초록

A system is designed and fabricated for the automatic low temperature calibration by comparison of industrial resistance thermometers. The cryostat is suitable for the calibration of four capsule type thermometers and four long stem thermometers. By the automatic temperature control of the cryostat, we could calibrate thermometers from $-200^{\circ}C$ to $0^{\circ}C$ for ${\sim}15$ hours by one fill of liquid nitrogen in the test run of the system. The uncertainty of the calibration for industrial platinum resistance thermometers using the automatic system is about 30mK with a 95.% confidence interval.

키워드

참고문헌

  1. H. Preston-Tomas, "The international temperature scale of 1990 (ITS-90)", Metrologia, vol. 27, no. 1, pp. 3-10, 1990 https://doi.org/10.1088/0026-1394/27/1/002
  2. B.W. Mangum, P. Bloembergen, M. V. Chattle, B. Fellmuth, P. Marcarino and A. I. Pokhodun, "On the international temperature scale of 1990 (ITS-90)", Metrologia, vol. 34, no. 5, pp. 427-429, 1997 https://doi.org/10.1088/0026-1394/34/5/8
  3. J. Ancsin and J. M. Philips, "Calibration apparatus for long-stem and capsule-type Pt resistance thermometers between the triple points of Ar and In", Rev. Sci. Instr., vol. 55, no. 8, pp. 1321-1314, 1984 https://doi.org/10.1063/1.1137924
  4. 강기훈, 김용규, 감기술, "캡슐형 백금저항온도계 국제비교를 위한 네온, 산소, 아르곤, 수은 및 물의 삼중점 실현", 센서학회지, 제9권, 제3호, pp. 153-162, 2000
  5. 양인석, 송창호, 강기훈, 김용규, "초정밀 온도센서의 저온 영역 교정 장치 개발", 센서학회지, 제17권, 제3 호, pp. 188-194, 2008 https://doi.org/10.5369/JSST.2008.17.3.188
  6. ASTM Standard E 1137/E 1137M - 04, "Standard specification for industrial platinum resistance thermometers", ASTM International, West Conshohocken, PA, www.astm.org, 2004