Abstract
In this paper, in order to enhance the image reconstruction performance of white light scanning interferometry(WLI), we demonstrate the scattered point noise filtering performance of post-processing methods. Median filtering is similar to using an averaging filter. Because the median value is less sensitive than the mean to extreme values, the median filter can remove the scattered point noise from a height-map without significantly reducing the sharpness of the image. In several specific cases, however, the median filter can't remove the scattered point noise. Therefore, we propose a comparative mean filter that uses order-statistic filtering and the mean of the neighborhood pixels. The performance is demonstrated by measuring an array of metal solder balls fabricated on PCB. The proposed method reduced the noise pixels by 4.4 percent.
본 논문에서는 백색광 주사 간섭계의 측정시 외부 잡음이나 진동에 의한 간섭무늬 신호 왜곡으로 인하여 높이 영상에 발생하는 점 노이즈 제거를 위해, 데이터 처리 과정에서 높이 영상에 잡음 제거 필터를 적용하여 백색광 간섭계의 잡음 제거 성능을 향상시켰다. 백색광 간섭계의 높이 영상에 산포되어 있는 점 노이즈의 제거 방법으로 주로 사용되는 미디안(median) 필터는 점 노이즈 제거에 좋은 성능을 보이지만, 특정 경우 노이즈 제거 성능이 좋지 못하여 본 논문에서는 비교 평균 필터(Comparative Mean filter)를 제안하였다. 비교 평균 필터는 중심 화소 수치와 주변의 화소 수치를 비교하여 노이즈를 판별하고, 노이즈를 참조구간 화소 수치의 평균값으로 대체하는 필터이다. 비교평균 필터는 높이영상에서 노이즈 판별을 통한 보정과정으로 인하여 복원 영상의 선명도를 유지하면서 점 노이즈 제거에 좋은 성능을 보인다. 표면이 거칠고 굴곡이 심한 시료를 측정한 결과 후처리 하지 않은 경우와 비교하여 잡음성 화소를 약 1/3로 감소시키는 성능 향상이 있었다.