An Image-Guided Robotic Surgery System for Spinal Fusion

  • Chung Goo Bong (School of Electrical Engineering and Computer Science, Hanyang University) ;
  • Kim Sungmin (Department of Biomedical Engineering, Hanyang University) ;
  • Lee Soo Gang (Automation R&D Center, LG Indestrial Systems Co., Ltd.) ;
  • Yi Byung-Ju (School of Electrical Engineering and Computer Science, Hanyang University) ;
  • Kim Wheekuk (Department of Control and Instrumentation Engineering, Korea University) ;
  • Oh Se Min (School of Electrical Engineering and Computer Science, Hanyang University) ;
  • Kim Young Soo (College of Medicine, Hanyang University) ;
  • So Byung Rok (School of Electrical Engineering and Computer Science, Hanyang University) ;
  • Park Jong Il (Devision of Electrical and Computer Engineering, Hanyang University) ;
  • Oh Seong Hoon (College of Medicine, Hanyang University)
  • 발행 : 2006.02.01

초록

The goal of this work is to develop and test a robot-assisted surgery system for spinal fusion. The system is composed of a robot, a surgical planning system, and a navigation system. It plays the role of assisting surgeons for inserting a pedicle screw in the spinal fusion procedure. Compared to conventional methods for spinal fusion, the proposed surgical procedure ensures minimum invasion and better accuracy by using robot and image information. The robot plays the role of positioning and guiding needles, drills, and other surgical instruments or conducts automatic boring and screwing. Pre-operative CT images intra-operative fluoroscopic images are integrated to provide the surgeon with information for surgical planning. Some experiments employing the developed robotic surgery system are conducted. The experimental results confirm that the system is not only able to guide the surgical tools by accurately pointing and orienting the specified location, but also successfully compensate the movement of the patient due to respiration.

키워드

참고문헌

  1. M. Shoham, M. Burman, L. Joskowicz, E. Batkilin, and Y. Kunicher, 'Bone-mounted miniature robot for surgical procedures: Concept and clinical applications,' IEEE Trans. on Robotics and Automation, vol. 19, no. 5, pp. 893-901, 2003 https://doi.org/10.1109/TRA.2003.817075
  2. G. B. Chung, S. K. Lee, S. M. Oh, B.-J. Yi, W. K. Kim, Y. S. Kim, J. I. Park, and S. H. Oh, 'Development of SPINEBOT for spine surgery,' Proc. of IEEE/RSJ Int. Conference on Intelligent Robots and Systems, pp. 3942-3947, 2004
  3. J.-M. M.-Thiong, H. Labelle, M. Rooze, V. Feipel, and C.-E. Aubin, 'Evaluation of a transpedicular drill guide for pedicle screw placement in the thoracic spine,' European Spine Journal, vol. 12, no. 5, pp. 542-547, 2003 https://doi.org/10.1007/s00586-003-0549-4
  4. S. Porada, P. A. Millner, N. Chibverton, E. Berry, and B. B. Seedhom, 'Computer aided surgery with lumbar vertebra drill-guides, using computer aided planning, design and visualization,' Poster Session in Proc. Medical Image Understanding and Analysis, 2001
  5. W. W. Choi, B. A. Green, and A. D. O. Levi, 'Computer-assisted fluoroscopic targeting system for pedicle screw insertion,' Neurosurgery, vol. 47, no. 4, pp. 872-878, Oct. 2000 https://doi.org/10.1097/00006123-200010000-00017
  6. J. Wahrburg, I. Gross, P. Knappe, S. Pieck, S.Kunzler, and F. erschbaumer, 'An interactive mechatronic assistance system to support surgical interventions,' Proc. of Computer Assisted Radiology and Surgery, pp. 431-436, 2004 https://doi.org/10.1016/j.ics.2004.03.163
  7. K. Birnbaum, E. Schkommodau, N. Decker, A. Prescher, U. Klapper, and K. Radermacher, 'Computer-assisted orthopedic surgery with individual templates and comparison conventional operation method,' Spine, vol. 26, no. 4, pp. 365- 370, 2001 https://doi.org/10.1097/00007632-200102150-00012
  8. E. W. Fritsch, 'Navigation in spinal surgery using fluoroscopy,' Navigation and Robotics in Total Joint and Spine Surgery, pp. 487-494, 2004
  9. H. C. Sagi, R. Manos, R. Benz, N. R. Ordway, and P. J. Connolly, 'Electromagnetic field-based image-guided spine surgery part one: Reults of a cadaveric study evaluating lumbar pedicle screw placement,' Spine, vol. 28, no. 17, pp. 2013-2018, 2003 https://doi.org/10.1097/01.BRS.0000087851.51547.00
  10. J. Geering, U. Berlemann, B. Frericks, M.kfuri, T. Hufner, and C. Krettek, 'Pedicle screw placement,' Navigation and Robotics in Total Joint and Spine Surgery, pp. 481-486, 2004
  11. J. de Wall Malefijt, 'Image_guided surgery of the spine,' Medica Mundi, vol. 42, no. 1, pp. 38-43, March 1998
  12. http://www.spine.org/
  13. http://www.spineuniverse.com/
  14. D. R. Jacobson and S. J. Larson, 'CT determination of pedicle canal diameter,' Proc .of IEEE Engineering in Medicine and Biology Society, p. 491, 1989
  15. D. P. Gwynne Jones, P. A. Robertson, B. Lunt, and S. A. Jackson, 'Radiation exposure during fluoroscopically assisted pedicle screw insertion in the lumbar spine,' Spine, vol. 25, no. 12, pp. 1538-1541, 2000 https://doi.org/10.1097/00007632-200006150-00013
  16. C. J. Schulze, E. Munzinger, and U. Weber, 'Clinical relevance of accuracy of pedicle screw placement: A computed tomographic-supported analysis,' Spine, vol. 23, no. 20, pp. 2215-2220, Oct. 1998 https://doi.org/10.1097/00007632-199810150-00014
  17. L. Wiesner, R. Kothe, K. P. Schulitz, and W. Ruther, 'Clinical evaluation and computed tomography scan analysis of screw tracts after percutaneous insertion of pedicle screws in the lumbar spine,' Spine, vol. 25, no. 5, pp. 615-621, 2000 https://doi.org/10.1097/00007632-200003010-00013
  18. W. H. M. Castro, H. Halm, J. Jerosch, J. Malms, J. Steinbeck, and S. Blasius, 'Accuracy of pedicle screw placement in lumbar vertebrae,' Spine, vol. 21, no. 11, pp. 1320-1324, June 1996 https://doi.org/10.1097/00007632-199606010-00008
  19. http://www.ndigital.com/polaris_technical.html
  20. G. B. Chung, B. -J. Yi, I. H. Suh, W. K. Kim, and W. K. Chung, 'Design and analysis of a spatial 3-DOF micro-manipulator for tele-operation,' Proc. of IEEE/RSJ Int. Conference on Intelligent Robots and Systems, pp. 337-342, 2001
  21. K. S. Arun, T. S. Huang, and S. D. Blostein, 'Least-squares fitting of two 3-D point sets,' IEEE Trans. on Pattern Analysis and Machine Intelligence, vol. 9, no. 5, pp. 698-700, 1987 https://doi.org/10.1109/TPAMI.1987.4767965
  22. J. H. Challis, 'A procedure for determining rigid body transformation parameters,' Journal of Biomechanics, vol. 28, no. 6, pp. 733-737, 1995 https://doi.org/10.1016/0021-9290(94)00116-L