초록

We have developed a high-$T_c$ SQUID magnetocardiogram (MCG) system for small laboratory animals. White noise of the measurement system was about 30 fT/$Hz^{1/2}$ when measured in a magnetically shielded room. We optimized the measurement position to obtain clear MCG wave from rat's small heart by using grid measurements. With the optimization, the MCG signal was successfully detected with the peak amplitude of about 30 pT. We could observe well defined P-, QRS-, and T-waves from the rat MCG. The results suggest that the developed system has a strong potential to monitor the progress of the heart disease model by using a laboratory rat.

키워드

참고문헌

  1. V. Pizzella, S. D. Penna, C. D. Gratta, and G. L. Romani, "SQUID systems for biomagnetic imaging", Supercond. Sci. Technol., vol. 14, pp. R79-R114, Apr. 2001. https://doi.org/10.1088/0953-2048/14/7/201
  2. I. Tavarozzi, S. Comani, C. D. Gratta, G. L. Romani, S. D. Luzio, D. Brisinda, S. Gallina, M. Zimarino, R. Fenici, R. D. Caterina", Magnetocardiography: current status and perspectives. Part I: Physical principles and instrumentation", Ital. Heart J. vol. 3, pp. 75-85, 2002.
  3. I. Tavarozzi, S. Comani, C. D. Gratta, S. D. Luzio, G. L. Romani, S. Gallina, M. Zimarino, D. Brisinda, R. Fenici, R. D. Caterina, "Magnetocardiography: current status and perspectives. Part II: Clinical applications", Ital. Heart J. vol. 3, pp. 151-165, 2002.
  4. S. Yamada, I. Yamaguchi, "Magnetocardiograms in clinical medicine: Unique information on cardiac ischemia, arrhythmias, and fetal diagnosis", Internal Medicine, vol. 44, pp. 1-19, 2005. https://doi.org/10.2169/internalmedicine.44.1
  5. K. Sternickel, A. I. Braginski, "Biomagnetism using SQUIDs: status and perspective", Supercon. Sci. Technol., vol. 19, pp. S160-S171, 2006. https://doi.org/10.1088/0953-2048/19/3/024
  6. D. Brisinda, M. E. Caristo, and Riccardo Fenici, "Contactless magnetocardiographic mapping in anesthetized Wistar rats: evidence of age-related changes of cardiac electrical activity", Am J Physiol Heart Circ Physiol 291: H368-H378, 2006. https://doi.org/10.1152/ajpheart.01048.2005
  7. R. Fischer, A. Gapelyuk, N. Wessel, K. Gruner, A. Gruner, D. Muller, R. Dietz, A. Schirdewan, "Time course of changes in cardiac magnetic field mapping after myocardial infarction in rats", International Congress Series 1300 pp. 484-487, 2007.
  8. Y. Aita, S. Kambe, T. Nagaishi, H. Itozaki, "Magnetocardiogram of a rat observed by a high-Tc SQUID sensor system", Physica C 378-381, 2002.
  9. S. Kuriki, A. Hayashi, Y. Hirata, T. Naiki, and K. Kawahara, "High-Tc SQUID Magnetometer for Measurements of Heart-Rate Signal of Small Animals", Chinese J. Physics vol. 42, pp.501-507, 2004.