Implementation of Strength Estimation Algorithm on the Metallic Plate Fixation

금속고정용의 강도 평가 알고리즘 구현

  • 김정래 (을지대학교 보건과학대학 의료공학과) ;
  • 김규호 (을지대학교 의료산업학부) ;
  • 이기영 (을지대학교 의료산업학부)
  • Received : 2009.04.23
  • Published : 2009.06.30

Abstract

This study was developed the metallic plate for bone fixation in the neurosurgery and general surgery and plates has a finn place in bone operating and treatment. The plates can be realized to bending strength and stiffness for strength estimation. Maximum point of bending curves has a bending point(P) with maximum load which to applied nearly 0.2% offset displacement. The device's sizing has a ${\Phi}13$ and ${\Phi}18$, and algorithm of strength estimation compared a plate(${\Phi}13$, ${\Phi}18$, ${\Phi}13-{\Phi}18$). The bending strength of the curved metallic plate has to evaluate maximum of a 311N, 387N, 410N, 474N. When a displacement preserve with a load, tensile stress through to press a plate is 274N, 324N, 382N, 394N. The algorithm of strength estimation can be used to support estimation of bending strength and stiffness. Their tool bring to settlement in the new basic algorithm for evidence with varied adjustment.

본 논문은 신경외과 및 일반외과 영역에서 골 고정용 금속판으로 골 수술 및 치료를 하는 과정에서 골 고정요 금속판의 굴곡강도와 강성을 평가할 수 있는 강요평가 알고리즘을 구현하였다. 굴곡곡선의 최대 포인트 점은 0.2% 상쇄변위(offset displacement)에서 평형하게 접근 할 수 있는 포인트 P점을 최대 하중에 부여하는 굴곡 점으로 적용하였다. 실험에 사용한 플레이트(Plate)는 ${\Phi}13$${\Phi}18$이며, 강도평가의 알고리즘은 플레이트 ${\Phi}13$, ${\Phi}18$, ${\Phi}13-{\Phi}18$ 비교하였으며, 플레이트에 압박하는 힘의 알고리즘을 산출한 결과 굽힘강도의 최대값이 311N, 387N, 410N, 474N로 나타났고, 변위(Displacement)에 따라 하중에 견디는 시점인 인장강도가274(N), 324(N), 382(N), 394(N)로 나타났음으로 이 시점이 금속판의 굴곡운동 값으로 결정되었다. 본 연구의 결과로 한 개의 축을 기준으로 발생하는 골 고정용 금속판의 강도 조절 문제를 평가알고리즘으로 구성함으로 해결할 수 있었고, 형태 변화에 따른 조절기능을 검증할 수 있는 시스템이 새로운 알고리즘 형성으로 가능할 것으로 예상된다.

Keywords

References

  1. Schatzker J, M.E. Muller-on his 80th Birthday. AO Dialogue, 11(1), pp.7-12. 1998.
  2. Bynum D. Ledbetter WB, Boyd CL, Ray DR. Holding characteristics of fasteners in bone. Exp Mech, Vol 11, pp.363-369, 1971. https://doi.org/10.1007/BF02320538
  3. Schatzker J, Anderson R, Murnahan P. The holding power of orthopaedic screws in vivo. Clin Orthop, Vol 108, pp.115-126, 1975. https://doi.org/10.1097/00003086-197505000-00019
  4. Perren S.M, Cordey J, The concept of interfragmentary strain. Berlin Heidelberg New York: Springer-Verlag, 1980.
  5. Nunamaker DN, Perren SM Force measurements in screw fixation. J Biomechanical, Vol 9, pp.669-675, 1976. https://doi.org/10.1016/0021-9290(76)90168-8
  6. Von Arx Ch. Schububertragung durch Reibung bei Plattenosteosynthese. AO Bulletin, MD Thesis, Basel 1975.
  7. Annual Book of ASTM Standards, American Society of Testing and Materials, West Conshohocken, PA 19428, Vol 13.01, 2001.
  8. Ruedi T. P, Murphy W.M, AO Principle of Practure Management, 2nd ed., Theime Stuttgart. New York, 2000.
  9. Annual Book of ASTM Standards, American Society of Testing and Materials, West Conshohocken, PA 19428, Vol 03.01, 2001.
  10. ASTM Committee F-4 on Medical and Surgical Materials and Devices, Originally published as F 382-73, Last previous edition F 382-98a, Published December 1999.
  11. Ansell RH, Scales JT. A study of some factors which affect the strength of screws and their insertion and holding power in bone. J Biomech, Vol 1, pp. 279-302, 1968. https://doi.org/10.1016/0021-9290(68)90023-7
  12. S.M. Perren, J. Cordey, F. Baumgart, B. A. Rahn, J Schatzker, Technical and biomechanical aspects of screws used for bone surgery. International Journal of Orthopaedic Truma, Vol 2, pp.31-48, 1992.
  13. T.G. Beckwith & R.D. Marangoni, Mechanical Measurements(4th Ed,) pp.23-26, pp.571-601, pp.611-628. 2000.
  14. Available from International Standards Organization, Rue de Varembe, Case Postale 56, CH-1211, Geneva 20, Switzerland. 2000.
  15. Available from America National Standards Institute, 11 W. 42nd St, New York, NY 10036. 2001.
  16. Cordey J, Widmer W, Rohner A, et al., Dosierung des Drehmoments beim Einsetzen von Knochenschrauben. (Exper imen telle Studie an Kortikalisschrauben mit Hilfeelektronischer Drehmomentsch raubenzie her). Z Ortbop, 115, pp.601-602, 1977.
  17. 황규섭, 센서활용기술, 기전연구사, pp.405, 1989.