Development of Asbestos Quality Control Sample for Proficiency Analytical Testing 1 - Development of Manufacturing Apparatus and Sample Preparing Procedure for Asbestos Quality Control Sample -

석면분석 정도관리용 표준시료 개발연구 I - 석면분석 정도관리용 표준시료 제조장치 개발 및 시료제조 방법 확립 -

  • Yi, Gwang Yong (Occupational Safety & Health Research Institute, KOSHA) ;
  • Lee, Jong- Han (Occupational Safety & Health Research Institute, KOSHA) ;
  • Jung, Sijeong (Occupational Safety & Health Research Institute, KOSHA) ;
  • Park, Doo Yong (Department of Mechanical Systems Engineering, Hansung University)
  • 이광용 (한국산업안전보건공단 산업안전보건연구원) ;
  • 이종한 (한국산업안전보건공단 산업안전보건연구원) ;
  • 정시정 (한국산업안전보건공단 산업안전보건연구원) ;
  • 박두용 (한성대학교)
  • Received : 2009.03.06
  • Accepted : 2009.05.04
  • Published : 2009.06.30

Abstract

Final purpose of this study was designed to develop the quality control(QC) sample for proficiency analytical testing of asbestos. This study consisted of two parts; first, development of manufacturing apparatus and sample preparing procedure for asbestos quality control(QC) sample: second, validation of the QC samples made by our developed method as asbestos proficiency analytical testing sample. The main results of the first part research are as followed We developed the apparatus for manufacturing the asbestos QC sample, consisted of filter hold, filter holder manifolder, vacuum system, and vacuum pump. The most proper filter of making the QC samples was a cellulose ester membrane filter with 25 mm diameter, pore size 0.8 um. And we presented the optimal procedure for preparing the asbestos QC sample by using the developed apparatus. We will verify the manufactured asbestos QC samples by this method, and present the validation results to confirm the reliability as a asbestos QC sample in next paper.

Keywords

References

  1. Niemeyer I.C, Gerchman LL. Preparation of asbestos filter samples for reference standard using a wet generation technique. American Industrial Hygiene Association Journal. 1981;42:757-759 https://doi.org/10.1080/15298668191420657
  2. Ortiz LW, Ettinger HJ, Fairchild CI. Calibration Standard for Counting Asbestos. American Industrial Hygiene Association Journal. 1975;36:104-112 https://doi.org/10.1080/0002889758507217
  3. Schlecht PC, Shulman SA. Performance of Asbestos fiber counting laboratories in the NIOSH Proficiency Analytical Testing(PAT) Program. American Industrial Hygiene Association Journal. 1986;47(5):259-269 https://doi.org/10.1080/15298668691389748
  4. Schlecht PC, Shulman SA. Phase contrast microscopy asbestos fiber counting performance in the Proficiency Analytical Testing Program. American Industrial Hygiene Association Journal. 1995;56:480-489 https://doi.org/10.1080/15428119591016908
  5. Pang TWS, Dicker WL, Nazer MA. An evaluation of the precision and accuracy of the direct transfer method for the analysis of asbestos fibers with comparison to the NIOSH method. American Industrial Hygiene Association Journal. 1984;45(5):329-335 https://doi.org/10.1080/15298668491399875
  6. Taylor DG, Baron PA. Shulman SA, Carter JW. Identification and counting of asbestos fibers. American Industrial Hygiene Association Journal. 1984;45(2):84-88 https://doi.org/10.1080/15298668491399415
  7. Baron PA, and Deye GJ. Generation of replicate asbestos aerosol samples for quality assurance. Applied Industrial Hygiene. 1994;2(3):114-118
  8. Skogstad A, Eduard W, Huser PO. A laboratory method for generation of replicate fiber samples of asbestos fibers in air. American Industrial Hygiene Association Journal.1996;57:741-745 https://doi.org/10.1080/15428119691014602
  9. Timbrell V, Rendall REG. Preparation of the UICC Standard Reference samples of asbestos. Powder Technology. 1971-1972;5:279-287 https://doi.org/10.1016/0032-5910(72)80032-9
  10. Timbrell V, Skidmore JW, Hyett AW, Wagner JC. Exposure chambers for inhalation experiments with standard reference samples of asbestos of the International Union Against Cancer(UICC). Aerosol Science. 1970;Vol 1:215-223 https://doi.org/10.1016/0021-8502(70)90035-2
  11. National Institute of Occupational Safety and Health(NIOSH). NIOSH Manual of Analytical Method. 4th ed. NIOSH;1994