Consideration about Ozone Generation in the Treatment Room While Treating a Patient

방사선 치료 시 치료실 내에서 발생하는 오존에 관한 고찰

  • Kwak, Yong-Kuk (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Yoon, Il-Kyu (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Lee, Jae-Hee (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Yoo, Suk-Hyun (Department of Radiation Oncology, Seoul National University Hospital)
  • 곽용국 (서울대학교병원 방사선종양학과) ;
  • 윤일규 (서울대학교병원 방사선종양학과) ;
  • 이제희 (서울대학교병원 방사선종양학과) ;
  • 유숙현 (서울대학교병원 방사선종양학과)
  • Received : 2009.06.02
  • Accepted : 2009.07.10
  • Published : 2009.09.30

Abstract

Purpose: Measure the ozone level in the treatment room while treating a patient so want to know the degree of contamination caused by ozone occurrence. Materials and Methods: Use the linear accelerator (Clinac 21EX, Varian, USA) with the ozone meter (series-200, aeroQual, New Zealand) and water phantom (Wellhofer, IBA, Germany) is irradiated the radiation so that measured the ozone generation level according to MU, dose-rate, SSD, field size, energy, delay time and put the ozone meter in the treatment room actually while treating a patient so measured the daily ozone level variation. Results: While irradiating the radiation, degree of ozone contamination wasn't affected by the energy but mostly in case of electron beam, ozone level was higher than photon beam. The higher dose-rate (0.016~0.025 ppm/hr), the farther SSD (0.018~0.030 ppm/hr), the wider field sizes (0.016~0.025 ppm/hr), the more MU (0.018~0.046 ppm/hr), it occurred high ozone level. Ozone decrement according to delay time changed the background level (0.016 ppm/hr) after elapsed time of 10 minutes from irradiating radiation. And daily ozone occurrence level in the treatment room was below ozone standard level 0.1 ppm/hr (average:0.06 ppm/8 hr) but it could confirm that ozone generation level was included the level (max:0.038 ppm/hr) above 0.02 ppm/hr which patient could perceive. Conclusion: Through ozone level according to variation of certain conditions, actually in the treatment room ozone generation level didn't damaged to patients or workers. Commonly peoples think that ozone was harmful gas but it thought that small amount of ozone generation level while treating a patient was beneficial in the treatment room through air purge action of pathogenic germ or virus sterilization.

목 적: 방사선 치료 시 치료실 내에서 발생하는 오존량을 측정하여 오존 발생으로 인한 오염 정도를 알아보고자 한다. 대상 및 방법: 선형가속기(Clinac 21EX, Varian, USA)와 오존 측정기(series-200, aeroQual, New Zealand)를 이용하여 water phantom (Wellhofer, IBA, Germany)에 방사선을 조사하여 MU, 선량율, 선원-표면간 거리(SSD), 조사야, 에너지, 경과 시간에 따라 발생되는 오존량을 측정 분석하였으며, 실제 환자치료 시 치료실 내에 오존 측정기를 위치하여 일일 오존 변화량을 측정하였다. 결 과: 방사선 발생 중 생성되는 오존의 오염도는 에너지에 따른 영향을 크게 받지 않지만 대체로 광자선보다 전자선 조사 시(0.016~0.028 ppm/hr) 많이 발생하였다. 또한, Dose-Rate가 높을수록(0.016~0.025 ppm/hr), SSD가 멀어지고(0.018~0.030 ppm/hr), 조사야가 넓어지고(0.016~0.025 ppm/hr), MU가 많을수록(0.018~0.046 ppm/hr) 오존 농도가 높아졌다. 시간 경과에 따른 오존의 감소량은 방사선을 조사한 후 10분이 경과된 후부터 background 농도(0.016 ppm/hr)로 변화하였다. 그리고 방사선 치료실 내의 일일 발생하는 오존의 농도는 실내의 오존 허용기준인 0.1 ppm/hr (측정 평균 0.06 ppm/8 hr) 이하이지만, 냄새에 민감한 환자들이 감지할 수 있는 수준인 0.02 ppm/hr 이상의 농도(최대: 0.038 ppm/hr)를 포함하는 것을 확인할 수 있었다. 결 론: 일정한 조건의 변화에 따른 오존 농도를 통하여 실제 치료실 내 오존 발생량은 환자나 작업종사자에 대해 유해한 작용을 미치는 수준은 아니었다. 흔히 오존을 해로운 기체라고 생각하지만 방사선 치료 시 발생되는 미량의 치료실 내 오존은 오히려 병원성 세균이나 바이러스 살균 등의 공기 정화 작용을 함으로써 치료실 내 이로운 오존의 역할을 할 것이라 사료된다.

Keywords