Characteristics of Superconductive Pb shield for a Whole Head MEG System

헬멧형 뇌자도 장치로의 활용을 위한 Pb 초전도 차폐의 특성

  • Yu, K.K. (Korea Research Institute of Standards and Science) ;
  • Kim, K. (Korea Research Institute of Standards and Science) ;
  • Lee, Y.H. (Korea Research Institute of Standards and Science) ;
  • Kwon, H. (Korea Research Institute of Standards and Science)
  • Published : 2009.10.30

Abstract

We have investigated the characteristics of a superconductive Pb shield for hemispherical shape and plate to improving signal-to-noise ratio(SNR) of biomagnetism. We measured the shielding factor for the position of helmet shape Pb and for changing the distance from Pb surface. To make a uniform magnetic field, a $1.5m{\times}1.5m$ set of the helmholtz coils activated at several frequencies. The shielding factor of hemispherical shape Pb was from 20 to 57 dB and of Pb plate was about $6{\sim}26dB$ as a function of distance from the lead surface. The shielding factor was rapidly reduced as increasing the distance from Pb surface. The white noise of superconductive quantum interference device(SQUID) with a superconductive shield was about $12fT/Hz^{1/2}$ at 1 Hz, $7fT/Hz^{1/2}$ at 100 Hz. The white noise was more increased about two times than conventional SQUID system without Pb shielding. An auditory signal was measured by first order gradiometer and magnetometer with Pb superconductive shield and compared the SNR. The SQUID system with Pb shield had better performance at low frequency noise level.

Keywords

References

  1. 이용호, 권혁찬, 김진목, “생체자기계측을 위한 자기차폐실의 설치 및 특성”, Journal of the Korean Mag. Soc., Vol. 6, NO. 4, pp. 123-129, August, 1996.
  2. G. stroink and B. Blackford, “Aluminum shielded room for biomagnetic measurements”, Rev. Sci. Instrum., 52(3), pp. 463-468, Mar. 1981. https://doi.org/10.1063/1.1136612
  3. Gen Uehara, Yoshiaki Adachi, Masahiro Shimogawara, and Hisashi Kado, “Multi-Channel SQUID Systems for Biomagnetic Measurement”, IEICE Trans. Electron., Vol. E86-C, pp. 43-53, 2003.
  4. 이용호, 김진목, 권혁찬, “자속-전압 변환계수가 큰 이중이완발진 초전도양자간섭소자의 제작 및 동작 조건에 따른 잡음특성”, Sae Mulli, Vol. 39, Number 1, pp. 86-93, February, 1999.
  5. Keiji Iramina, and Shoogo Ueno, “Measurement of Brainstem Auditory Evoked Magnetic Fields Using a Highly Sensitive SQUID Magnetometer with a Variable Baseline”, IEEE Trans. on Mag., Vol. 31, pp. 4271-4273, November 1995. https://doi.org/10.1109/20.489949
  6. Y. H. Lee, K. Kim, J. M. Kim, and Y. K. Park, “A low noise multichannel magnetocardiogram system for the diagnosis of heart electric activity”, Korean society of medical and biological Eng., Vol. 27, pp.154-163, 2006.
  7. R. H. Kraus, E. R. Flynn, M. A. Espy, A. Matlashov, W. Overton, C. C. Wood, M. V. Perters, and P. Ruminer, “ First Results for a Superconducting Imaging-Surface Sensor Array for Magnetocardiography”, Recent Advances in Biomagentism Proceeding, pp. 43-46, 1999.
  8. D. B. Hulsteyn, A. G. Petsch, E. R. Flynn, and W.C. Overton, “Superconducting Imaging Surface Magnetometery”, Rev. Sci. Instrum., Vol. 66, pp. 3777-3784.
  9. H. Ohta, T. Matsui, M. Aono, and K. Kamijo, “A 64-channel whole head SQUID system in a superconducting magnetic shield”, Supercon. Sci. Technol., Vol.12, pp. 762-765, 1999. https://doi.org/10.1088/0953-2048/12/11/321