이중나사산 임플란트의 안정성에 대한 평가 - 오스텔 멘토를 이용한 초기 안정성 ; PART I

Evaluation of Stability of Double Threaded Implant-Emphasis on Initial Stability Using Osstell MentorTM; Part I

  • 김시엽 (부산대학교 치의학전문대학원 구강악안면외과학 교실) ;
  • 김병국 (오스템 임플란트 연구소) ;
  • 허진호 (부산대학교 치의학전문대학원 구강악안면외과학 교실) ;
  • 이주연 (부산대학교 치의학전문대학원 치주과학 교실) ;
  • 정창모 (부산대학교 치의학전문대학원 보철과학 교실) ;
  • 김용덕 (부산대학교 치의학전문대학원 구강악안면외과학 교실)
  • Kim, Si-Yeob (Department of Oral & Maxillofacial Surgery, School of Dentistry, Pusan National University) ;
  • Kim, Byung-Kook (Osstem R&D Center) ;
  • Heo, Jin-Ho (Department of Oral & Maxillofacial Surgery, School of Dentistry, Pusan National University) ;
  • Lee, Ju-Youn (Department of Periodontology, School of Dentistry, Pusan National University) ;
  • Jeong, Chang-Mo (Department of Prosthodontics, School of Dentistry, Pusan National University) ;
  • Kim, Yong-Deok (Department of Oral & Maxillofacial Surgery, School of Dentistry, Pusan National University)
  • 발행 : 2007.12.30

초록

Fixture의 안정도를 평가하는 장비인 Osstell Mentor와 기존 Osstell과의 비교 시험 및 측정환경에 따른 노이즈 발생 시험 결과를 통해 다음과 같은 결과를 얻었다. 1. US II Fixture와 SS II Fixture의 ISQ 값을 측정한 1차 검증에서는 Osstell과 Osstell Mentor의 ISQ 값 경향의 차이가 크게 나타나지 않음을 확인하였다. 2. SS II Fixture를 얕게 매식한 것과 정상적인 매식한 것의 ISQ 값 비교를 한 2차 검증에서는, Osstell과 Osstell Mentor 모두 매식 깊이에 따른 동일한 경향의 결과를 보였다. 3. 3차 검증에서는 골질별로 Osstell과 Osstell Mentor의 ISQ 값이 일정한 경향을 보이는 것을 확인하였고, 특히 Osstell Mentor 장비는 Osstell 장비보다 사용함에 있어서 편리성이 뛰어나다고 판단된다. 또한 ISQ 값을 100% 신뢰하기는 어렵지만 Osstell Mentor를 임상 및 실험에 적용하는데 있어서는 문제가 없을 것으로 판단된다. 4. 측정환경에 따라 전자파를 발생시키는 환경이 존재하면 노이즈가 발생할 수 있으나, 실제 측정결과 휴대폰이 작동 중인 환경을 제외하고는 노이즈가 발생하지 않았다. 실제 치과 내에서의 사용에서도 노이즈는 발생하지 않음을 확인하였다. 5. 비교적 안전하게 장치를 장착 및 탈착 할 수 있어 측정과정에 의해 발생되는 임플란트와 주위골에 대한 잠재적 위험도가 낮고 측정위치에 따른 오차가 적어 임상적으로 불리한 상황과 다양한 환경에서의 임플란트 안정성 측정이 가능해 임상적 활용도가 높다고 판단된다.

Purpose This study was planned to compare and evaluate the stability of implant using $Osstell^{TM}$ and Osstell $Mentor^{TM}$. Material and methods Artificial bone and RBM(resorbable blasting media) surface blasted implants(Osstem US II, SS II implants - diameter: 4mm, length: 13mm) were used. To measure the stability of installed implants, $Osstell^{TM}$ and Osstell $Mentor^{TM}$ were used. In the first experiment, five implants were installed in D1(external type implants) and D3(internal type implants). In the second experiment, 4 internal type implants were divided in two groups and installed in D1 artificial bone with different depth. In the third experiment, two external implants were installed in D1 and D3 artificial bone each and two internal implants were installed in D1 and D3 artificial bone. In all groups, their stability were measured by $Osstell^{TM}$ and Osstell $Mentor^{TM}$. Results In all groups, $Osstell^{TM}$ and Osstell $Mentor^{TM}$ both showed reliable measurement values. The value difference between $Osstell^{TM}$ and Osstell $Mentor^{TM}$ was observed but the difference was small and clinically acceptable. Conclusion These results suggest that the use of Osstell $Mentor^{TM}$ has clinical relevance in the assessment of implant stability.

키워드

참고문헌

  1. Meredith N. Assessment of implant stability as a prognostic determinant. Int. J Prosthodont 1988;11:491-501
  2. Sennerby L. Meredith N. Resonance frequency analysis: Measuring implant stability and osseointegration. Compendium 1998:19:493-502
  3. Meredith N, Alleyne D, Cawley P. Quantitative determination of the stability of the implant-tissue interface using resonance frequency analysis. Clin Oral Impl Res 1996;7:261-7 https://doi.org/10.1034/j.1600-0501.1996.070308.x
  4. Barewal RM, Oates TW, Meredith N, Cochran DL. Resonance frequency measurement of implant stability in vivo on implants with a sandblasted and acidetched surface. Int J Oral Maxillofac Implants 2003: 18: 641 - 651
  5. Manz M, Morris HF, Shigeru O. An evaluation of the periotest system. Part I examiner reliability and repeatability of readings. Implant Dent 1992: 1: 142-146 https://doi.org/10.1097/00008505-199205000-00006
  6. Stenport VF, Olsson B, Morberg P, Tomell J, Johansson CB. Systemically administered human growth hormone improves initial implant stability: an experimental study in the rabbit. Clin Implant Dent Relat Res 2001: 3: 135 - 141 https://doi.org/10.1111/j.1708-8208.2001.tb00133.x
  7. Derhami K, Woolfardt JF, Faulkner G, Grace M. Assessment of the Periotest device in baseline mobility measurements of craniofacial implants. Int J Oral Maxillofac Implants. 1995: 10: 221 - 229