Abstract
The evaluation of Varian enhanced dynamic wedges (EDW) were performed in terms of quality assurance in external radiotherapy. The seven (10, 15, 20, 25, 30, 45, 60 deg.) EDW angles were evaluated for 6 and 15 MV x-rays in Varian Linac. The STT (segmented treatment table) for a field were calculated and compared with actual movement of the jaw using Dynalog files in order to evaluate mechanical operation. Two dimensional array detector and an ionization chamber were used to measure dose distributions in phantom from Linac. The mechanical movement of jaw was agreed with its expectation and two dimensional dose distributions including beam profiles were in agreement with RTP data approximately. In comparison with RTP calculations the percentage difference of output dose values for 100 MU irradiation was less than 2.9% and measured wedge factor was less than 2.6%. These results are shown that there is no problem in clinical applications of EDW equipped on this linac.
선형가속기의 동적쐐기(enhanced dynamic wedge: EDW)의 품질보증을 위하여 다양한 방법으로 동적쐐기의 특성을 평가하였다. 본 연구에서는 6 MV와 15 MV 엑스선에 대하여 각각 7종(10, 15, 20, 25, 30, 45, 60도)의 EDW를 평가하였다. EDW 작동에 필요한 STT(segmented treatment table)를 계산으로 구하고 로그파일을 통하여 기계적 작동을 평가하였으며, 이차원배열형검출기와 전리함을 사용하여 팬텀속 선량분포를 측정하고 치료계획시스템(RTP)과 비교하였다. EDW의 기계적 작동은 예상과 잘 일치하였으며, 빔측면도를 포함한 이차원선량분포는 근사적으로 RTP 계산과 일치하였다. 선형가속기 조사량 100 MU에 대한 출력선량은 RTP 계산과 2.9% 이내로 일치하였으며, 측정된 쐐기인자는 RTP 계산과 최대 2.6%를 보였다. 이 결과들은 본 선형가속기에 장착된 EDW의 임상적 적용에 문제가 없음을 의미한다.