Evaluation of Usefulness in New Immobilization Device with 3D CT Angiography for Lower Extremity
전산화단층촬영검사에서 하지의 3D 혈관조영검사를 위한 새로운 고정기구의 유용성 평가
- Published : 2007.04.28
Abstract
Our objective was to develop and evaluate a non-invasive device for rigid immobilization and surface disease non-contact of the table in the lower extremity during CT angiography. The immobilization device consists of two components. The patient's lower limb device stabilizing elements made of polyethylene resin soft materials, and pelvis parts foam pad is used for non-contact surface. In a Prospective study the lower extremity device was used in patients who underwent a CT angiography of the lower limb. Immobilization with our device was well tolerated by all patients. The quality of the resulting images in the popliteal and infrapopliteal region was rated by five-point scale. The rigid immobilization resulted in a complete absence of motion artifacts. The new device is an effective, well tolerated and easily used immobilization that is suitable of use in 3D lower extremity CT angiography.
3차원 전산화단층촬영 검사에서 하지 혈관조영술을 위해서는 움직임이 없어야 한다. 표재질환의 검사는 환자의 하지가 테이블에 접촉하지 않도록 해야 한다. 새로운 고정기구를 하지 조영술을 위해 새롭게 개발 되었다. 본 연구에서는 새로운 하지조영술용 고정기구를 고안하여 개발하였다. 개발된 고정기구는 2개로 구성되었고, 폴리우레탄 레진의 재료를 이용하였다. 둔부에 위치하는 패드와 족부를 고정할 수 있도록 만들었다. 새로운 고정기구를 100명의 환자를 대상으로 전산화단층촬영영상을 획득하였으며 이들 영상을 5등급으로 구분하여 고정기구를 사용하지 않은 환자와 영상를 비교하여 평가하였다. 고정기구를 사용하여 전산화단층촬영으로 하지 혈관조영검사를 시행한 경우 매우 편안하고 움직임이 없이 혈관을 묘출하는데 효과적으로 사용됨을 확인하였다.