The Development of Extravasation Detection Accessory System for the Preventive Contrast Media Extravasation in the Computed Tomography: A Preliminary Report

CT 검사에서 조영제의 혈관외유출 예방을 위한 EDA 시스템 개발의 예비 보고

  • Kweon Dae-Cheol (Departments of Diagnostic Radiology, Seoul National University Hospital) ;
  • Jeong Seok-Hee (Departments of Biomedical Engineering, Seoul National University Hospital) ;
  • Kim Tae-Hyung (Department of Radiology, Asan Medical Center) ;
  • Kim Jeong-Goo (Department of Radiological Science, Hanseo University) ;
  • Park Peom (Department of Industrial and Information Systems Engineering, Ajou University)
  • 권대철 (서울대학교병원 진단방사선과) ;
  • 정석희 (서울대학교병원 의공학과) ;
  • 김태형 (서울아산병원 방사선과) ;
  • 김정구 (한서대학교 방사선학과) ;
  • 박범 (아주대학교 산업정보시스템공학과)
  • Published : 2006.03.01

Abstract

To assess the ability of an extravasation detection accessory (EDA) system to detect clinically important extravascular injection of iodinated contrast material delivered with an automated power injector. Fifty patients referred for contrast material enhanced body computed tomography studied in a prospective, observation study in which the EDA system was used to identify and interrupt any injection associated with clinically Important extravasation. The presence or absence of extravasation was definitively established with multi-detector CT at the injection site (injection rate, $2.0{\sim}2.5$ mL/sec). There were two true positive, extravasation volumes $22{\sim}25$ mL. The EDA system had sensitivity in the detection of clinically important extravasation. The EDA system is easy to use, safe, and accurate In the monitoring of intravenous injections for extravasation, which may prove especially useful in CT applications.

CT 검사에서 조영제의 혈관외유출을 예방하기 위해 EDA시스템을 증폭기, ADC, 스트레인 게이지로 구성하여 개발하였고, 스트레인 게이지의 신호의 수치를 나타내도록 프로그램을 설계하였다. 신호비율은 500/1000 sec, 분해능 값은 0에서 4,096범위로 표시하고, 입력범위 $0{\sim}5$ V이다. 혈관외유출 검출 문턱 값은 50으로 하였고, 10 간격으로 조정하도록 하였다. 개발된 EDA 시스템을 성인 50명을 대상으로 적용하였다. 2명의 환자에서 혈관외유출이 검출되었고, 조영제 주입속도는 2.0 mL/sec, 2.5 mL/sec, 압력은 137 psi, 158 psi, 혈관외유출된 조영제 용량은 22 mL, 25 mL였다. 설정된 주입 조영제에 비해 20% 정도의 적은 용량이 혈관외유출되었다. 개발된 EDA 시스템은 소량의 조영제를 검출하여 혈관외유출을 예방하는 효과가 있다.