DOI QR코드

DOI QR Code

Uncertainty assessment of industrial platinum resistance thermometers for different lead-wire connection methods

산업용 백금저항온도센서의 결선방식에 따른 측정불확도 평가방법

  • Kim, Yong-Gyoo (Div. Physical Metrology, Korea Research Institute of Standards and Science) ;
  • Gam, Kee-Sool (Div. Physical Metrology, Korea Research Institute of Standards and Science) ;
  • Yang, In-Seok (Div. Physical Metrology, Korea Research Institute of Standards and Science)
  • 김용규 (한국표준과학연구원 기반표준본부) ;
  • 감기술 (한국표준과학연구원 기반표준본부) ;
  • 양인석 (한국표준과학연구원 기반표준본부)
  • Published : 2009.07.31

Abstract

To estimate the measurement uncertainty for industrial platinum resistance thermometers(IPRTs) made with 3-wire connection, the immersion temperature profile was investigated using a liquid bath. Two types of IPRTs having lead wires made of silver and nickel were constructed and the immersion profiles were measured at temperatures from -50 $^{\circ}C$ to 250 $^{\circ}C$ using 3-wire and 4-wire method. As immersion depth and temperature increased, the resistances measured by 3-wire method increased linearly but not for 4-wire method. To calibrate a 3-wire IPRT, the immersion effect must be accounted for. We propose a linear equation to assess correctly the measurement uncertainty.

Keywords

References

  1. Techniques for approximating the international temperature scale of 1990, BIPM, 2005
  2. P.Marcarino, A.Merlone, P.P.M.Steur, A.Actis, and M.Antinori, 'Proposal:new reference functions for industrial platinum resistance thermometers', TEMPMEKO 2004, vol. 2, pp. 807-812, 2004
  3. T.Weckstrom, 'Pt-100 thermometers and the ITS-90', TEMPMEKO 2004, vol. 2, pp. 813-816, 2004
  4. 산업용 저항온도계 표준교정절차서, KASTO 03-50106-077, 한국계량측정협회, 2003
  5. Smithells Metals Reference Book, 8th edition, Elsevier Butterworth-Heinemann, 2004
  6. H. Preston-Thomas, 'The International Temperature Scale of 1990 (ITS-90)', Metrologia, vol. 27, pp. 3-10, 1990 https://doi.org/10.1088/0026-1394/27/1/002
  7. 측정불확도 표현지침, KRISS-98-096-SP, 한국표준과학연구원, 1998