Biomechanical Changes of the Lumbar Segment after Total Disc Replacement : Charite$^{(R)}$, Prodisc$^{(R)}$ and Maverick$^{(R)}$ Using Finite Element Model Study

  • Kim, Ki-Tack (Department of Orthopedic Surgery, School of Medicine, Kyung Hee University) ;
  • Lee, Sang-Hun (Department of Orthopedic Surgery, School of Medicine, Kyung Hee University) ;
  • Suk, Kyung-Soo (Department of Orthopedic Surgery, School of Medicine, Kyung Hee University) ;
  • Lee, Jung-Hee (Department of Orthopedic Surgery, School of Medicine, Kyung Hee University) ;
  • Jeong, Bi-O (Department of Orthopedic Surgery, School of Medicine, Kyung Hee University)
  • Received : 2009.10.30
  • Accepted : 2010.05.17
  • Published : 2010.06.28

Abstract

Objective : The purpose of this study was to analyze the biomechanical effects of three different constrained types of an artificial disc on the implanted and adjacent segments in the lumbar spine using a finite element model (FEM). Methods : The created intact model was validated by comparing the flexion-extension response without pre-load with the corresponding results obtained from the published experimental studies. The validated intact lumbar model was tested after implantation of three artificial discs at L4-5. Each implanted model was subjected to a combination of 400 N follower load and 5 Nm of flexion/extension moments. ABAQUS$^{TM}$ version 6.5 (ABAQUS Inc., Providence, RI, USA) and FEMAP version 8.20 (Electronic Data Systems Corp., Plano, TX, USA) were used for meshing and analysis of geometry of the intact and implanted models. Results : Under the flexion load, the intersegmental rotation angles of all the implanted models were similar to that of the intact model, but under the extension load, the values were greater than that of the intact model. The facet contact loads of three implanted models were greater than the loads observed with the intact model. Conclusion : Under the flexion load, three types of the implanted model at the L4-5 level showed the intersegmental rotation angle similar to the one measured with the intact model. Under the extension load, all of the artificial disc implanted models demonstrated an increased extension rotational angle at the operated level (L4-5), resulting in an increase under the facet contact load when compared with the adjacent segments. The increased facet load may lead to facet degeneration.

