DOI QR코드

DOI QR Code

Blood Pressure Measurement using the Modified Step-wise Deflation Method

개선된 계단 배기 방법을 이용한 혈압 측정

  • Oh, Hong-Sic (Department of Biomedical Engineering, Hanyang University) ;
  • Lee, Jong-Shill (Department of Biomedical Engineering, Hanyang University) ;
  • Chee, Young-Joon (Department of Biomedical Engineering, College of Engineering, University of Ulsan) ;
  • Kim, In-Young (Department of Biomedical Engineering, Hanyang University)
  • 오홍식 (한양대학교 의용생체공학과) ;
  • 이종실 (한양대학교 의용생체공학과) ;
  • 지영준 (울산대학교 공과대학 의공학과) ;
  • 김인영 (한양대학교 의용생체공학과)
  • Received : 2010.03.10
  • Accepted : 2010.09.20
  • Published : 2010.10.31

Abstract

In the automatic non-invasive blood pressure measurement device, the oscillometric method iswidely used. In the oscillometric method, the step-wise deflation has the advantage of the robustness for the motion artifacts than the linear deflation method. But it has the disadvantage of its longer measurement time because we need to detect two or more pulses in a certain cuff pressure step. In this study, we suggest the modified step-wise deflation method to overcome this limitation while maintaining the general concept of step-wise deflation. Using one valid pulse in each step and the deflating valve control during the diastolic period, the measurement time could be reduced. In order to verify the accuracy of the proposed algorithm, we compared the blood pressure values from the suggested method and the blood pressure values from the conventional auscultation method. The mean and standard deviation were -0.50${\pm}$5.3mmHg and 2.08${\pm}$4.75mmHg, for systolic and diastolic blood pressure respectively. The measurement time can be reduced up to the half of conventional step-wise deflation method.

Keywords

References

  1. M.Ramsey "Noninvasive automatic determination of mean arterial pressure", Med. Biol. Eng. Comput, vol.17, no.1, pp.17-18, 1979.
  2. B.D.Zislin, A.B.Chistyakov, V.A.Bagin, E.N.Soloukhin, V.A.Polunin, E.V.Devaikin and D.V.Pochepko, "Ways of Improving the Accuracy of Arterial Pressure Oscillometry", Biomed. Eng., vol.39 no.4, pp.174-178, 2005. https://doi.org/10.1007/s10527-005-0074-3
  3. Nelson R.M., Stebor A.D., Groh C.M., Timoney P.M., Theobald K.S. and Friedman B.A. "Determination of accuracy in Neonates for non-invasive blood pressure device using an improved algorithm", Blood Press Mon, 2002 7:123-9. https://doi.org/10.1097/00126097-200204000-00006
  4. Todd J. Brinton et al, "Development and Validation of a Noninvasive Method to Determine Arterial Pressure and Vascular Compliance", Am J Cardiol, vol.80, Issue.3, pp.323-330, 1997. https://doi.org/10.1016/S0002-9149(97)00353-6
  5. H.S.Oh, Y.J.Chee, J.S.Lee, I.Y.Kim, S.I.Kim, "BP Measurement algorithm based on the oscillometric method and The Single Peak Detection, using a controlled step deflation technique", Proc. 38th KOSOMBE, pp.12-15, 2008.
  6. A.F.Martin and C.B.Robert, "Random Sample Consensus: A Paradigm for Model Fitting with Applications to Image Analysis and Automated Cartography", Comm. ACM., vol.24, pp.381-395, 1981. https://doi.org/10.1145/358669.358692
  7. L.A.Geddes, M.Voelz. C.Combs. D.Reiner and C.F.Babbs, "Characterization of the oscillometric method for measuring indirect blood pressure," Ann Biomed Eng, vol.10, pp.271-280, 1982. https://doi.org/10.1007/BF02367308
  8. S.S.Lim, K.J.Lee, "Analysis of Oscillometric Model based on Shape of Arterial Pressureˮ, J. Biomed Eng. Res., vol.21, No.4, pp.411-417, 2000.
  9. Eoin O'Brien, Thomas Pickering, Roland Asmar, et al. on behalf of the Working Group on Blood Pressure Monitoring of the European Society of Hypertension, Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults, Blood Pressure Monitoring, vol.7, pp.3-17, 2002. https://doi.org/10.1097/00126097-200202000-00002
  10. Jongshill Lee, Daekyu Park, Hongsic Oh, Inyoung Kim, Dongfan Shen and Youngjoon Chee, "Digital recording system of sphygmomanometry," Blood Press Mon, vol.14, pp.77-81, 2009 https://doi.org/10.1097/MBP.0b013e3283262f45
  11. S.K. Hanneman, "Design, analysis, and interpretation of method-comparison studiesˮ, AACN Adv. Crit. Care. vol.19, no.2, pp.223-34, 2008. https://doi.org/10.1097/01.AACN.0000318125.41512.a3
  12. Association for the Advancement of Medical Instrumentation. ANSI/AAMI SP10:2002, Manual, electronic, or automated sphygmomanometers.