An Initial Study on the Reliability Assurance in PET/CT Standardized Uptake Values
PET/CT 에서 표준섭취계수(SUV)의 신뢰성 확보를 위한 초기연구
- Park, Hoon-Hee (Department of Nuclear Medicine, Severance Hospital, Yonsei University Health System) ;
- Kim, Jung-Yul (Department of Nuclear Medicine, Severance Hospital, Yonsei University Health System) ;
- Lee, Seung-Jae (Department of Nuclear Medicine, Severance Hospital, Yonsei University Health System) ;
- Park, Min-Soo (Department of Nuclear Medicine, Severance Hospital, Yonsei University Health System) ;
- NamKoong, Hyuk (Department of Nuclear Medicine, Severance Hospital, Yonsei University Health System) ;
- Lim, Han-Sang (Department of Nuclear Medicine, Severance Hospital, Yonsei University Health System) ;
- Oh, Ki-Baek (Department of Nuclear Medicine, Severance Hospital, Yonsei University Health System) ;
- Kim, Jae-Sam (Department of Nuclear Medicine, Severance Hospital, Yonsei University Health System) ;
- Lee, Chang-Ho (Department of Nuclear Medicine, Severance Hospital, Yonsei University Health System) ;
- Jin, Gye-Hwan (Department of Radiology, Nambu University)
- 박훈희 (연세의료원 세브란스병원 핵의학과) ;
- 김정열 (연세의료원 세브란스병원 핵의학과) ;
- 이승재 (연세의료원 세브란스병원 핵의학과) ;
- 박민수 (연세의료원 세브란스병원 핵의학과) ;
- 남궁혁 (연세의료원 세브란스병원 핵의학과) ;
- 임한상 (연세의료원 세브란스병원 핵의학과) ;
- 오기백 (연세의료원 세브란스병원 핵의학과) ;
- 김재삼 (연세의료원 세브란스병원 핵의학과) ;
- 이창호 (연세의료원 세브란스병원 핵의학과) ;
- 진계환 (남부대학교 방사선과)
- Received : 2009.06.15
- Accepted : 2009.07.13
- Published : 2009.10.10
Abstract
Purpose: As the number of domestic medical institutions installing PET/CT is increasing rapidly, the transfer of PET/CT images among medical institutions is also increasing. Thus, it is necessary to collect the comparative SUV data from several medical institutions' PET/CT systems through a phantom study which semi-quantitatively compares the SUV on one bed, the change scale of the SUV on the slices, and the time of measuring. The phantom study to find differences among the SUVs from various PET/CT offers the opportunity to obtain the reliability of the SUV in PET/CT images. Materials and Methods: Ten PET/CT systems from medical institutions in Korea were used. To obtain the accurate data, the study has been using the radiation detector of Korea Research Institute of Standards and Science to verify. The internal structures of NEMA
최근 PET/CT가 급격하게 증가하면서 의료기관 사이에 영상의 이동도 증가하고 있다. 이에 서로 다른 의료기관 간의 시스템 별 표준섭취계수 차이를 반영하기 위하여 1 Bed에서 표준섭취계수, 슬라이스 내의 표준섭취계수 변화율과 측정시간에 따른 표준섭취계수를 정량적으로 비교할 수 있는 팬텀을 이용한 비교측정이 필요하다. 본 연구에서는 임상에서 사용하는 다양한 PET/CT 시스템의 표준섭취계수 차이에 대한 연구를 통하여, PET/CT 영상의 표준화섭취계수의 신뢰성을 확보하고자 하였다. 대한민국 전국에 분포되어 의료기관에 설치된 PET/CT 장비 10대를 대상으로 하였으므로, 정확한 방사능 산출을 위하여, 한국표준과학원의 검출기로 검증을 통하여 실험하였다. NEMA PET