Keywords

References

  1. Ahn MW, Shin KS, Lee JW, Chung IS, Lee CY : Analysis of compression behavior in lumbar spine under simple vertical load vs follower load. J Korean Soc Spine Surg 9 : 78-83, 2002 https://doi.org/10.4184/jkss.2002.9.2.78
  2. Baker DA, Hastings RS, Pruitt L : Study of fatigue resistance of chemical and radiation cross linked medical grade ultrahigh molecular weight polyethylene. J Biomed Mater Res 46 : 573-581, 1999 https://doi.org/10.1002/(SICI)1097-4636(19990915)46:4<573::AID-JBM16>3.0.CO;2-A
  3. Freeman BJ, Davenport J : Total disc replacement in the lumbar spine : a systematic review of the literature. Eur Spine J 15 : S439-S447, 2006 https://doi.org/10.1007/s00586-006-0186-9
  4. Chen CS, Feng CK, Cheng CK, Tzeng MJ, Liu CL, Chen WJ : Biomechanical analysis of the disc adjacent to posterolateral fusion with laminectomy in lumbar spine. J Spinal Disorders Tech 18 : 58-65, 2005 https://doi.org/10.1097/01.bsd.0000123426.12852.e7
  5. Chen WJ, Lai PL, Niu CC, Chen LH, Fu TS, Wong CB : Surgical treatment of adjacent instability after lumbar spine fusion. Spine 26 : E519-E524, 2001 https://doi.org/10.1097/00007632-200111150-00024
  6. Chung SK, Kim YE, Wang KC : Biomechanical effect of constraint in lumbar total disc replacement : a study with finite element analysis. Spine (Phila Pa 1976) 34 :1281-1286, 2009
  7. Cunningham BW, Gordon JD, Dmitriev AE, Hu N McAfee PC : Biomechanical evaluation of total disc replacement arthroplasty : an in vitro human cadaveric model. Spine (Phila Pa 1976) 28 : S110-S117, 2003 https://doi.org/10.1097/01.BRS.0000092209.27573.90
  8. David T : Revision of a Charite artificial disc 9.5 years in vivo to a new Charite artificial disc : case report and explant analysis. Eur Spine J 14 : 507-511, 2005 https://doi.org/10.1007/s00586-004-0842-x
  9. de Kleuver M, Oner FC, Jacobs WC : Total disc replacement for chronic low back pain : background and a systematic review of the literature. Eur Spine J 12 : 108-116, 2003
  10. Dooris AP, Goel VK, Grosland NM, Gilbertson LG, Wilder DG : Load-sharing between anterior and posterior elements in a lumbar motion segment implanted with an artificial disc. Spine (Phila Pa 1976) 26 : E122-E129, 2001 https://doi.org/10.1097/00007632-200103150-00004
  11. Geisler FH, Blumenthal SL, Guyer RD, McAfee PC, Regan JJ, Johnson JP, et al. : Neurological complications of lumbar artificial disc replacement and comparison of clinical results with those related to lumbar arthrodesis in the literature : results of a multicenter, prospective, randomized investigational device exemption study of Charite intervertebral disc. Invited submission from the Joint Section Meeting on Disorders of the Spine and Peripheral Nerves, March 2004. J Neurosurg Spine 1 : 143-154, 2004 https://doi.org/10.3171/spi.2004.1.2.0143
  12. Goto K, Tajima N, Chosa E, Totoribe K, Kubo S, Kuroki H, et al. : Effects of lumbar spinal fusion on the other lumbar intervertebral levels (three-dimensional finite element analysis). J Orthop Sci 8 : 577-584, 2003 https://doi.org/10.1007/s00776-003-0675-1
  13. Ha SK : Finite element modeling of multi-level cervical spinal segments (C3-C6) and biomechanical analysis of an elastomer-type prosthetic disc. Med Eng Phys 28 : 534-541, 2006 https://doi.org/10.1016/j.medengphy.2005.09.006
  14. Hitchon PW, Eichholz K, Barry C, Rubenbauer P, Ingalhalikar A, Nakamura S, et al. : Biomechanical studies of an artificial disc implant in the human cadaveric spine. J Neurosurg Spine 2 : 339-343, 2005 https://doi.org/10.3171/spi.2005.2.3.0339
  15. Kim YE, Yun SS, Jung SK : Biomechanical analysis of the implanted constrained and unconstrained ICR types of artificial disc using FE model. J of the KSPE 23 : 176-182, 2006
  16. Kumaresan S, Yoganandan N, Pintar FA, Maiman DJ : Finite element modeling of the cervical spine : role of intervertebral disc under axial and eccentric loads. Med Eng Physics 21 : 689-700, 1999 https://doi.org/10.1016/S1350-4533(00)00002-3
  17. Le Huec JC, Basso Y, Aunoble S, Friesem T, Bruno MB : Influence of facet and posterior muscle degeneration on clinical results of lumbar total disc replacement : two-year follow-up. J Spinal Disord Tech 18 : 219-223, 2005
  18. McAfee PC, Cunningham B, Holsapple G, Adams K, Blumenthal S, Guyer RD, et al. : A prospective, randomized, multicenter Food and Drug Administration investigational device exemption study of lumbar total disc replacement with the CHARITE artificial disc versus lumbar fusion : part II : evaluation of radiographic outcomes and correlation of surgical technique accuracy with clinical outcomes. Spine (Phila Pa 1976) 30 : 1576-1583; discussion E388-E390, 2005 https://doi.org/10.1097/01.brs.0000170561.25636.1c
  19. Natarajan RN, Andersson GB : The influence of lumbar disc height and cross-sectional area on the mechanical response of the disc to physiologic loading. Spine (Phila Pa 1976) 24 : 1873-1881, 1999 https://doi.org/10.1097/00007632-199909150-00003
  20. Ng HW, Teo EC, Lee VS : Statistical factorial analysis on the material property sensitivity of the mechanical responses of the C4-C6 under compression, anterior and posterior shear. J Biomech 37 : 771-777, 2004 https://doi.org/10.1016/j.jbiomech.2003.09.025
  21. Patwardhan AG, Havey RM, Meade KP, Lee B, Dunlap B : A follower load increases the load-carrying capacity of the lumbar spine in compression. Spine (Phila Pa 1976) 24 : 1003-1009, 1999 https://doi.org/10.1097/00007632-199905150-00014
  22. Rohlmann A, Neller S, Claes L, Bergmann G, Wilke HJ : Influence of a follower load on intradiscal pressure and intersegmental rotation of the lumbar spine. Spine (Phila Pa 1976) 26 : E557-E561, 2001 https://doi.org/10.1097/00007632-200112150-00014
  23. Rohlmann A, Zander T, Bergmann G : Effect of total disc replacement with prodisc on intersegmental rotation of the lumbar spine. Spine (Phila Pa 1976) 30 : 738-743, 2005 https://doi.org/10.1097/01.brs.0000157413.72276.c4
  24. van Ooij A, Oner FC, Verbout AJ : Complications of artificial disc replacement : a report of 27 patients with the SB Charité disc. J Spinal Disord Tech 16 : 369-383, 2003 https://doi.org/10.1097/00024720-200308000-00009
  25. Wang JL, Parnianpour M, Shirazi-Adl A, Engin AE : Rate effect on sharing of passive lumbar motion segment under load-controlled sagittal flexion : viscoelastic finite element analysis. Theoretical and Applied Fracture Mechanics 32 : 119-128, 1999 https://doi.org/10.1016/S0167-8442(99)00032-4
  26. White AA III, Panjabi MM : Clinical biomechanics of the spine. Philadelphia : Lippincott, 1978, pp81
  27. Yamamoto I, Panjabi MM, Crisco T, Oxland T : Three-dimensional movements of the whole lumbar spine and lumbosacral joint. Spine (Phila Pa 1976) 14 : 1256-1260, 1989 https://doi.org/10.1097/00007632-198911000-00020

