DOI QR코드

DOI QR Code

Pulse wave analysis system using wrist type oximeter for u-Health service

u-Health 서비스 지원을 위한 착용형 옥시미터를 이용한 맥파 분석 시스템

  • Jung, Sang-Joong (Department of Electronic Engineering, Graduate school, Pukyong National University) ;
  • Seo, Yong-Su (Department of Electronic Engineering, Dongseo University) ;
  • Chung, Wan-Young (Department of Electronic Engineering, Graduate school, Pukyong National University)
  • 정상중 (부경대학교 대학원 전자공학과) ;
  • 서용수 (동서대학교 전자공학과) ;
  • 정완영 (부경대학교 대학원 전자공학과)
  • Published : 2010.01.30

Abstract

This paper describes a real time reliable monitoring method and analysis system using wrist type oximeter for ubiquitous healthcare service based on IEEE 802.15.4 standard. Photoplethysmograph(PPG) is simple and cost effective technique to measure blood volume change. In order to obtain and monitor physiological body signals continuously, a small size and low power consumption wrist type oximeter is designed for the measurement of oxygen saturation of a patient unobtrusively. The measured data is transferred to a central PC or server computer by using wireless sensor nodes in wireless sensor network for storage and analysis purposes. LabVIEW server program is designed to monitor stress indicator from heart rate variability(HRV) and process the measured PPG to accelerated plethysmograph(APG) by appling second order derivatives in server PC. These experimental results demonstrate that APG can precisely describe the features of an individual's PPG and be used as estimation of vascular elasticity for blood circulation.

Keywords

References

  1. I. Akyildiz, W. Su, Y. Sankarasubramaniam and E. Cayirci, “A survey on sensor networks”, IEEE Communications Magazine, vol. 40, no. 8, 2002.
  2. Y. D. Lee, E. Alassarela, and W. Y. Chung, “A wireless health monitoring system in multi-hop body sensor networks”, Procceding of ISMICT'07, 2007.
  3. Y. Mendelson, “Noninvasive pulse oximetry utilizing skin reflectance photoplethysmography”, IEEE Transactionson Biomedical Engineering, vol. 35, no. 10, pp. 798-805, 1988. https://doi.org/10.1109/10.7286
  4. 허영정, 윤길원, “광용적맥파(PPG)를 이용한 호흡수 측정에 있어서 동잡음을 이용한 정확도 향상”, 센서학회지, 제17권, 제6호, pp. 447-453, 2008. https://doi.org/10.5369/JSST.2008.17.6.447
  5. 새롬메디칼, “자율신경계 균형검사 및 가속도 맥파검사를 위한 Canopy9pro 측정결과 설명서”, IEMBIO Inc. 02-929-8188, www.srmc.biz/
  6. K. Takazawa, N. Tanaka, M. Fujita, O. Matsuoka, T. Saiki, M. Aikawa, S. Tamura, and C. Ibukiyama, “Assessment of vasoactive agents and vascular aging by the second derivative of photoplethysmogram waveform”, Hypertension, vol. 32, pp. 365-370, 1998. https://doi.org/10.1161/01.HYP.32.2.365
  7. J. Simek, D. Wichterle, V. Melenovsk, J. Malik, S. Svacina, and J. Widimsk, “Second derivative of the finger arterial pressure waveform: An insight into dynamics of the peripheral arterial pressure pulse”, Physiological Research, vol. 54, pp. 505-513, 2005
  8. Y. Iketani, T. Iketane, K. Tanakawa, and M. Murata, “Second derivate of photoplethysmogram in children and young people”, Jpn Circ J, vol. 64, pp. 110-116, 2000. https://doi.org/10.1253/jcj.64.110
  9. 우종민, “심박동수 변이(heart rate variability) 측정법의 개념과 임상적 활용”, 정신신체의학, 제12권, 제1호, pp. 3-14, 2004.
  10. 정기삼, “HRV의 개요”, 가정의학회지, 제25권, 제11호, pp. 528-532, Suppl. 2004.
  11. Task Force of The European Society of Cardiology and The North American Society of Pacing and Electrophysiology, “Heart rate variability: standards of measurement, physiological interpretation, and clinical use”, Circulation, vol. 93, pp. 1043-1065, 1996. https://doi.org/10.1161/01.CIR.93.5.1043
  12. W. Y. Chung, G. Walia, Y. D. Lee and R. Myllyla, “Design issues and implementation of query-driven healthcare system using wireless sensor Ad-hoc network”, 4th International Workshop on Wearable and Implantable Body Sensor Networks, March 26- 28, pp. 99-104, RWTH Achen University, Germany, 2007.
  13. 정완영, 신광식, 장성균, “확장성이 고려된 IEEE 802.15.4 기반의 저전력 범용 센서노드 설계 및 제작”, 센서학회지, 제16권, 제4호, pp. 247-253, 2007. https://doi.org/10.5369/JSST.2007.16.4.247
  14. J. G. Webster, “Design of pulse oximeters”, Taylor & Francis Group, 1997.

Cited by

  1. Movie Contents Design of One-Person Production Using IP Cameras vol.11, pp.2, 2011, https://doi.org/10.5392/JKCA.2011.11.2.001
  2. Adrenergic regulation during acute hepatic infection with Entamoeba histolytica in the hamster: involvement of oxidative stress, Nrf2 and NF-KappaB vol.24, 2017, https://doi.org/10.1051/parasite/2017048
  3. Fabrication and Evaluation of Sensor for Measuring Pulse Wave Velocity using Piezo Film and Conductive Textile vol.21, pp.2, 2012, https://doi.org/10.5369/JSST.2012.21.2.135
  4. Design and Implementation of a Customized Contents Service System for Relieving Users' Stress vol.11, pp.2, 2011, https://doi.org/10.5392/JKCA.2011.11.2.101