DOI QR코드

DOI QR Code

Analysis of heavy metal in polymer materials by combustion ICP-AES Method

연소법 전처리에 의한 고분자 물질 중 중금속의 분석

  • Lim, Heon-Sung (Chemical Analysis Laboratory, Korea Research Institute of Chemical Technology) ;
  • Lee, Bo-Reum (Chemical Analysis Laboratory, Korea Research Institute of Chemical Technology) ;
  • Lee, Sueg-Geun (Chemical Analysis Laboratory, Korea Research Institute of Chemical Technology)
  • 임헌성 (한국화학연구원 화학분석센터) ;
  • 이보름 (한국화학연구원 화학분석센터) ;
  • 이석근 (한국화학연구원 화학분석센터)
  • Received : 2010.08.19
  • Accepted : 2010.12.02
  • Published : 2010.12.25

Abstract

An oxygen bomb combustion procedure were studied for determination of heavy metal in polymer materials such as polyvinyl chloride and polypropylene by ICP-AES. This method is proposed as a rapid and simple sample preparation for decomposition of polymer containing certified contents of the analytes by teflon coated oxygen bomb combustion using different absorbing solution. The recoveries of metal were found to be 30~102% from absorbed solution by ICP-AES method using polypropylene certified reference material(PP CRM-As, Ba, Cd, Hg, Pb, Zn). The recoveries of metal using PVC CRM(Cd, Hg, Pb) was found to be 45 ~101% with same procedures.

Keywords

References

  1. K. E. Haque, Int. J. Miner. Process., 57, 1-24(1999). https://doi.org/10.1016/S0301-7516(99)00009-5
  2. E. M. Flores, J S. Barin, M. F. Mesko and G. Knapp, Spectrochim. Acta B, 62, 1051-1064(2007). https://doi.org/10.1016/j.sab.2007.04.018
  3. J. Jung, H. Kim, Y. Lee, L. Lee, J. Shin and S. Lee, Anal. Sci. Tech., 21(4), 284-295(2008).
  4. Y. Miyake, N. Yamashita, P. Rostkowski, M. K. So, S. Taniyasu, P. K. Lam and K. Kannan, J. Chromatogra. A, 1154, 214-221(2007). https://doi.org/10.1016/j.chroma.2007.03.084
  5. Y. Miyake, M. Kato and K. Urano, J. Chromatogra. A, 1139, 63-69(2007). https://doi.org/10.1016/j.chroma.2006.10.078
  6. Y. S. Fung and K. L. Dao, Anal. chim. Acta, 315, 347-355(1995). https://doi.org/10.1016/0003-2670(95)00317-S
  7. Y. Miyake, N. Yamashita, P. Rostkowski, M. K. So, S. Taniyasu, P. K. Lam and K. Kannan, J. Chromatogra. A, 1143, 98-104(2007). https://doi.org/10.1016/j.chroma.2006.12.071

Cited by

  1. Evaluation of Accuracy and Precision of Analysis of Metals with Polyvinyl Chloride Membrane Filters vol.26, pp.1, 2016, https://doi.org/10.15269/JKSOEH.2016.26.1.48