DOI QR코드

DOI QR Code

정적형 유량계를 이용한 소닉노즐 유출계수 교정 방법에 관한 연구

Calibration of Discharge Coefficient of Sonic Nozzle Using CVFM

  • 신진현 (한국표준과학연구원 진공센터) ;
  • 강상백 (한국표준과학연구원 진공센터) ;
  • 박경암 (한국표준과학연구원 유동음향센터) ;
  • 임종연 (한국표준과학연구원 진공센터) ;
  • 정완섭 (한국표준과학연구원 유동음향센터)
  • Shin, J.H. (Vacuum center, Korea Research Institute of Standards and Science) ;
  • Kang, S.B. (Vacuum center, Korea Research Institute of Standards and Science) ;
  • Park, K.A. (Acoustics & Vibration group, Korea Research Institute of Standards and Science) ;
  • Lim, J.Y. (Vacuum center, Korea Research Institute of Standards and Science) ;
  • Cheung, W.S. (Acoustics & Vibration group, Korea Research Institute of Standards and Science)
  • 투고 : 2010.05.03
  • 심사 : 2010.07.05
  • 발행 : 2010.07.30

초록

진공용 기체 유동측정 표준기로 사용하고자 소닉노즐을 ISO 9300에서 제시한 사양에 맞추어 목직경 0.03 mm와 0.2 mm의 소닉 노즐을 제작하였다. 한국표준과학연구원에서 진공용 유량측정 장치로 개발된 정적형 유량계를 이용하여 제작된 2종의 소닉노즐의 유출계수를 확장불확도 3% 이내로 교정하였다. 교정된 소닉노즐의 유량 측정범위는 약 0.6~1,800 cc/min 범위를 갖는 것으로 나타났으며, 사용유동 조건에 해당되는 레이놀드 수(Reynolds number) 범위는 26~12,100으로 확인되었다. 이러한 결과는 교정된 소닉노즐을 이용하여 진공공정에서 필요한 극 미세 유량의 정밀측정을 가능하게 한 새로운 연구결과로 판단된다. 교정된 소닉노즐을 이용하여 진공펌프의 배기속도 측정결과는 기 구축된 정적법을 이용한 배기속도 측정결과와 1% 이내의 오차범위내로 매우 잘 일치함을 보였다. 교정된 소닉노즐은 향후 반도체 및 디스플레이 공정에 사용되는 다양한 건식 진공펌프들의 배기속도를 현장에서 간단하게 평가할 수 있는 현장 성능평가 장치에 활용할 예정이며, 현재 공정현장에서 배기속도 측정에 널리 사용중인 MFC를 대체할 수 있을 것으로 예상된다.

Sonic nozzles have been a standard device for measurement of steady state gas flow, as recommended in ISO 9300. This paper introduces two sonic nozzles of diameter ${\Phi}$ 0.03 mm and ${\Phi}$ 0.2 mm precisely machined according to ISO 9300. The constant volume flow meter(CVFM), readily set up in the Vacuum center of KRISS. was used to calibrate the discharge coefficients of both nozzles. The calibration results were shown to determine them within the 3% expanded measurement uncertainty. Calibrated sonic nozzles were found to be applicable for precision measurement of steady state gas flow in the vacuum process in the ranges of 0.6~1,800 cc/min. Those flow conditions are equivalent to the fine gas flow with Reynolds numbers of 26~12,100. Those encouraging results confirm that calibrated sonic nozzles enable precision measurement of extremely low gas flow encountered very often in th vacuum processes. Both calibrated sonic nozzles are proven to provide the precision measurement of the volume flow rate of the dry vacuum pump within one percent difference in reference to CVFM. Calibrated sonic nozzles are applied to a new 'in-situ and in-field' equipment designed to measure the volume flow rate of vacuum pumps in the semiconductor and flat display processes. Furthermore, they can provide other applications to flow control devices in vacuum, such as MFC, etc.

키워드

참고문헌

  1. Low Vacuum Generation and Control on BIEN Technology: Mass Flow and Dry Pumping Characteristics, Key Engineering Material, Vols. 277- 279, pp. 1000-1015 (2005). https://doi.org/10.4028/www.scientific.net/KEM.277-279.1000
  2. ISO 9300: Measurement of gas flow by means of critical flow venturi nozzles, (1990).
  3. 음속노즐의 임계 압력비에 대한 저 레이놀즈 수의 영향, 대한기계학회논문집, 24, 1535-1539 (2000).
  4. 배석희, 인상렬, 정광화, 이영백, 신용현, 진공공학 (한국경제신문, 서울, 2000).
  5. 한국표준과학연구원 교정 절차서, 상압가스 유량측정 표준시스템-벨 푸루버, C-10-4-0090-2002.