The Study on Clinical Conditions and Skin Dose of Upper-Gastrointestinal X-ray Fluoroscopy

위장 X선 투시검사에 따른 실태 및 선량에 관한 연구

  • Kim, Sung-Chul (Department of Radiological Science, Gachon University of Medicine & Science) ;
  • Ahn, Sung-Min (Department of Radiological Science, Gachon University of Medicine & Science) ;
  • Jang, Sang-Sup (Department of Radiological Technology, Deajeon Health Sciences College)
  • 김성철 (가천의과학대학교 방사선학과) ;
  • 안성민 (가천의과학대학교 방사선학과) ;
  • 장상섭 (대전보건대학 방사선과)
  • Published : 2007.03.31

Abstract

This study examined present conditions of upper-gastrointestinal X-ray fluoroscopy and patient skin dose. The authors elected 21 equipments to check the X-ray equipment and exposure factor of fluoroscopy & spot exposure in university hospitals, hospitals, and clinics where perform upper-gastrointestinal X-ray fluoroscopy more than five times every day in Incheon areas. The amount of patient's skin dose during upper-gastrointestinal X-ray fluoroscopy was measured by ionization chamber.

위암은 $1999{\sim}2001$년 국내 암발생율 1위로 조기진단이 중요시 된다. 위암의 검사방법 중 하나인 위장 X선 투시검사는 중요성이 높기 때문에 저자 등은 인천지역을 중심으로 종합 대학병원, 병원, 개인의원급 21대의 X선장치를 선정하여 X선장치의 현황과 투시검사 및 저격촬영시의 검사조건 등을 조사하였다. 또한 검사중 환자표면선량을 전리조를 이용하여 측정하였다. 본 연구는 위장 X선 투시검사의 실태 및 환자표면선량의 정도를 파악하여 그 현황을 알리고자 실시하였다.