Abstract
A lab scale distillation experiment is conducted with small size packing used in lab scale multi-tray distillation equipment for the performance evaluation of distillation system. A sufficient surface are yielded with 6.7 mm cylindrical packings made of stainless steel, and a good liquid holdup and residence time are resulted. The comparison between the theoretical tray from the HYSYS and the experimental distillation outcome indicates that a 7 cm HETP from 27 cm packing height and a 8 cm HETP from 45 cm packing height are obtained. Comparing with the 8 cm HETP of commercial structured packing shows a similar experimental results obtained here. The 7 cm HETP is available with a complete insulation, and the importance of the insulation is proved. The results of this study indicates that a practical distillation column used in field can be tested in lab.
증류시스템 검증용 소형 다단 증류 장치에 활용할 수 있는 소형 충전물을 사용하여 실험실 규모의 증류실험을 실시하였다. 직경 6.7 mm의 실린더형 스테인레스 스틸 충전물을 이용하여 충분한 겉보기 표면적을 얻었고, 액의 체류량과 체류 속도도 양호한 결과를 얻었다. HYSYS를 이용한 이론단수의 계산치와 증류실험치를 비교하여 27 cm 충전층에 대해서는 7 cm이고, 45 cm 충전층에 대해서는 8 cm의 충전층이 이론단 1 단에 해당함을 알았다. 상업용 일정구조 배열의 충전탑에서 8 cm의 충전층이 이론단 1 단에 해당함과 비교하면 유사한 증류효율을 얻을 수 있음을 보여 주었다. 이러한 결과는 증류관의 보온이 완벽할 때 가능하였으며, 실험실에서의 증류 실험에서는 보온이 중요함을 입증하였다. 본 실험 결과로부터 실험실에서도 실제 현장에서 사용하는 증류탑과 유사한 증류 시스템의 시험이 가능함을 보여 준다.