Abstract
Breast cancer is the most common form of cancer among korean woman. Therefore, the early detection activities of breast cancer such as breast self-examinations, clinical breast examinations, mammography are important. A yearly mammography examination has been recommended for women aged 40 and older for the early detection of breast cancer in asymptomatic periods. However, the glandular tissue of breast is the most radiation-sensitive tissue, and the determination of average glandular dose (AGD) forms an important part of the quality control of the mammographic systems. Because of the difficulty of estimating AGD directly, it is often estimated from the measurements of the incident air kerma and by applying the appropriate conversion factors. The primary objective of this study was to standardize the method of measuring AGD. The secondary objective was to evaluate the relationships between AGD per various composition and thickness of the breast using Monte Carlo simulations. As a result, we standardized the method of measuring AGD according to International Atomic Energy Agency (IAEA) guidelines (CoP: an international code of practice). Overall, AGD for mammographic practice in Korea was less than 3.0 mGy recommended by the Korea Food and Drug Adminstration (KFDA) protocol, and Korean Institute for Accreditation of Medical Image (KIAMI). The measured and simulated AGD for a given condition were calculated as 1.7 and 1.6 mGy, respectively. For the AGDs obtained, there was no significant difference between them. The simulated AGD was dependent on the fraction of glandular tissue of the breast. The AGD increases with increasing of the breast glandularity due to increasing absorption of low energy photons. The AGD also increases as a function of breast thickness. In conclusion, the results of this study could be used as a baseline to establish a reference level of radiation dose in mammography.
유방암은 우리나라 여성에게 발생하는 암 중에서 발병률이 높은 암이다. 따라서 유방자가진단, 임상적 검사, 유방 X선촬영 등과 같은 유방암 초기 검출검사는 중요하다. 이들 중 유방 X선촬영은 무자각증상 시기에 유방암 초기 검출을 위해 40대 이상의 여성들에게 매년마다 시행하기를 권고하고 있다. 그러나 유방의 유선조직은 방사선 민감성 조직이다. 이에 유방 X선촬영장치의 평가 중 평균 유선선량 측정은 매우 중요한 부분이 된다. 평균 유선선량의 직접측정은 어렵기 때문에 incident air kerma 측정, 적절한 변환계수 등을 적용하여 계산하게 된다. 따라서 본 연구의 첫 번째 목적은 평균 유선선량의 측정 방법의 표준화이다. 두 번째는 몬테칼로 시뮬레이션을 이용하여 평균 유선선량과 유방 실질양상 및 두께와 의 상관관계를 분석하고자 했다. 본 연구 결과, IAEA 가이드라인(CoP)에 따라 평균 유선선량 측정방법을 제시하였다. 전반적으로 우리나라는 유방촬영 시 받는 평균 유선선량이 식품의약품안전청 및 한국의료영상품질관리원에서 제시하는 3mGy 이하로 측정되었다. 측정된 평균 유선선량과 시뮬레이션된 평균 유선선량은 각각 1.7과 1.6 mGy로 큰 차이를 보이지 않았다. 시뮬레이션된 평균 유선선량은 주로 유방의 유선조직의 비율에 따라 의존한다. 유방의 glandularity 증가에 따라 낮은 에너지의 광자의 흡수가 증가하여 평균 유선선량도 증가하였다. 또한 유방의 두께가 두꺼울수록 평균 유선선량은 증가하였다. 결과적으로, 본 연구는 유방촬영의 진단 참고준위 확립을 위한 기초자료로 활용될 것이다.