Abstract
Computational and experimental dosimetry of Henschke applicator with respect to high dose rate brachytherapy using the MIRD phantom and a remote control afterloader were performed. A comparison of computational dosimetry was made between the simulated Monte Carlo dosimetry and GAMMADOT brachytherapy Planning system's dosimetry. Dose measurements was performed using ion chamber in a water phantom. Dose rates are calculated using Monte Carlo code MCNP4B and the GAMMADOT. Thecomputational models include the detailed geometry of Ir-192 source, tandem tube, and shielded ovoids for accurate estimation. And transit dose delivered during source extension to and retraction from a given dwell position was estimated by Monte Carlo simulations. Point doses at ICRU bladder/rectal pointswhich have been recommened by ICRU 38 was assessed. Calculated and measured dose distribution data agreed within 4% each other. The shielding effect of ovoids leads to 19% and 20% dose reduction at bladder surface and rectal points.
Manchester system 타입의 장착기중 상, 하부에 차폐체가 장착되어 있는 Henschke 장착기를 이용하여 자궁암 근접치료시 자궁 및 주변장기의 선량분포를 평가하기 위하여 치료계획수립에 사용되는 실용프로그램 결과와 몬테칼로 모의계산 결과를 비교하였다. 또한 자궁 및 주변 정상조직이 받은 선량을 계산하기 위해 ORNL(Oak Ridge National Laboratory)에서 수립한 여성의 MIRD (Medical Internal Radiation Dose)형 모의피폭체를 이용 하여 주변장기가 받는 선량을 MCNP로 계산하였다. 몬테칼로 모사에는 MCNP 4B코드를 사용하였으며, 실용계산프로그램에는 GAMMADOT를 이용하였다 MCNP계산에는 $^{192}Ir$ 선원과 장착기의 기하학적 모양을 정밀하게 모사하여 계산 오차를 줄이도록 하였으며, 치료계획용 실용계산프로그램의 계산 조건과 동일하게 치료선원의 강내 체류시간과 체류위치를 적용하여 선량을 계산하였다. 주요 선량 비교 평가점은 Manchester system에서 사용되는 4곳과 ICRU 38에서 Manchester system을 보완하기 위해 제시한 방광표면 및 직장이였다. 실용계산 결과는 MCNP모의계산의 결과와 비교했을 때 대부분 위치에서 상대오차 4% 이내의 결과를 보였고, 난형체의 차폐체 장착효과로 인한 방광과 직장에서의 선량감쇠효과는 각각 19%, 20%였다.