Abstract

We describe a simple method to obtain an optical sectioning in a conventional wide-field microscope by projecting a single spatial frequency grid pattern onto the object. Using a patterned beam, we have measured the height of BGA with a rough surface that provide the coherence noise. The configuration of the height measurement system using pattern beam is simple. The image acquired by this system is not depend on the coherence noises. This system is also applicable to the sample reference plan that has no pattern on ground. The reappearance and accuracy are outstanding and applicable to many industrial optical metrology.

본 연구에서는 패턴 빔을 이용하여 표면이 거친 물체의 단차를 비접촉 방식으로 측정 하는 방법에 대하여 연구하였다. 본 연구에 사용된 방법은 장치가 매우 간단하고 스페클 노이지가 영상에 미치는 영향이 작아 재현성 면에서 매우 우수하였다. 특히 기준면에 문양이 없는 경우에는 기존의 자동 초점 측정 장치로는 측정하기 어려우나, 본 연구 방법에 의해 측정이 가능함을 보였다. 그리고 측정 시간이 매우 짧고 재현성이 좋으며 장치가 간단하여 산업 현장에서 응용하기 적절한 방법이다.

Keywords

References

  1. T. Dresel, G. Hausler, and H. Venzke, “Three-dimensional sensing of rough surfaces by coherence radar,” Appl. Opt. 31, 919-925 (1992) https://doi.org/10.1364/AO.31.000919
  2. S. Brinkmann, T. Dresel, R. Schreiner, and J. Schwider, “Testing of rod objects by grazing-incidence interferometry: experiment,” Appl. Opt. 38, 121-125 (1999) https://doi.org/10.1364/AO.38.000121
  3. T. Zhang and I. Yamaguchi, “Three-dimensional microscopy with phase-shifting digital holography,” Opt. Lett. 23, 1221-1223 (1998) https://doi.org/10.1364/OL.23.001221
  4. I. Yamaguchi, J. I. Kato, S. Otha, and J. Mizuno, “Image formation in phase-shifting digital holography and applications to microscopy,” Appl. Opt. 40, 6177-6186 (2001) https://doi.org/10.1364/AO.40.006177
  5. T. Wilson, Confocal Microscopy (Academic Press, London, UK, 1990)
  6. S. Zhang and P. Huang, “High-resolution, real-time 3-D shape measurement,” Opt. Eng. 45, 123601-123607 (2006) https://doi.org/10.1117/1.2402128
  7. S. Kumar, D. Snyder, D. Duncan, J. Cohen, and J. Cooper, “Digital preservation of ancient cuneiform tablets using 3D-scanning,” in Proc. Fourth International Conference on 3-D Digital Imaging and Modeling (Banff, Alberta, Canada,2003) pp. 326-333 https://doi.org/10.1109/IM.2003.1240266
  8. P. Hariharan and Z. S. Hegedus, “Reduction of speckle in coherent imaging by spatial frequency sampling,” Opt. Acta. 21, 345-356 (1974) https://doi.org/10.1080/713818900
  9. N. Bertaux, Y. Frauel, P. Refregier, and B. Javidi, “Speckle removal using a maximum likelihood technique with isoline gray-level regularization.” J. Opt. Soc. Am. A 21, 2283-2291 (2004) https://doi.org/10.1364/JOSAA.21.002283
  10. S. P. Kelly, J. E. Ward, U. Gopinathan, and J. T. Sheridan, “Controlling speckle using lenses and free space,” Opt. Lett. 32, 3394-3396 (2007) https://doi.org/10.1364/OL.32.003394
  11. L. Firestone, K. Cook, K. Gulp, N. Talsania, and K. Preston, “Comparison of autofocus methods for automated microscopy,” Cytometry 12, 195-206 (1991) https://doi.org/10.1002/cyto.990120302
  12. X. Y. Liu, W. H. Wang, and Y. Sun, “Dynamic evaluation of autofocusing for automated microscopic analysis of blood smear and pap smear,” J. Microsc. 227, 15-23 (2007) https://doi.org/10.1111/j.1365-2818.2007.01779.x
  13. Y. Liron, Y. Paran, G. Zatorsky, B. Geiger, and Z. Kam, “Laser autofocusing for high-resolution cell biological imaging,” J. Microsc. 221, 145-151 (2006) https://doi.org/10.1111/j.1365-2818.2006.01550.x
  14. M. A. A. Neil, R. Juskaitis, and T. Wilson, “Method of obtaining optical sectioning by using structured light in a conventional microscope,” Opt. Lett. 22, 1905-1907 (1997) https://doi.org/10.1364/OL.22.001905
  15. J. N. Helbert, Handbook of VLSI Microlithography: Principles, Technology, and Applications (Noyes Press, NY, USA, 2001), Chapter 5
  16. J. R. Sheats and B. W. Smith, Microlithography: Science and Technology (CRC press, NY, USA, 1998), Chapter 3
  17. M. A. A. Neil, R. Juˇskaitis, and T. Wilson, “Method of obtaining optical sectioning by using structured light in a conventional microscope,” Opt. Lett. 22, 1905-1907 (1997) https://doi.org/10.1364/OL.22.001905