A Study on the Accuracy of the Electronic Apex Locator Using a Micro-Computed Tomography

미세단층촬영기(Micro-CT)를 이용한 전자 근관장 측정기의 정확성에 관한 연구

  • Published : 2007.08.01

Abstract

The length of root canal has to be measured for endodontic treatment. Several electronic apex locators were developed to measure the length of root canal by other researchers. And their accuracies were verified by X-ray or micrometer method. But these methods did not consider the non-linear bends of pulp and had ${\pm}0.5mm$ error which was large enough to measure the length of root canal. The purpose of this study is the introduction of a new method to measure the length of root canal and the verification of the accuracy of an electronic apex locator using a Micro-CT. The length of root canal of 6 teeth were measured with the electronic apex locator. When the electronic apex locator reads 0.5, 0.6, 0.7, 0.8, 0.9 and 1.0 mm length of the file which was inserted in the hole of the tooth to measure the length of root canal. The average (${\pm}$Standard deviation) length of root canal of 6 teeth measured by the Micro-CT was $0.49{\pm}0.03,\;0.59{\pm}0.04,\;0.68{\pm}0.03,\;0.78{\pm}0.03,\;0.90{\pm}0.04\;and\;1.01{\pm}0.03mm$, respectively. The maximum error of the electronic apex locator was 0.06 mm.

Keywords

References

  1. Johnson, W. T., 'Color Atlas of Endodontics,' Daehan Narae Publishing, pp. 67-98, 2003
  2. Nam, K. C., Kim, S. C., Lee, S. J., Kim, Y. J., Kim, N. G. and Kim, D. W., 'Root Canal Length Measurement in Teeth with Electrolyte Compensation,' Med. Biol. Eng. Comput., Vol. 40, No. 2, pp. 200-204, 2002 https://doi.org/10.1007/BF02348125
  3. Sunada, I., 'New Method for Measuring the Length of the Root Canal,' Journal of Dent. Res., Vol. 41, pp. 376-387, 1962
  4. Arens, D. E., 'Surgical Endodontics for Clinician,' Daehan Narae Publishing, pp. 3-22, 2002
  5. Chun, K. J., Lee, H. J. and Chung, D. T., 'A Study of the Standardization in the Mandibular First Premolar of the Middle Aged Korean,' J. of the KSPE, Vol. 23, No.2, pp.154-163, 2006
  6. Yamaoka, M., Yamashita, Y. and Saito, T., 'Electrical Root Canal Measuring Instrument Based on a New Principlem,' Tokyo, Nihon University School of Dentistry, 1989
  7. Hwang, H. K, Shin, Y. G. and Kim, P. S., 'A Study on the Accuracy of the Root-ZX According to the Various Conditions of Root Canals,' The Korean Academy of Conservative Dentistry, Vol. 25, No. 3, pp. 477-481, 2000
  8. Kim, E. S. and Lee, S. J., 'Electronic Apex Locator,' Dent. Clin. N. Am., Vol. 48, No. 1, pp. 35-54, 2004 https://doi.org/10.1016/j.cden.2003.10.005
  9. Shin, J. W., 'Dental Morphology,' Jungmoonkak publish, second editon, 1995
  10. Kobayashi, C., Matoba, K., Suda, H. and Sunada, I., 'New 'Practical Model of the Division Method Electronic Root Canal Length Measuring Device,' Journal of Jpn. Endodon Assoc., Vol. 12, pp. 143-148, 1991