How Image Quality Affects Determination of Target Displacement When Using kV Cone-beam Computed Tomography (CBCT)

kV Cone-beam CT를 사용한 치료준비에서 재구성 영상의 품질이 표적 위치 결정에 미치는 영향

  • Oh, Seung-Jong (Department of Biomedical Engineering, The Catholic University of Korea College of Medicine) ;
  • Kim, Si-Yong (Department of Radiation Oncology Mayo Clinic, Jacksonville, USA) ;
  • Suh, Tae-Suk (Department of Biomedical Engineering, The Catholic University of Korea College of Medicine)
  • Published : 2006.12.31

Abstract

The advent of kV cone-beam computed tomography (CBCT) integrated with a linear accelerator allows for more accurate Image-guided radiotherapy (IGRT). IGRT is the technique that corrects target displacement based on internal body information. To do this, the CBCT Image set is acquired just before the beam is delivered and registered with the simulation CT Image set. In this study, we compare the registration results according to the CBCT's reconstruction quality (either high or medium). A total of 56 CBCT projection data from 6 patients were analyzed. The translation vector differences were within 1 mm in all but 3 cases. For rotation displacement difference, components of all three axes were considered and 3 out of 168 ($56{\times}3$ axes) cases showed more than lo of rotation differences.

kV cone-beam CT (CBCT)가 결합된 선형가속기의 등장으로 더욱 정확한 영상유도 방사선치료(Image-guided radiation therapy, IGRT)가 가능해졌다. IGRT는 영상장비를 이용해 표적의 이동을 보정할 수 있는 방사선치료기술로, CBCT를 이용한 IGRT의 경우 내부 장기의 정보를 바탕으로 병변의 이동을 정확히 알 수 있다. 내부 장기의 정보를 얻기 위해서는 방사선이 조사되기 바로 전에 환자의 CBCT 영상을 획득하여 치료계획 시 사용한 모의치료 CT영상과 정합을 수행하게 된다. 본 연구에서는 CBCT 영상의 재구성 품질에 따른 정합결과를 비교해 보았다. 6명의 환자로부터 총 56개의 CBCT 투과 정보를 획득하여 분석한 결과, 평행이동벡터의 차가 1 mm를 초과하는 경우는 단 3개에 불과하였다. 회전이동의 경우, x, y, z, 세 축을 모두 고려하였으며, 그 결과 총 168 ($56{\times}3$)개에서 단 3개만이 $1^{\circ}$ 이상의 차이를 나타냈다.

Keywords