A Development of Eddy Current Sensor System for An Axial-flow type Blood Pump with The Magnetic Bearing

축류형 인공심장의 자기베어링 제어를 위한 와전류 센서 시스템 개발

  • Ahn, C.B. (Biomedical Engineering of Brain Korea 21 Program, Korea University) ;
  • Moon, K.C. (Korea Artificial Organ Center, Korea University) ;
  • Jeong, G.S. (Biomedical Engineering of Brain Korea 21 Program, Korea University) ;
  • Nam, K.W. (Korea Artificial Organ Center, Korea University) ;
  • Lee, J.J. (Korea Artificial Organ Center, Korea University) ;
  • Sun, K. (Department of Biomedical Engineering College of Medicine, Korea University)
  • 안치범 (고려대학교 BK21 의용생체공학) ;
  • 문기철 (고려대학교 한국인공장기센터) ;
  • 정기석 (고려대학교 BK21 의용생체공학) ;
  • 남경원 (고려대학교 한국인공장기센터) ;
  • 이정주 (고려대학교 한국인공장기센터) ;
  • 선경 (고려대학교 의과대학 의공학교실)
  • Published : 2007.04.30

Abstract

The axial-flow type blood pump(XVAD) which has been developed in our group consists of mechanical parts (an impeller, a diffuser and a flow straightener) and electrical parts (a motor and a magnetic bearing). The magnetic bearing system fully levitates the impeller to remove mechanical coupling with other parts of the pump with constant gap, which needs non-contact type gap sensing. Conventional gap sensors are too large to be adopted to the implantable axial -flow type blood pump. Thus, in this paper, the compact eddy current type gap sensor system proper for the implantable axial-flow type blood pump was developed and its performance was evaluated in vitro. The developed eddy current type gap sensor system is a transformer type and has a differential probe. Sensor coil(probe) has small dimensions(6 mm diameter, 2 mm thickness) and its optimal inductance was determined as 0.068 mH for the measurement range of $0\sim3mm$. It could be manufactured with 130 turns of the 0.04 mm diameter copper coil. The characteristics of the developed eddy current type gap sensor system was evaluated by in vitro experiment. At experiment, it showed satis(actory performance to apply to the magnetic bearing system of the XVAD. It could measure the gap up to 3mm, but the linearity was decreased at the range of $1.8\sim3.0mm$. Moreover, it showed no difference in different media such as the water and the blood at the temperature range of $35\sim40^{\circ}C$.

Keywords

References

  1. Min Woo Jung, Chang Mo Hwang, Gi Seok Jeong, Jung Soo kang, Chi Bum Ahn, et al, 'A study of electromagnetic actuator for electro-pneumatic driven ventricular assist device,' J. Biomedical Engineering Research, vol. 26, no. 6, pp. 293-298, Dec. 2005
  2. XINWEI SONG, AMY L. THROCKMORTON, ALEXANDRINA UNTAROIU, SONNA PATEL, PAUL E. ALLAIRE, et al., 'Axial flow blood pumps,' J. ASAIO, vol. 49, pp.355-364, 2003
  3. Myounggyu Noh, Min-Kyoung Jeong, 'The effects of eddy currents and hysteresis on the performance of inductive position sensor for magnetic bearing,' The Korean Society of Mechanical Engineers Conference, vol. 1, no. 2, pp.36-41, 11. 2001
  4. Hideo Hoshi, Kiroyuki Katakoa, Katsuhiro Ohuchi, Juni Chi Asama, Tadahiko Shinshi, Akira Shimokohbe, And Setsuo Takatani, 'Magnetically suspended centrifugal bloos pump with a radial magnetic driver,' J. ASAIO, vol. 51. pp. 60-64. 2005 https://doi.org/10.1097/01.MAT.0000151148.72891.55
  5. G. S. Jeong, C. M. Hwang, J. J. Lee, M. W. Jung, J. S. Kang, et al. 'Development of the small size impeller for axial flow artificial heart: KX -VAD' The 32th Korea Society of Medical&Biological Engineering conference. pp75. May. 2005
  6. E.J. Lee. 'Nondestructive Evaluation VI Eddy Current Testing' Gold 1999
  7. Y. M. Jung, 'Eddy current testing(II),' Journal of the Korean Society for Nondestructive Testing, vol. 13, no. 3, 1993
  8. S. D. Kim, J. M. Shim. 'Impedance analysis and experimental study of a solenoid eddy current sensor to detect the cross-sectional area of non-ferromagnetic stranded conductors,' Journal of the Korean Society of Sensor, vol. 6, no. 2, Mar. 1997