Cited by

  1. Effect of centers of rotation on spinal loads and muscle forces in total disk replacement of lumbar spine vol.227, pp.5, 2010, https://doi.org/10.1177/0954411912474742
  2. The Effects of a Semiconstrained Integrated Artificial Disc on Zygapophyseal Joint Pressure and Displacement vol.39, pp.25, 2010, https://doi.org/10.1097/brs.0000000000000638
  3. Analysis of risk factors for adjacent superior vertebral pedicle-induced facet joint violation during the minimally invasive surgery transforaminal lumbar interbody fusion: a retrospective study vol.20, pp.20, 2010, https://doi.org/10.1186/s40001-015-0174-9
  4. ISASS Policy Statement – Lumbar Artificial Disc vol.9, pp.None, 2015, https://doi.org/10.14444/2007
  5. The Change of Sagittal Alignment of the Lumbar Spine after Dynesys Stabilization and Proposal of a Refinement vol.58, pp.1, 2010, https://doi.org/10.3340/jkns.2015.58.1.43
  6. Biomechanical Effects of the Geometry of Ball-and-Socket Artificial Disc on Lumbar Spine: A Finite Element Study vol.42, pp.6, 2010, https://doi.org/10.1097/brs.0000000000001789
  7. Effect of sacral slope on the biomechanical behavior of the low lumbar spine vol.13, pp.5, 2010, https://doi.org/10.3892/etm.2017.4251
  8. Biomechanical analysis of lumbar interbody fusion cages with various lordotic angles: a finite element study vol.21, pp.3, 2018, https://doi.org/10.1080/10255842.2018.1442443
  9. Biomechanical evaluation of four surgical scenarios of lumbar fusion with hyperlordotic interbody cage: A finite element study vol.29, pp.4, 2018, https://doi.org/10.3233/bme-181004
  10. Development of a Biconcave Mobile‐Bearing Lumbar Total Disc Arthroplasty Concept Using Finite Element Analysis and Design Optimization vol.37, pp.8, 2019, https://doi.org/10.1002/jor.24315
  11. Finite element analysis of a ball‐and‐socket artificial disc design to suppress excessive loading on facet joints: A comparative study with ProDisc vol.35, pp.9, 2010, https://doi.org/10.1002/cnm.3214
  12. Growing a backbone - functional biomaterials and structures for intervertebral disc (IVD) repair and regeneration: challenges, innovations, and future directions vol.8, pp.5, 2020, https://doi.org/10.1039/c9bm01288e
  13. Sensitivities of lumbar segmental kinematics and functional tissue loads in sagittal bending to design parameters of a ball-in-socket total disc arthroplasty prosthesis vol.23, pp.9, 2010, https://doi.org/10.1080/10255842.2020.1745783