THE EFFECT OF AUTOCLAVE STERILIZATION AND REUSE OF $SMARTPEG^{TM}$ ON THE IMPLANT STABILITY QUOTIENT (ISQ) MEASUREMENT

$Smartpeg^{TM}$의 고압멸균소독 및 재사용이 임플랜트 안정성 지수(ISQ) 측정에 미치는 영향

  • Kang, In-Ho (Department of Prosthodontics, Graduate School, Seoul National University) ;
  • Kim, Myung-Joo (Department of Prosthodontics, Graduate School, Seoul National University) ;
  • Lim, Young-Jun (Department of Prosthodontics, Graduate School, Seoul National University) ;
  • Kim, Chang-Whe (Department of Prosthodontics, Graduate School, Seoul National University)
  • 강인호 (서울대학교 치과대학 치과보철학교실) ;
  • 김명주 (서울대학교 치과대학 치과보철학교실) ;
  • 임영준 (서울대학교 치과대학 치과보철학교실) ;
  • 김창회 (서울대학교 치과대학 치과보철학교실)
  • Published : 2007.10.31

Abstract

Statement of problem: Resonance frequency analysis is evaluated as the non-invasive and objective method for the evaluation of implant stability and has been increasingly used. It is necessary to evaluate the factors affect the ISQ measurement stability. Purpose: This study was performed to evaluate the effect of the autoclave sterilization and reuse of $Smartpeg^{TM}$ on ISQ measurement. Material and methods: $SmartPeg^{TM}$ (Integration Diagnostics Ltd., $G\ddot{o}teborg$, Sweden) of autoclave group (A) was autoclave sterilized 9 times and $Smartpeg^{TM}$ of reuse group (R) was reused 9 times. Ten $SmartPeg^{TM}s$ were allocated to each group and after each autoclave sterilization and reuse, implant stability quotient (ISQ) values were measured 3 times from the two directions a and b at a right angle. $Osstell^{TM}$ mentor (Integration Diagnostics Ltd. $G\ddot{o}teborg$, Sweden) was used and type 1 (article no. 100353) $Smartpeg^{TM}$ was selected according to $Smartpeg^{TM}$ reference list. Osstem Implant US II future (Osstem Co., Seoul, Korea) in $4.0mm{\times}11.5mm$ was embedded in the self-curing acrylic resin ($Orthojet^{(R)}$, Lang Dental, U.S.A.). Data was statistically analyzed by one-way ANOVA $({\alpha}=.05)$ and scheffe test was done where a significant difference exist. Correlation test was also done between ISQ value and the number of autoclave sterilization or reuse. Results: 1. In autoclave group, the means and sd. of ISQ value before autoclave sterilization were $84.97{\pm}0.41,\;84.93{\pm}0.74$ at direction a and b. There was significant differences between autoclave groups at direction a and b (P=.000). 2. In reuse group, the means and sd. of ISQ value before reuse were $85.40{\pm}0.62,\;85.50{\pm}0.57$ at direction a and b. There was no significant difference between reuse groups at direction a and b (P>.05). 3. There was a weak positive correlation between the number of reuse and ISQ value at direction a and b (${\gamma}=.207$ and .246, P<.01). Conclusion: Within the limitations of this study, the following conclusions were drawn. Till ninth reuse of $Smartpeg^{TM}$, the ISQ measurement stability did not be affected. After twice autoclave sterilization of $Smartpeg^{TM}$ the ISQ measurement stability was affected.

Keywords

References

  1. Sennerby L, Meredith N. Resonance Frequency Analysis: Measuring Implant Stability and Osseointegration. Compendium 1998:19(5):493-502
  2. Roberts WE, Misch CE. Contemporary Implant Dentistry. St. Louis: CV Mosby: 1993. p.327-54
  3. Lekholm U, Zarb GA, Branemark PI, Albrektsson T. Tissue-Integrated Prostheses. Osseointegration in Clinical Dentistry. Chicago: Quintessence Int: 1985. p.199-210
  4. Strid KG, Branemark PI, Zarb GA, Albresktsson T. Tissue-Integrated Prostheses. Osseointegration in Clinical Dentistry. Chicago: Quintessence Int: 1985. p.187-198
  5. Schulte W et al. Ein neues Verfahren und Cerat zur Messung der Function des Parodontiums. Zahnartzl Mitt. 1983:73:12-29
  6. Teerlinck J, Quirynen M, Darius P, Van Steemberghe D. An Obejective clinical diagnosis of bone apposition toward implants. Int J Oral Maxillofac Implants 1991:6:55-61
  7. Derharmi K, Wolfaardt JF, Faulkner G, Grace M. Assessment of the Periotest device in baseline mobility measurements of craniofacial implants. Int J Oral Maxillofac Implants 1995:10:221-9
  8. Olive J, Apricio C. The Periotest method as a measure of osseointegrated oral implant stability. Int J Oral Maxillofac Implants 1990:5:390-400
  9. Sullivan DY, Sherwood RL, Collins T et al. The reverse torque test: a clinical report. Int J Oral Maxiilofac Implants 1996:11:179-85
  10. Johansson CB. Albrektsson T. A removal torque and histomorphometric study of commercially pure niobium and titanium implants in rabbit bone. Clin Oral Impl Res 1991:2:24-9 https://doi.org/10.1034/j.1600-0501.1991.020103.x
  11. Wennerberg A, Albrektsson T, Andersson B, Kro JJ. A histomorphometric and removal torque study of screw-shaped titanium implants with three different surface topographies. Clin Oral Impl Res 1995:6:24-30 https://doi.org/10.1034/j.1600-0501.1995.060103.x
  12. Johansson P, Strid KG. Assessment of bone quality from cutting resistance during implant surgery. Int J Oral Maxillofac Implants 1994:9:279-88
  13. Friberg B, Sennerby L, Roos J, Lekholm U. Identification of bone quality in conjunction with insertion of titanium implants. Clin Oral Impl Res 1995:6:213-9 https://doi.org/10.1034/j.1600-0501.1995.060403.x
  14. Friberg B, Sennerby L, Roos J, Johansson P, Strid K-G, Lekholm U. Evaluation of bone density using cutting resistance measurements and radiography. Clin Oral Impl Res 1995:6:164-71 https://doi.org/10.1034/j.1600-0501.1995.060305.x
  15. Meredith N, Alleyne D, Cawley P. Quantitative determination of the stability of the implant-tissue interface using resonance frequency analysis. Clin Oral Impl Res 1996:7:261-7 https://doi.org/10.1034/j.1600-0501.1996.070308.x
  16. Meredith N. Assessment of implant stability as a prognostic determinant. Int J Prosthodont 1988:11:491-501
  17. Meredith N, Shagaldi F, Alleyne D, Sennerby L, Cawley P. The application of resonance frquency measurements to study the stability of titanium implants during healing in the rabbit tibia. Clin Oral Impl Res 1997:8:234-43 https://doi.org/10.1034/j.1600-0501.1997.080310.x
  18. Sennerby L, Meredith N. Analisi della frequenza di risonanza (RFA). Osteointegrazione e Carico Immediato 2002:A-32
  19. Valderrama et al. Evaluation of Two different resonacne frequency devices to detect implant stability: A Clinical Trial. J Periodontol 2007:78:262-72 https://doi.org/10.1902/jop.2007.060143
  20. Van Oosterwyck H, Duyck J, Vander Sloten J, Van Der Perre G, Naert I. Periimplant bone tissue strains in cases of dehiscence: A finite element study. Clin Oral Implants Res 2002:13:327-33 https://doi.org/10.1034/j.1600-0501.2002.130314.x