전류변성기 비교기를 이용한 인덕터의 절대 평가

Absolute Evaluation of Inductor Using Current Transformer Comparator

  • 김윤형 (충남대학교 전기공학과) ;
  • 정재갑 (한국표준과학연구원 기반표준부 전자기그룹) ;
  • 한상길 (충남대학교 전기공학과) ;
  • 김한준 (한국표준과학연구원 전자기센터) ;
  • 한상옥 (충남대학교 전기공학과)
  • 발행 : 2008.09.01

초록

We have developed two absolute evaluation technology of inductor using current transformer (CT) comparator. One is the method that the reactance of inductor is obtained by analysing the equivalent circuit of CT with inductor connected to series at secondary terminal of CT. The other is the method that the reactance of inductor is obtained by comparing phase displacement of current flowing on inductor by using CT comparator. These technologies have the advantage to apply up to rated current and voltage of inductor. The method was applied to inductors under test in the range of $100 {\mu}H{\sim}1\;H$. The inductance of the inductor under test obtained in this study are consistent with those measured by LCR meter using the same inductor within an expanded uncertainty (k = 2) in the overall range of inductance.

키워드

참고문헌

  1. Agilent, "Agilent 4980A Precision LCR Meter, 20 Hz to 1 MHz", Agilent Technologies, 2008. (www.agilent.com)
  2. 김한준, "인덕턴스 브리지 개발", 한국표준과학연구원, 1999
  3. 정재갑, 이상화, 강전홍, 김명수, 김윤형, 한상길, 한상옥, "20,000 A 전류변성기 국가표준 시스템 구축", 대한전기학회지, 57P, 6-13, 2007
  4. J. D. Ramboz and O. Petersons, "A calibration service for current transformers", NIST Special Publication, vol. 250, No 36, p60, 1991
  5. 한국표준협회 "측정 결과의 불확도 추정 및 표현을 위한 지침", KS A 3000, 2005
  6. 정재갑, 이상화, 권성원, 김명수, "전류변성기 비교기와 정밀션트저항을 이용한 전류변성기용 부담의 평가기 술", 대한전기학회지, 55C, 250-6, 2006
  7. Tinsley, "Precision Resistance Measurement", Tinsley Precision Instruments, 1995. (www.tinsley.co.uk)