Abstract
In this study, in order to overcome the weakness of acoustic field analysis which is generally used for ultrasonic probe performance evaluation, automatic acoustic field measurement system and evaluation parameters were proposed. The comparisons between acoustic field simulation and measured acoustic distribution data of normal and abnormal channels were conducted to evaluate the availability of proposed system and evaluation parameters. First, the impulse response characteristic of sample probe was investigated to classify the normal elements and abnormal elements. And then, normal channels and abnormal channels with abnormal element were chosen. The suggested 12 evaluation parameters were calculated using the acoustic fields of these channels. The availability of proposed automatic acoustic field measurement system and evaluation parameters was confirmed. And the performance evaluation of ultrasonic probe using acoustic field analysis could be easier and faster.