A study on the radiation exposure of simple abdomen Radiation in Radiography

복부 단순 방사선 검사 시 피폭선량에 대한 연구

  • Yeo, Jin-Dong (Department of Health Sciences Kosin University Graduate School) ;
  • Kim, Mi-Sook (Department of Health Sciences Kosin University Graduate School)
  • 여진동 (고신대학교 대학원 보건과학과) ;
  • Published : 2007.12.30

Abstract

This study was performed to measure about exposure dose during simple abdmon radiation radiography. The exposure dose was measured by PDD, surface dose, respectively. The result was as followed: 1. When tube voltage were increased with 60-85kv, surface dose were increased. When FFD(focus film distance) at the 50-150cm and mAs were increased, surface dose were decreased. 2. The percentage depth dose(PDD) were appeared 50% below depth dose at 4cm with 60-75kv, and 6cm depth with 80-85kv, 5% below depth dose at 12cm with 60kv, and depth with 65-85kv. 3. The percentage forward scatter increased from 10% to 11.78% at the 60-85kv. The back scatter dose were increase from 25% to 37% at the 60-85kv. The side scatter dose were affected to heel effect.

복부 단순방사선 촬영은 흉부촬영에 이어 가장 많이 검사가 이루어지고 있는 실정인 것이 임상현장이므로 환자의 피폭선량에 대한 연구가 필수적으로 진행이 이루어 질 필요성이 대두가 된다. 그 결과는 다음과 같았다. 1. 관전압이 60-85kV 증가 시 표면선량은 증가하고, 촬영거리를 50-150cm로 20cm씩 증가함에 mAs도 증가시킨 결과 표면선량은 감소되었다. 2. 심부선량 백분율은 60-75kV에서는 4cm 깊이, 80-85kV에서는 6cm 깊이에서 50% 이하를 나타내었으며, 60kV에서는 12cm 깊이, 65-85kV에서는 14cm 깊이에서 5% 이하를 나타났다. 3. 전방산란율은 60-85kV까지 관전압이 증가 시 10-11.78%까지 증가함을 나타내었다. 후방 산란율은 관전압이 60-85kV 증가 시 25-37%로 증가하였다. 측방산란율은 음극 측에서는 관전압이 60-85kV 증가 시 1.63-2.95%로 완만하게 증가하나 양극측은 그 반대로 감소하는데 그 원인은 heel effect 현상 때문인가 사료된다. 후방산란율이 가장 크고, 그 다음이 전방산란율, 측방산란율 순으로 작아짐을 알 수 있다.