DOI QR코드

DOI QR Code

Comparative Evaluation of Resolution according to Frequency Change for SPECT

단일광자방출전산화단층촬영(SPECT)을 위한 주파수 변동에 따른 분해능 비교평가

  • 김성철 (가천의과학대학교 방사선학과) ;
  • 박수연 (가천의과학대학교 방사선학과) ;
  • 조영권 (세란동물병원) ;
  • 안성민 (가천의과학대학교 방사선학과)
  • Received : 2009.08.04
  • Accepted : 2009.09.29
  • Published : 2009.12.28

Abstract

In clinical SPECT, FWHM provides quality information of the images obtained under different observing conditions. Therefore, the purpose of this study was to determine the optimum cut off level with comparison of FWHM according to cut off levels in each filter - Band limited, Sheep-Logan, Hanning, Hamming, Low pass cosine, Parazen and Butterworth filter in a SPECT camera. I recorded images along the X, Y, Z-axis with 99mTcO4 point source and measured FWHM with profile curves. In conclusion, all filters showed the longest figures of FWHM with cut off level 0.4, which has the worst image resolution. The images with cut off level 0.7 showed best image resolution. The shortest average of FWHM in MS2 was $11.07\pm0.07mm$ using the Butterworth filter, in MS3 it was $8.44\pm0.19mm$ through using the Hanning filter.

SPECT 장치에서 반치폭은 분해능을 평가할 수 있는 척도가 되므로, 본 연구에서는 검출기가 각각 2개와 3개가 있는 장치를 이용하여 우리나라에서 가장 많이 사용되는 7개의 필터 - Band limited, Sheep-logan, Hanning, Hamming, Low pass cosine, Parazen and Butterworth - 에 대해 세 개의 점선원으로 filter별로 MTF curve를 얻어 각각의 축방향에 따른 cut off level 과 profile 곡선을 이용하여 반치폭(FWHM)의 크기의 변화를 관찰하였다. 실험 결과 모든 filter에서 cut off level 0.4였을 때 가장 큰 반치폭의 길이가 측정되어 분해능이 가장 떨어졌고, cut off level 0.7에서 가장 높은 분해능을 보였다. MS2에서 가장 짧은 반치폭은 Butterworth filter사용시 $11.07\pm0.07mm$이었으며, MS3에서는 Hanning filter사용시 $8.44\pm0.19mm$로 나타났다.

Keywords

References

  1. 정진호, 최용, 홍기조, 민병준, 호위, 강병준, "핵의학 영상기기의 최근 진보", 핵의학분자영상, 제42권, 제2호, pp.98-111, 2008.
  2. 박석건, 송재범, "SPECT", 대한핵의학기술학회, pp.40-68, 2001.
  3. 이철주, 강준규, 김형태, 박인덕, 정영미, "좌내흉 동맥과 복재정맥편을 사용한 관상동맥우회로술 환자에서의 추적조사에서 MDCT의 유용성", 대한흉부외과학회지, 제38권, 제6호, pp.410-414, 2005.
  4. 유병수, 신윤철, 김건일, 김응중, 지현근, "관상동맥조영술과 MDCT를 이용한 관상동맥 이식편의 비교평가", 대한흉부외과학회지, 제39권, 제1호, pp.42-47, 2006.
  5. 최윤석, 윤호중, 정승은, 최용원, 이동현, 박철수, 오용석, 정욱성, 승기배, 김재형, 최규보 "MDCT 에서의 관상 동맥 석회화 병변과 관상동맥 조영술에서의 협착과의 상관관계", Korean circulation journal, 제37권, 제4호, pp.167-172, 2007. https://doi.org/10.4070/kcj.2007.37.4.167
  6. 윤연이, 전은주, 최의근, 조영진, 이원재, 최상일, 최동주, 장혁재, "64열 심장다절편 CT에서 발견되는 비심장소견:흉부 CT와의 비교", Korean circulation journal, 제38권, 제5호, pp.276-283, 2008. https://doi.org/10.4070/kcj.2008.38.5.276
  7. E. Paul Christian and Kristen M. Waterstram -Rich, "Nuclear medicine and PET/CT : technology and techniques," Mosby, 2007.
  8. Pete Shackett, "Nuclear medicine technology : procedures and quick reference," Lippincott Williams & Wilkins, 2009.
  9. 정준기, 이명철, "고창순 핵의학", 고려의학, 2008.
  10. Cyril Riddell, Ricahrd E.Carson, Jorge A. Carrasquillo, Steven K. Libutti, David N. Danforth, Millie Whatley and Stephen L. Bacharach, "Noise reduction in oncology FDG PET images by iterative reconstruction : A Quantitative assessment, Journal of Nuclear Medicine," Vol.42, No.9, pp.1316-1323, 2001.
  11. "Performance measurements of scintillation camera NEMA Standards Publication NU1-2001," pp.11-13, 2001.
  12. Shepp LA and Vardi Y, "Maximum likelihood reconstruction for emission tomography," IEEE Trans Med Imaging, Vol.1, No.2, pp.113-122, 1982. https://doi.org/10.1109/TMI.1982.4307558
  13. Avid R. Gilland, Benjamin M. W. Tsui, William H. McCartney. J. Randolph Perry and Jan Berg, "Determination of the optimum filter function for SPECT imaging." Journal of Nuclear Medicine, Vol.29, pp.643-650, 1988.
  14. R. James. Galt,H. Lee Hise. Ernest V. Garcia. and David J. Nowak, "Filtering in frequency space," Journal of Nuclear Medicine Technology, Vol.14, pp.152-160, 1986.
  15. R. L. Eisnr, D. J. Nowak, R. Pettigrew, and W. Fajman, "Fundamentals of 180 degree acquisition in SPECT Imaging," Journal of Nuclear Medicine, Vol.27, No.11, pp.1717-1728, 1986.