Abstract
Performance tests on the 3-Dimensional Flow-Structure-Interactions-Measurement System(FSIMS) have been carried out. Experimental data obtained by the FSIMS on a floating cylinder have been used to generate a set of artificial images. Comparisons between the data obtained by the use of the artificial images and those original experimental data have been made. Another set of artificial images have also been generated based on theoretically modulated sinusoidal motions, and comparisons between the data obtained by the use of these artificial images and the theoretical ones have been carried out. It has been verified that the FSIMS has a measurement uncertainty of 0.04-0.06mm/frame for velocity vectors and 0.002-0.01mm for the cylinder's positions.