Abstract
In this study, an attempt was made to measure uncertainties involved in the VOC analysis for the VOC working standards prepared by a dilution technique using Tedlar bags. For this purpose, VOC standard gases of benzene, toluene, xylene, and styrene were prepared at four different concentrations (4, 8, 20, and 40 ppb). These standard samples were then loaded on to the GC system equipped with air server/thermal desorption (AS/TD) system. Each of these four standard concentrations was analyzed individually to derive their respective calibration results. These calibration data sets were then compared across four different compounds. According to this comparison, differences in calibration patterns were moderately insignificant within the selected concentration range of 4~40 ppb. It was also observed that the loss of styrene standard was fairly high compared to other VOCs investigated simultaneously. The results of our study suggest that the analytical uncertainty associated with the preparation of VOC starndard gas using a dilution technique can be assessed in a fairly reasonable manner for samples with a narrow concentration range.
본 연구에서는 Tedlar bag을 사용하여 조제한 작업용 기체상 표준시료를 열탈착과 GC/FID를 조합한 시스템으로 분석할 때, 발생 가능한 오차를 평가하고자 하였다. 이를 위해, 여러 가지 VOC성분(Benzene, Toluene, p-Xylene, Styrene)을 함유하는 표준시료를 4 가지 초기농도대(4~40 ppb)에서 조제하였다. 그리고 이들 작업용 시료를 이용하여, 각각의 검량결과를 도출하고 비교분석을 실시 하였다. 본 실험에서 설정한 4~40 ppb 농도대의 검량결과를 비교하기 위해, 공기분배/저온농축/열탈착 전용기(Air server/thermal desorber (AS/TD))와 조합한 GC/FID 시스템을 사용하였다. 본 연구의 결과, 표준시료의 조제방식이나 그와 연관된 초기농도의 차이가 검량 특성에 영향을 미친다고 하기는 어렵다는 결론을 내릴 수 있었다. 그러나 테들러백을 사용하는 경우, 여타 성분들에 비해 Styrene성분에서 감도가 떨어지는 경향이 크게 나타나는 점을 확인할 수 있었다.