Abstract
EAS (electronic article surveillance) systems have increased rapidly for article surveillance. In this paper, the strength of the EMF (electromagnetic fields) of EAS systems were measured. Pacemaker and ICD were investigated for inappropriate response resulting from EM (electromagnetic) EAS systems. The strength of EMF and the response of pacemaker and ICD were measured in the inner left side, outer right sides and the center of gates of the 6.3 kHz and 14.25 kHz EAS systems at a height of 130cm. As the result, EMF of the EAS system using 14.25 kHz was stronger than that of 6.3 kHz. AU interferences were observed only for 14.25 kHz, and the noisy ECG was found in three static positions on the pacemaker. The ICD resulted in noise reversion and VF (ventricular fibrillation) both static and moving positions by the EMP of 14.25 kHz EAS system. Therefore, it is necessary to post a message warning radiation of EMF from every EAS systems and possible risk of pacemakers and ICDs.
EAS (electronic article surveillance) 시스템은 상품의 도난 방지를 위한 것으로 상품 판매 자동화 증가로 인해 사용이 급속히 늘어나고 있다. 이에 본 연구에서는 EAS 시스템의 전자기장 세기를 측정하였고, EAS 시스템이 pacemaker와 ICD에 미치는 영향에 대해 알아보았다. 일반 성인 심장 높이인 130cm에서 6.3 kHz와 14.25 kHz EAS 시스템의 게이트 안쪽의 왼쪽, 바깥쪽의 오른쪽과 가운데 위치에서 전자기장의 세기를 측정하였고, pacemaker와 ICD의 동작 여부를 관찰하였다. 측정 결과 전자기장의 세기는 EAS 시스템에 따라 큰 차이가 났는데 14.25 kHz를 사용하는 EAS 시스템이 가장 강했다. Pacemaker는 고정시 세 위치 모두에서 파형 변화가 관찰되었고, ICD는 이동시와 고정시한 위치에서만 noise reversion, 심실세동이 관찰되었으며, 모두 전자기장 세기가 가장 센 14.25 kHz의 EAS 시스템의 전자기장에 의해 발생하였다. 이에 현재 국내에서는 시행되고 있지 않은 전자기장 방출에 대한 안내문 부착과 이식형 의료기기 시술을 받은 환자의 접근에 대한 위험성 경고문 등의 조치가 필요하다.