An Initial Study on the Reliability Assurance in PET/CT Standardized Uptake Values

PET/CT 에서 표준섭취계수(SUV)의 신뢰성 확보를 위한 초기연구

  • Park, Hoon-Hee (Department of Nuclear Medicine, Severance Hospital, Yonsei University Health System) ;
  • Kim, Jung-Yul (Department of Nuclear Medicine, Severance Hospital, Yonsei University Health System) ;
  • Lee, Seung-Jae (Department of Nuclear Medicine, Severance Hospital, Yonsei University Health System) ;
  • Park, Min-Soo (Department of Nuclear Medicine, Severance Hospital, Yonsei University Health System) ;
  • NamKoong, Hyuk (Department of Nuclear Medicine, Severance Hospital, Yonsei University Health System) ;
  • Lim, Han-Sang (Department of Nuclear Medicine, Severance Hospital, Yonsei University Health System) ;
  • Oh, Ki-Baek (Department of Nuclear Medicine, Severance Hospital, Yonsei University Health System) ;
  • Kim, Jae-Sam (Department of Nuclear Medicine, Severance Hospital, Yonsei University Health System) ;
  • Lee, Chang-Ho (Department of Nuclear Medicine, Severance Hospital, Yonsei University Health System) ;
  • Jin, Gye-Hwan (Department of Radiology, Nambu University)
  • 박훈희 (연세의료원 세브란스병원 핵의학과) ;
  • 김정열 (연세의료원 세브란스병원 핵의학과) ;
  • 이승재 (연세의료원 세브란스병원 핵의학과) ;
  • 박민수 (연세의료원 세브란스병원 핵의학과) ;
  • 남궁혁 (연세의료원 세브란스병원 핵의학과) ;
  • 임한상 (연세의료원 세브란스병원 핵의학과) ;
  • 오기백 (연세의료원 세브란스병원 핵의학과) ;
  • 김재삼 (연세의료원 세브란스병원 핵의학과) ;
  • 이창호 (연세의료원 세브란스병원 핵의학과) ;
  • 진계환 (남부대학교 방사선과)
  • Received : 2009.06.15
  • Accepted : 2009.07.13
  • Published : 2009.10.10

Abstract

Purpose: As the number of domestic medical institutions installing PET/CT is increasing rapidly, the transfer of PET/CT images among medical institutions is also increasing. Thus, it is necessary to collect the comparative SUV data from several medical institutions' PET/CT systems through a phantom study which semi-quantitatively compares the SUV on one bed, the change scale of the SUV on the slices, and the time of measuring. The phantom study to find differences among the SUVs from various PET/CT offers the opportunity to obtain the reliability of the SUV in PET/CT images. Materials and Methods: Ten PET/CT systems from medical institutions in Korea were used. To obtain the accurate data, the study has been using the radiation detector of Korea Research Institute of Standards and Science to verify. The internal structures of NEMA $phantom^{TM}$ were removed and Six thousand milliliters of distilled water which has 1mCi of $^{18}F$-FDG put into the phantom. The water was properly integrated with $^{18}F$-FDG using magnetic stirrer. The images were acquired at 60, 70, 80, 90, 100, 110 and 120-minutes for 3 minute each. Two hundred square centimeters of region of interests were placed and analyzed. To confirm the usefulness, the correction-table came out from patients' data. Results: The coefficient of variability of the SUV from -11.0 to 9.90 % fell into the range of international standards(${\pm}10%$) along with the SUV on a bed, the change scale of the SUV on the slices, and the time of measuring, except one PET/CT system. Using the data of the differences among the SUVs, we came to withdraw the correction-table ranging from 0.803 to 1.246. The correction-table was confirmed its usefulness through Linear Regression Analysis which was applied to normal cases. Conclusions: Although studies have been made on the variation of the SUV, there is little attention on the standardization of the SUV. Based on this study of the quantitatively comparable data about the SUV accommodating the correction-table, it would help to have more corrective diagnosis.

최근 PET/CT가 급격하게 증가하면서 의료기관 사이에 영상의 이동도 증가하고 있다. 이에 서로 다른 의료기관 간의 시스템 별 표준섭취계수 차이를 반영하기 위하여 1 Bed에서 표준섭취계수, 슬라이스 내의 표준섭취계수 변화율과 측정시간에 따른 표준섭취계수를 정량적으로 비교할 수 있는 팬텀을 이용한 비교측정이 필요하다. 본 연구에서는 임상에서 사용하는 다양한 PET/CT 시스템의 표준섭취계수 차이에 대한 연구를 통하여, PET/CT 영상의 표준화섭취계수의 신뢰성을 확보하고자 하였다. 대한민국 전국에 분포되어 의료기관에 설치된 PET/CT 장비 10대를 대상으로 하였으므로, 정확한 방사능 산출을 위하여, 한국표준과학원의 검출기로 검증을 통하여 실험하였다. NEMA PET $Phantom^{TM}$의 내부구조물을 제거하고 $^{18}F$-FDG 1 mCi를 6,000 mL 증류수에 균일하고 분포하도록 마그네틱 스터러와 마그네틱 바의 회전력으로 희석하여, 팬텀에 주입하였다. 주입 후 60분, 70분, 80분, 90분, 100분, 110분, 120분에 3분간 영상을 획득하고, 관심영역 $200\;cm^2$에 대하여 분석하였으며, 유용성 확인을 위하여 임상환자를 대상으로 교정표를 산출하였다. 1 Bed에서 표준섭취계수, 슬라이스 내의 표준섭취계수 변화율, 측정시간에 따른 표준 섭취계수 변화율과 함께 표준섭취계수의 변이계수가 -11.0~9.90%로 국제적으로 통용되는 기준인 ${\pm}10%$를 1개의 장비를 제외하고 모두 만족하였다. 또한, 시스템 별 평균 표준섭쉬계수 차이를 이용하여 0.803~1.246으로 이루어진 교정표를 도출하였고, 정상인을 통한 임상 적용에서 선형회귀분석을 통하여 유의함을 확인하였다. 본 연구를 통하여 PET/CT 장비간의 표준섭취계수 차이를 교정표를 이용하여 정량적으로 비교할 있는 근거를 제시한다면 정확한 진단에 도움이 되며, 아울러 이에 대한 세계적 기준이 명확하지 않기에 유사 연구에 도움이 되리라 사료된다.