Study on the Exposure Dose(mAs) and acquisition Image set up Density Display and Sensitivity of control Panel for the Digital Flat-Panel-Detector

디지털 평판형 검출기에서 Control Panel의 Density Display와 Sensitivity 설정이 조사선량(mAs)과 획득영상에 미치는 영향에 관한 연구

  • Kim, Byung-Ki (Department of Radiology, Konkuk University Hospital) ;
  • Kim, Sang-Keun (Department of Radiology, Konkuk University Hospital) ;
  • Cha, Seon-Hwa (Department of Radiology, Konkuk University Hospital) ;
  • Choi, Jun-Gu (Department of Radiology, Konkuk University Hospital) ;
  • Lee, Jun (Department of Radiology, Konkuk University Hospital) ;
  • Kim, Min-Woo (Department of Radiology, Konkuk University Hospital) ;
  • Kim, Sun-Bae (Department of Radiology, Konkuk University Hospital) ;
  • Kim, Gyeong-Sun (Department of Radiology, Konkuk University Hospital)
  • 김병기 (건국대학교병원 영상의학과) ;
  • 김상건 (건국대학교병원 영상의학과) ;
  • 차선화 (건국대학교병원 영상의학과) ;
  • 최준구 (건국대학교병원 영상의학과) ;
  • 이준 (건국대학교병원 영상의학과) ;
  • 이민우 (건국대학교병원 영상의학과) ;
  • 김순배 (건국대학교병원 영상의학과) ;
  • 김경수 (건국대학교병원 영상의학과)
  • Published : 2007.12.30

Abstract

The purpose to recognize change of average pixel value of acquisition image by control panel's density and right set up method of speed (sensitivity) and exposure dose(mAs) change that dose in purpose digital flatpanel-detector. X -ray generator DHF-158H2(Hitachi, Japan). Detector CXDI 4OG(Canon, Japan), 12 : 1 grid and exposure ray 135 kVp, 250 mA, 10 ms. focus-detector distance 180 cm and used AEC mode. DICOM reflex analysis program used image J that is digital reflex analysis program that offer in United States America National Health Center(National Institutes of Health : NlH) phantom used chest phantom(Anthromorphic : Flukebrome.medicaI USA). An experiment chest phantom that consist by formation equivalence material use because density value( -3${\sim}$+3) in X-ray control panel and seep that is speed step(slow, medium, fast) each control experimentalize. image analysis reflex neted through an experiment using image j each image compare. These was change in dose according to slow, medium, fast and density's change in an experiment result. According to detector sensitivity and density condition set, dose was relationship dissimilarity 500% from 200%. The dose came highest when is density +3 to slow. and dose more increases gray scale's extent could know that rise. Could know whether how equipment set is important through this experiment. cause of disease which change by digital radiography system forward is thought to increase more, it is considered that suitable education by this and continuous interest about equipment need absolutely.

Keywords