The Usefulness Evaluation of Radiation Shielding Devices in PET Scan Procedures

PET 검사 프러시저별 방사선 차폐기구의 유용성 평가

  • Kim, Yeong-Seon (Department of Nuclear Medicine, Dankook University Hospital) ;
  • Seo, Myeong-Deok (Department of Nuclear Medicine, Dankook University Hospital) ;
  • Lee, Wan-Kyu (Department of Nuclear Medicine, Dankook University Hospital) ;
  • Jeong, Yo-Cheon (Department of Radiology, Dankook University Hospital) ;
  • Kim, Sang-Wook (Seoul Radiology Service Co. ltd.) ;
  • Seo, Il-Teak (The Korean Society of Nuclear Medicine) ;
  • Song, Jae-Beom (Department of Nuclear Medicine, Dankook University Hospital)
  • 김영선 (단국대학교병원 핵의학과) ;
  • 서명덕 (단국대학교병원 핵의학과) ;
  • 이완규 (단국대학교병원 핵의학과) ;
  • 정요천 (단국대학교병원 영상의학과) ;
  • 김상욱 ((주)서울방사선서비스) ;
  • 서일택 (대한 핵의학회) ;
  • 송재범 (단국대학교병원 핵의학과)
  • Received : 2010.08.18
  • Accepted : 2010.08.31
  • Published : 2010.10.30

Abstract

Purpose: he use of PET scanners and the number of patient in Korea have been increased for recent several years dramatically. For this reason, technologists have more possibilities to be exposed to the radiation. The hospitals using PET scanners should make an effort to reduce the radiation exposure dose. The purpose of this study was to evaluate the radiation exposure does when using radiation shielding devices. The evaluation was performed through questionnaire survey and experiment. Materials and Methods: First, the technologists who had experience working in PET center in 2008-2009 were surveyed with questionnaire and TLD Figures, personal opinion of utilization of radiation shielding devices are analyzed. Second, we measured the shielding rate of shielding devices which have been using in PET study procedures. We divided the procedures into four steps; distribution, moving, injection of $^{18}F$-FDG and patient setup. Results: First, the results of this survey, using of L-block+Syringe shield, L-block, Syringe shield, No shield during the injection, were each 58.5%, 20%, 9%, 12.3%. The TLD values according to utilization of radiation shield, using both L-block+Syringe Shield and L-block showed the lower TLD values, and Syringe shield only or No shield showed the higher TLD values. Second, the results of experiments according to PET study procedures measured the shielding rates as follows. The shielding rates during the distribution using L-block, L-block+Apron shield were measured 97.4%, 97.7%. The shielding rates during the $^{18}F$-FDG delivery to the injection room using mobile Syringe shield, Syringe holder, Syringe shield carrier were each 81.7%, 98.9%, 99.7%. The shielding rates during the injection using Syringe shield, L-block, L-block+Syringe shield were measured each 51.9%, 98.3%, 98.7%. The shielding rates of Apron were measured in each 30, 60, 90, 120, 150 cm distance. The measurement were each 16.9%, 14.2%, 16.6%, 17.1%, 18.1%, 18.6%. Conclusion: The most effective method for radiation shielding is to using L-block during the $^{18}F$-FDG distribution and Syringe shield carrier during in moving $^{18}F$-FDG. For the $^{18}F$-FDG injection, L-block+Syringe shield have to be used. The shielding effect of Apron has shown average 16.4%. According to the survey of questionnaire, the operators recognized well risk of the radiation exposure but, tended ignore in working. The radiation dose according to recognition of radiation exposure risk was not relevant. but radiation dose according to utilization of radiation shield lower the more use it. The main reason of no use of shielding devices is cumbersome, 55% of the respondents answered. I'm sure, by use of radiation shield in all PET procedure, radiation exposure will be reduced considerably.

1994년 PET의 국내 도입 이후, 현재까지 양적으로나 질적으로 많은 발전이 있었다. 하지만 이와 함께 방사성의약품의 사용량 또한 급증하면서 검사에 있어서 가장 중요한 역할을 하는 방사선사의 개인피폭선량이 높아지는 요인으로 작용했던 것이 사실이다. 식품의약품안전청의 피폭선량관리센터에서 발표한 자료에 따르면 2008년도 방사선작업종사자의 전체 평균 피폭선량은 0.67 mSv였고, 방사선사 전체 피폭선량은 1.33 mSv로 해마다 감소하는 것으로 나타났다. 설문결과를 살펴보면 PET 검사 담당자의 평균피폭선량은 1.69 mSv였고, 1.0 mSv를 초과하는 구간에 75.3%의 방사선사가 포함되어 있었다. 이는 고 노출구간에 속한 방사선사의 비율이 높다는 것을 의미하므로, 피폭저감을 위한 노력이 요구된다. 본 연구에서는 설문조사를 통하여 PET 검사 프러시저별 차폐기구의 활용에 따른 방사선 피폭 정도를 TLD 수치를 이용하여 분석하고, 실험을 통하여 각 차폐기구의 차폐율을 측정함으로써, 그 효용성을 입증하고 적합한 차폐방법을 강구하고자 하였다. 방사선 피폭의 위험에 대해서 어느 누구보다 잘 인지하고 있는 방사선사임에도 불구하고 안이하게 생각하는 경향이 있었다. 설문결과를 살펴보면 방사선 피폭의 위험성에 대한 인식도와 피폭선량과는 크게 관련이 없었으며, 차폐기구의 활용도에 따라서는 피폭선량의 차이가 확연하게 나타났다. 프러시저 중 피폭선량에 가장 많은 영향을 미치는 단계는 방사성의약품의 투여였으며, 투여시 차폐방법에 따라 이동시 차폐방법도 달라졌다. 투여 시 차폐기구 사용현황을 보면 Both shield는 58.5%, L-Block은 20%, Syringe shield는 9%, No shield는 12.3%를 차지하였다. TLD 수치에 따른 투여 시 차폐방법은 수치가 낮을수록 Both shield와 L-block을 많이 사용하였으며, 수치가 높을수록 Syringe shield와 차폐를 시행하지 않는 경우가 많았다. 실험결과, 가장 이상적인 차폐방법은 분배 시엔 L-block을 사용하고, 이동시엔 Syringe shield carrier를 사용하며, 투여 시엔 L-block과 Syringe shield를 함께 사용하는 것이다. 환자 포지셔닝 시 Apron의 차폐율은 평균 16.4%로 차폐효과가 있다. PET 검사를 시행함에 있어서 강한 실천의지를 가지고 프러시저별로 차폐기구를 적극 활용한다면 점증하는 방사선사의 피폭 수준을 크게 감소시킬 수 있을 것으로 판단된다.

Keywords