S0142-9612(97)00018-5
  27. Deguchi T, Nasu M, Murakami K, Yabuuchi T, Kamioka H, Takano-Yamamoto T. Quantitative evaluation of cortical bone thickness with computed tomographic scanning for orthodontic implants. Am J Orthod Dentofacial Orthop 2006;129:721.e7-12.
  28. Carr AB, Papazoglou E, Larsen PE. The relationship of Periotest values, biomaterial, and torque to failure in adult baboons. Int J Prosthodont 1995;8:15-20.
  29. Aparicio C, Orozco P. Use of 5-mm-diameter implants: periotest values related to a clinical and radiographic evaluation. Clin Oral Implants Res 1998;9:398-406. https://doi.org/10.1034/j.1600-0501.1996.090605.x
  30. Haas R, Bernhart T, Dortbudak O, Mailath G. Experimental study of the damping behaviour of IMZ implants. J Oral Rehabil 1999;26:19-24. https://doi.org/10.1046/j.1365-2842.1999.00351.x
  31. Huang HM, Chiu CL, Yeh CY, Lee SY. Factors influencing the resonance frequency of dental implants. J Oral Maxillofac Surg 2003;61:1184-8. https://doi.org/10.1016/S0278-2391(03)00680-3
  32. Al-Nawas B, Wagner W, Grotz KA. Insertion torque and resonance frequency analysis of dental implant systems in an animal model with loaded implants. Int J Oral Maxillofac Implants 2006;21:726-32.
  33. Nkenke E, Hahn M, Weinzierl K, Radespiel-Troger M, Neukam FW, Engelke K. Implant stability and histomorphometry: a correlation study in human cadavers using stepped cylinder implants. Clin Oral Implants Res 2003;14:601-9. https://doi.org/10.1034/j.1600-0501.2003.00937.x
  34. Schliephake H, Sewing A, Aref A. Resonance frequency measurements of implant stability in the dog mandible: experimental comparison with histomorphometric data. Int J Oral Maxillofac Surg 2006;35:941-6. https://doi.org/10.1016/j.ijom.2006.05.002
  35. Moon CH, Lee DG, Lee HS, Im JS, Baek SH. Factors associated with the success rate of orthodontic miniscrews placed in the upper and lower posterior buccal region. Angle Orthod 2008;78:101-6. https://doi.org/10.2319/121706-515.1
  36. Park HS, Jeong SH, Kwon OW. Factors affecting the clinical success of screw implants used as orthodontic anchorage. Am J Orthod Dentofacial Orthop 2006;130:18-25. https://doi.org/10.1016/j.ajodo.2004.11.032

Cited by

  1. Stability and success rate of dual-thread miniscrews: vol.91, pp.4, 2010, https://doi.org/10.2319/083020-756.1