Abstract
To estimate the measurement uncertainty for industrial platinum resistance thermometers(IPRTs) made with 3-wire connection, the immersion temperature profile was investigated using a liquid bath. Two types of IPRTs having lead wires made of silver and nickel were constructed and the immersion profiles were measured at temperatures from -50 $^{\circ}C$ to 250 $^{\circ}C$ using 3-wire and 4-wire method. As immersion depth and temperature increased, the resistances measured by 3-wire method increased linearly but not for 4-wire method. To calibrate a 3-wire IPRT, the immersion effect must be accounted for. We propose a linear equation to assess correctly the measurement uncertainty.