Development of QA Phantom Prototype for Imaged Based Radiation Treatment System

영상기반 방사선 치료기기를 위한 QA 팬텀 시작품 개발

  • Chang, Jin-A (Department of Biomedical Engineering, The Catholic University of Korea) ;
  • Oh, Seoung-Jong (Department of Biomedical Engineering, The Catholic University of Korea) ;
  • Jung, Won-Kyun (Department of Biomedical Engineering, The Catholic University of Korea) ;
  • Jang, Hong-Suk (Department of Radiation Oncology, Kangnam St. Mary's Hospital, The Catholic University of Korea) ;
  • Kim, Hoi-Nam (Department of Radiation Oncology, Kangnam St. Mary's Hospital, The Catholic University of Korea) ;
  • Kang, Dae-Gyu (Department of Radiation Oncology, St. Vincent's Hospital, The Catholic University of Korea) ;
  • Lee, Doo-Hyun (Department of Radiation Oncology, National Cancer Center) ;
  • Suh, Tae-Suk (Department of Biomedical Engineering, The Catholic University of Korea)
  • 장지나 (가톨릭대학교 의과대학 의공학교실) ;
  • 오승종 (가톨릭대학교 의과대학 의공학교실) ;
  • 정원균 (가톨릭대학교 의과대학 의공학교실) ;
  • 장홍석 (가톨릭대학교 강남성모병원 방사선종양학과) ;
  • 김회남 (가톨릭대학교 강남성모병원 방사선종양학과) ;
  • 강대규 (가톨릭대학교 성빈센트병원 방사선종양학과) ;
  • 이두현 (국립암센터 방사선종양학과) ;
  • 서태석 (가톨릭대학교 의과대학 의공학교실)
  • Published : 2008.06.30

Abstract

In this study, we developed the protopype of QA phantom for image QA including an additional component for image based radiation treatment system. The new phantom considered two main parts: Image quality and fusion accuracy. Image quality part included for daily CT number linearity and spatial resolution, and fusion accuracy part designed to simulate a simple translation-rotation setting. The CT scans of the phantom obtained from conventional CT, MVCT of Tomotherapy unit, and both image sets were satisfied the recommendation of spatial resolution. This phantom was simple and efficient for daily imaging QA, and it is important to provide a new concept of verification of image registration.

본 연구에서는 영상 기반 방사선 치료기기의 영상 QA를 위한 필요한 추가적인 요소를 포함시킨 QA 팬텀 시제품을 개발하였다. 개발된 팬텀은 크게 영상 질 평가 부분과 정합 정확도 평가 부분의 두 가지로 이루어져 있다. 영상 질 평가를 위해 CT number 선형성과 공간 해상도 평가가 가능하게 설계되었고, 정합 정확도 평가를 위하여 간단한 이동 및 회전조절이 가능하게 설계되었다. 제작된 팬텀의 영상은 일반 CT와 Tomotherapy의 MVCT에서 획득되었고, 두 영상 모두 해상도 평가 기준을 충족하였다. 이 팬텀은 일간 영상 QA를 위한 간단하고, 효율적이며, 그리고 영상 정합 평가의 새로운 개념을 제시했다는 데에 큰 의의가 있다.