The Effect of Using Two Different Type of Dose Calibrators on In Vivo Standard Uptake Value of FDG PET

FDG 사용 시 Dose Calibrator에 따른 SUV에 미치는 영향

  • Park, Young-Jae (Department of Nuclear Medicine, Seoul National University Bundang Hospital) ;
  • Bang, Seong-Ae (Department of Nuclear Medicine, Seoul National University Bundang Hospital) ;
  • Lee, Seung-Min (Department of Nuclear Medicine, Seoul National University Bundang Hospital) ;
  • Kim, Sang-Un (Department of Nuclear Medicine, Seoul National University Bundang Hospital) ;
  • Ko, Gil-Man (Department of Nuclear Medicine, Seoul National University Bundang Hospital) ;
  • Lee, Kyung-Jae (Department of Nuclear Medicine, Seoul National University Bundang Hospital) ;
  • Lee, In-Won (Department of Nuclear Medicine, Seoul National University Bundang Hospital)
  • 박영재 (분당서울대학교병원 핵의학과) ;
  • 방성애 (분당서울대학교병원 핵의학과) ;
  • 이승민 (분당서울대학교병원 핵의학과) ;
  • 김상언 (분당서울대학교병원 핵의학과) ;
  • 고길만 (분당서울대학교병원 핵의학과) ;
  • 이경재 (분당서울대학교병원 핵의학과) ;
  • 이인원 (분당서울대학교병원 핵의학과)
  • Received : 2010.03.23
  • Accepted : 2010.03.31
  • Published : 2010.06.05

Abstract

Purpose: The purpose of this study is to measure F-18 FDG with two different types of dose calibrator measuring radionuclide and radioactivity and investigate the effect of F-18 FDG on SUV (Standard Uptake Value) in human body. Materials and Methods: Two different dose calibrators used in this study are CRC-15 Dual PET (Capintec) and CRC-15R (Capintec). Inject 1 mL, 2 mL, 3 mL of F-18 FDG into three 2 mL syringes, respectively, and measure initial radioactivity from each dose calibrator. Then measure and record radioactivity at 30 minute interval for 270 minutes. According to the initial radioactivity, linearity between decay factor driven from radioactive decay formula and the values measured by dose calibrator have been analyzed by simple linear regression. Fine linear regression line optimizing values measured with CRC-15 through regression analysis on the basis of the volume of which the measured value is close to the most ideal one in CRC-15 Dual PET. Create ROI on lung, liver, and region part of 50 persons who has taken PET/CT test, applying values from linear regression equation, and find SUV. We have also performed paired t-test to examine statistically significant difference in the radioactivity measured with CRC-15 Dual PET, CRC-15R and its SUV. Results: Regression analysis of radioactivity measured with CRC-15 Dual PET and CRC-15R shows results as follows: in the case 1 mL, the r statistic representing correlation was 0.9999 and linear regression equation was y=1.0345x+0.2601; in 2 mL case, r=0.9999, linear regression equation y=1.0226x+0.1669; in 3 mL case, r=0.9999, linear regression equation y=1.0094x+0.1577. Based on the linear regression equation from each volume, t-test results show significant difference in SUV of ROI in lung, liver, region part in all three case. P-values in each case are as follows: in 1 mL case, lung, liver and region (p<0.0001); in 2 mL case, lung (p<0.002), liver and region (p<0.0001); in 3 mL case, lung (p<0.044), liver and region (p<0.0001). Conclusion: Radioactivity measured with CRC-15 Dual PET, CRC-15R, dose calibrator for F-18 FDG test, do not show difference correlation, while these values infer that SUV has significant differences in the aspect of uptake in human body. Therefore, it is necessary to consider the difference of SUV in human body when using these dose calibrator.

방사성핵종과 방사능을 측정하는 dose calibrator의 서로 다른 두개의 기기에서 F-18 FDG을 각각 측정하고, 이 측정값에 의한 인체 내 SUV (Standard Uptake Value)에 미치는 영향은 없는지에 대하여 알아보고자 하였다. 이 연구에서 두개의 다른 dose calibrator는 CRC-15 Dual PET, CRC-15R을 사용하였다. 각 dose calibrator에서 F-18 FDG를 2 mL 주사기 3개에 볼륨 1 mL, 2 mL, 3 mL를 만들어 각각 초기 radio activity를 측정한 후 4시간 30분(270분)까지 30분 간격으로 radio activity를 각각 측정하고 기록한다. 초기 radio activity 값을 기준으로 방사선붕괴 공식으로 산출된 값(decay factor)과 dose calibrator로 측정된 값 간의 직선성을 단순선형을 통하여 분석하였다. CRC-15 Dual PET에서 가장 이상적인 값에 가까운 측정값을 볼륨을 기준으로 하여, CRC-15R에서 측정된 값을 최적화하기 위한 선형회귀식을 회귀분석을 통하여 구한다. 선형회귀식의 산출값을 적용하여 PET/CT 검사를 시행한 50명을 대상으로 lung, liver, region 부위에 ROI을 그려 SUV를 구한다. CRC-15 Dual PET, CRC-15R에서 측정한 radio activity와 그 값을 이용한 SUV를 paired t-test로 통계적으로 유의한 차이가 있는지 알아보았다. CRC-15 Dual PET과 CRC-15R에서 측정한 radio activity의 상관분석결과 1 mL의 경우 상관성을 보여주는 r값은 r=0.999이었으며, 회귀분석에 위한 선형식은 y=1.0345x+ 0.2601이었다. 2 mL의 경우 r=0.999이며, 선형회귀식은 y=1.0226x+0.1669이었다. 3 mL의 경우 r=0.999이며, 선형회귀식은 y=1.0094x+0.1577이었다. 각 세 가지 볼륨에서 얻어진 선형회귀식의 산출 값을 이용하여 구한 lung, liver, region 부위에 ROI의 SUV에서의 차이는 1 mL의 경우, lung, liver, region은 (p<0.0001)로 t-test 결과 모두 유의한 차이가 있었다. 2mL의 경우, lung (p<0.002), liver, region은 ROI의 SUV에서 유의한 차이(p<0.0001)를 보였다. 또한 3 mL의 경우, lung(p<0.044), liver, region의 유의한 차이(p<0.0001)가 있었다. F-18 FDG 검사에서 사용되는 두 개의 dose calibrator CRC-15 Dual PET, CRC-15R의 radioactivity에 대한 측정치에서는 상관관계의 차이가 없다는 것을 알 수 있었다. 그러나 이 두 값을 통해 SUV는 인체 내의 uptake 정도에 있어서는 유의한 차이를 보인다는 것을 알 수 있었다. 따라서 이 두개의 dose calibrator를 사용함에 있어서, 이 두 값이 인체 내 SUV의 차이를 고려하여 사용하여야 할 필요성이 요구된다.