Abstract
The aim of this study is to acquire useful information of lower urinary tract symptom (LUTS) diagnosis through urophonography signal as a noninvasive method. The hardware and software which could evaluate the function of compensatory hypertrophy with noninvasive and comfortable method was implemented to measure uroflow and urophonography signal during urination. The PSD (power spectrum density) and the log-log plot gradient analysis were accomplished in frequency domain. For evaluation of the system and analysis method, a model system for the lower urinary system of men was used. From the evaluation of the model system, the PDS and the log-log plot gradient were dependent on the occlusion degree significantly. In a pilot study on normal and abnormal male subjects, the PSD and the log-log plot gradient were highly correlated with the artificial urethral obstruction.