초록
Stroke patients, etc. can't use their hands because of the paralysis of their fingers, and their fingers could be recovered by rehabilitating training. In order to judge the rehabilitating extent of their fingers, the patients should grasp a cylindrical object in hospital. At present, the used object in hospital is only a cylindrical object, and can't measure the force of fingers. Therefore, doctors judge the rehabilitating extent by touching and watching at their fingers. So, the four-finger force measuring system which can measure the force of their fingers should be developed. In this paper, four-finger force measuring system with four force sensors which can measure the grasping force is developed. The force sensors are designed and fabricated, and the force measuring device is designed and manufactured by using DSP(digital signal processing). Also, the grasping force test of men is performed by using the developed four-finger force measuring system. It was confirmed that the finger average force of right hand is about 214.6 N and that of left hand is about 212.8 N.