Analysis of Space Radiation Dose Rate using portable X-ray Generating Device for Abdomen

이동형 X-ray 발생장치를 이용한 복부 촬영 시 공간 선량률에 관한 연구

  • Received : 2010.08.11
  • Accepted : 2010.10.28
  • Published : 2010.11.30

Abstract

This experimental study is carried out one of the General Hospital in Kyungbok providence. Abdomen Phantom being located Anterior-posterior(AP) position on portable bed, and the portable X-ray generating device was placed the phantom at $-90^{\circ}$ direction. The experiment were set 65 kVp, 10 mAs, $10{\times}10\;cm^2$, 100 cm(FOD) for the measurement. Digital proportional counting tube survey meter was used for measuring the space scatter dose. Measurement points of horizontal distribution was set up at $30^{\circ}$ interval by increasing 50 cm radius of upside, downside, left and right. Vertical distribution of measurement points were set up for the vertical plane with a radius of at $30^{\circ}$ intervals with 50cm increments. It is concluded that longer distance from the soure of X-ray significantly decrease radiation dose to the patient and use of the radiation protection device should be applied in clinical practice to reduce dose to the patient.